User's Manual

Patient Monitor User Manual Alarm Information
- 52 -
RESP alarm limits are listed as follows: unit (rpm)
Patient Type ALM HI ALM LO
ADU 120 6
PED 150 6
NEO 150 6
SpO
2
alarm limits are listed as follows (unit %):
ALM HI ALM LO
SpO
2
100 0
PR
alarm limits is listed as follows: unit (bpm)
ALM HI ALM LO
PR 300 30
NIBP alarm limits are listed as follows (EDAN module): unit (mmHg)
Patient Type ALM HI ALM LO
ADU SYS
DIA
MAP
270
215
235
40
10
20
PED SYS
DIA
MAP
200
150
165
40
10
20
NEO SYS
DIA
MAP
135
100
110
40
10
20
NIBP alarm limits are listed as follows (Omron module): unit (mmHg)
Patient Type ALM HI ALM LO
ADU(PED)
SYS
DIA
MAP
250
200
235
60
40
45
NEO
SYS
DIA
MAP
120
90
100
40
20
30