Use Instructions

Vital Signs Monitor User Manual Care and Cleaning
- 83 -
water at least for 1 minute until no visible cleaning agent remains.
7. Dry the host, the protection shell and the charger stand in a ventilated and cool place.
8. If the host, the protection shell or the charger stand is not visually clean at the end of the
cleaning steps, please repeat the cleaning steps through step 4 to step 7.
9. Inspect the host, the protection shell or the charger stand to ensure that there is no damage.
10. Reassemble the protection shell to the host, and then reassemble the host with the protective
shell to the charger stand.
19.3.2 Cleaning the Reusable Accessories
19.3.2.1 Cleaning the Blood Pressure Cuff
Cleaning the Cuff:
1. Disassemble NIBP Cuff from the monitor, and take out the air bladder.
2. Remove all residual foreign matters from the surface of cuff and air bladder using sterile
cloth or paper towel immediately after examination until the surface is visually clean.
3. Hand wash the cuff with the cleaning solution; clean the air bladder with a soft cloth
dampened with the cleaning solution. Clean the surface of the cuff and the air bladder
thoroughly until no visible contaminants remain
4. Rinse the cuff and after cleaning, wipe off the cleaning solution with a fresh cloth or towel
dampened with tap water until no visible cleaning agent remains.
5. Wipe off residual moisture with a dry cloth.
6. Air dry the cuff thoroughly after cleaning.
7. If the cuff and the air bladder are not visually clean at the end of the cleaning steps, please
repeat the cleaning steps through step 4 to step 6.
8. Inspect the cuff and the air bladder to ensure that there is no damage.
Replacing the Air Bladder:
After cleaning, replace the air bladder into the cuff following the steps below:
1. Roll the bladder lengthwise and insert it into the cuff from the large opening at one end of
the cuff.
2. Thread the hose from within the cuff and out through the small hole at the top of the cuff.
3. Adjust the bladder until it is in position.
19.3.2.2 Cleaning the SpO
2
Sensor
1. Disassemble SpO
2
sensor from the monitor.
2. Remove all residual foreign matters from the surface of SpO
2
Sensors, including cables,
using sterile cloth or paper towel immediately after examination until the surface is visually
clean.