Patient Monitor User Manual Alarm Information
- 41 -
Message Cause Alarm Level Action Taken
NIBP Leak
NIBP pump, valve, cuff or
tube has a leakage.
Low
Properly wrap the
cuff and connect the
tube. Measure again;
if failure persists,
stop measuring
function of NIBP
module and notify
biomedical engineer
or manufacturer’s
service staff.
NIBP Excessive
Pressure
Pressure has exceeded the
specified upper safety
limit.
Low
Measure again, if
failure persists, stop
measuring function
of NIBP module and
notify biomedical
engineer or
manufacturer’s
service staff.
NIBP Init Pressure
High
The initial pressure is too
high during measuring
Low
Measure again, if
failure persists, stop
measuring function
of NIBP module and
notify biomedical
engineer or
manufacturer’s
service staff.
NIBP Self Test Error
Sensor or other hardware
errors.
High
If failure persists,
stop using measuring
function of NIBP
module and notify
biomedical engineer
or manufacturer’s
service staff.
NIBP Cuff Type
Error
The cuff type used isn’t
consistent with the patient
type.
Low
Confirm the patient
type and change the
cuff.