User's Manual

Patient Monitor User Manual Alarm Information
- 53 -
TEMP alarm limits are listed as follows:
ALM HI ALM LO
T1
50qC (122 º F) 0qC (32 º F)
T2
50qC (122 º F)
0qC (32 º F)
TD
50qC (90 º F)
0qC (0 º F)
IBP alarm limits are listed as follows: unit (mmHg)
ALM HI ALM LO
Art 300 0
RAP 40 -10
LAP 40 -10
CVP 40 -10
PA 120 -6
ICP 40 -10
P1 300 -50
P2 300 -50
CO
2
alarm limits are listed as follows:
ALM HI ALM LO
EtCO
2
150 mmHg 0
FiCO
2
50 mmHg 3
AwRR 150 rpm 2 rpm
C.O. alarm limits are listed as follows:
ALM HI ALM LO
TB
43qC (109.4) 23qC(73.4)
Quick Temp alarm limits are listed as follows:
Patient Type ALM HI ALM LO
ADU
42qC (107.6) 35.5qC (95.9)
PED