Vital Signs Monitor User Manual Networking
- 31 -
If TLS is enabled, click Import Certificate to upgrade the Certificate via USB flash drive.
The self-signed certificate of manufacturer is provided by default. However, the certificate
issued by Certificate Authority (CA) is recommended and self-signed certificate should be
avoided.
The certificate format is as follows:
For CA certificate, the fixed name ca.cer‘ is supported;
For client certificate, the fixed name client.cer‘ is supported;
For private key file of the client, the fixed name server.cer‘ is supported;
For sever certificate, the fixed name client.pem‘ is supported;
For private key file of the server, the fixed name server.pem‘ is supported;
NOTE
1 Be aware that some network-based functions may be limited for monitors on wireless
networks in comparison with those on wired networks.
2 The obstacle may interfere with data transmission and even cause data loss.
3 Use the wireless device recommended by EDAN, otherwise some exceptional
situations such as frequent network disconnection may occur on the monitor.
4 When signal intensity is level 2 or less, signal may be unstable and quality of the
signal transmission may be degraded.
5 The device can be connected to the following encrypted Wi-Fi hotspots:
OPEN (unencrypted);
WEP/WPA/WPA2;
WARNING
If Wi-Fi is unavailable, please restart the monitor (refer to Section Host) to restore Wi-Fi
function under the precondition of ensuring patients safety.
8.3 e-Link Function
A monitor with e-link module can transmit data through e-link and also displays e-link
connection status in the icons area.
To enable e-link function:
1. Select Menu > e-link to enable the function.
2. User can view the name of monitor device.
5. Select data transmission direction: Passive Connection or Active Connection.
Passive Connection means sending the review data to network system of medical
institutions through HL7 protocol.
Before connection, please enter User Maintain to set e-link Key, and click Server >