1. Disconnect the patient from the monitor and stop all monitoring and measurement.
2. Switch the monitor power on and fully charge the battery.
3. Disconnect AC power and let the monitor run until there is no battery power left and the
monitor shuts off.
4. The running time of the battery reflects the battery performance.
If the running time is obviously less than the specified time in the specification, please change the
battery or contact the service personnel. If the running time meets the specification, fully charge
the battery again for use or charge it to 40%-60% for storage.
18.5 Replacing the Battery
To install or replace the battery, please follow the procedure:
Lithium-ion battery
1. Push the battery latch down according to icon indication.
2. Grab the handle on the bottom of battery to pull battery out.
3. Insert the new battery into the battery compartment.
4. Place the metal terminal of the new battery to right-front, and close to the front end of the
groove (as shown below). Gently press it (do not push battery from back to front) until the
battery is fully installed in the groove.
5. Place the bottom of battery door at the bottom of groove. Gently press battery door and bolt
up the latch to close the door.
AA battery
Please follow the positive and negative direction in the groove to install battery.
18.6 Recycling the Battery
When the battery no longer holds a charge, it should be replaced. Remove the old battery from
the monitor and recycle it properly.