Patient Monitor User Manual Product Specification
- 179 -
Stability
Short term drift: drift over 4 hours < 0.8 mmHg
Long term drift: 120 hours
Warm-up time
Display reading within 20s; reach to the designed accuracy within 2
minutes.
Rise time 400ms (typical value, using water trap, sample gas flowrate:100ml/min
Response time
<4s (water trap) with 2m gas sampling tube, sample gas flowrate:
100ml/min
Work mode Standby, measure; default: measure
Respiratory
inspection
The value of concentration change is greater than 1 vol%.
O
2
compensation
Range: 0%~100%
Resolution: 1%
Default: 16%
N
2
O
compensation
Range: 0%~100%
Resolution: 1%
Default: 0%
AG
compensation
Range: 0%~20%
Resolution: 0.1%
Default: 0%
Humidity
compensation
method
ATPD, BTPS (default)
Calibration Support
Alarm EtCO
2
, FiCO
2
, AwRR
Apnea alarm
delay
10s, 15s, 20s, 25s, 30s, 35s, 40s; default value is 20s.
Respironics Module
Intended patient Adult, pediatric, neonatal
Measurement method Infra-red Absorption Technique
Unit mmHg, %, Kpa
Measuring Range
EtCO
2
0 mmHg ~ 150 mmHg
FiCO
2
3 mmHg ~50 mmHg
AwRR 2 rpm ~ 150 rpm(sidestream)
0 rpm ~ 150 rpm(mainstream)
Resolution
EtCO
2
1mmHg