Use Instructions

Vital Signs Monitor User Manual Ward Round Mode
- 25 -
6.2 Editing Patient Information
To edit the patient information after a patient has been admitted, click the patient information
area in main interface to enter Patient Info window, and make the required changes on the popup
interface. Click the icon to confirm the operation.
If the monitor is powered off and restarted, the patient information will be restored by default,
that is, the patient type is adult and the other information is blank.
6.3 Import Patient
In Menu > Patient Info > Import Patient window:
If ADT Query is on and network server is connected, user can set the query conditions
(Admission Date, Department), and click query icon to show the patients. Choose the
patients as needed and click Import icon . After successful import, the imported patients
will be displayed in the window of Select Patient, and marked with icon .
NOTE:
In importing process, the monitor will import the patient information that meets the
requirements, and the wrong patient information will be prompted. The wrong patient
information needs to be revised and re-imported.
During importing, if the total patients in the monitor exceed upper limit (1000 patients), there will
be a prompt. User can delete part of patients and import again.
6.4 Refresh Patient List
Click Select Patient shortcut key to enter the window:
If ADT Query is on and network server is connected, click to set query conditions
(Admission Date, Department), click to confirm, and then click to refresh patient
list. All eligible patients will be refreshed to the window of Select Patient, and marked with icon
.
NOTE:
1 The operation of importing will keep the patients those exist in the monitor but not
found in network. However, the operation of refreshing will delete those patients.
2 If MRN or the bed No. has conflict, the monitor will replace all conflicting patient
information. But for MRN conflict, the patient measurement status will keep
unchanged as the previous patient, and for bed No. conflict, it will recover to the initial
unmeasured status.