User's Manual

Size:
Length*Width* Height: (48±1) mm× (39±1) mm× (145±3 ) mm
Weight:
<200g(including batteries)
LCD:
Size:
1.3 inch23.4 mm 23.4 mm
Resolution
240240(Pixels)
Coupling
Gel:
pH: 5.5~8.0
Acoustic Impedance: 1.5x10
6
Pa.s/m ~1.7x10
6
Pa.s/m (35°C/95ºF )
Environment
Working:
Temperature:+5°C ~ +40°C ( +41ºF ~ +104ºF)
Humidity:15% RH ~ 95% RH(non-condensing)
Atmospheric Pressure:70 kPa ~ 106 kPa
Transport and Storage:
Temperature:-25°C ~ +70°C (-13ºF ~ +158ºF)
Humidity:15% RH ~ 95% RH (non-condensing)
Atmospheric Pressure:70 kPa ~106 kPa
Note: The time required for the Doppler to warm from the minimum storage temperature between uses
until it is ready for intended use is at least 2 hours; the time required for the Doppler to cool from the
maximum storage temperature between uses until it is ready for intended use is at least 2 hours
Performance Specifications
FHR (Essential
Performance):
FHR Measuring Range: 30 bpm ~ 240 bpm
Accuracy: ±2 bpm
Note: FHR measurement result may not be accurate if the equipment is
measuring beyond its measuring range.
FHR Resolution:
1bpm
Audio Output:
Output Power: 2w
Background noise: <45dBA
Overall Sensitivity:
>90dB
Auto Power-off:
Power off when the Doppler receives no signal or operation for 1
minute.
Bluetooth:
TransmissionRange (Without Obstacles) :> 5m (Indoor range depends
on the building’s structure and material.)
Intelligent Denoising
Noise generated in static state100mV
Shock Resistance
Withstands a 1.3 m drop at least 6 times to concrete surface with
possible cosmetic damage only
Ultrasound:
Frequency: (3.0±10%) MHz; (2.0±10%) MHz
p_<1 MPa
Iob<10 mW/cm
2
Ispta<100 mW/cm
2
Isata<10 mW/cm
2
Isppa.3<190 W/cm
2
Ispta.3<94 mW/cm
2
Effective Radiating Area: 490mm
2
± 15%
Ultrasound output power<49mW
Working Mode: pulse wave
Pulse wave repetition rate:5KHz (±5%)
Pulse wave duration:75µs (±5%)
SpO2:
Measurement range:70%~100%
Resolution:1%
Accuracy:±2%
SpO2 Sensor:
Red light:660nm
Infrared light:905 nm
Wave lengthabout660nmand 900nm
Emitted light energy<15mW
PR:
Measurement range:30~240bpm
Resolution:1bpm
Accuracy:±2bpm or ±3%,whichever is greater
Battery Specifications
Specification:
Two AA alkaline batteriesTwo AA rechargeable NI-MH batteries
Working Duration:
4h
Bluetooth Specifications
Modulation:
GFSK π /4-DQPSK 8DPSK
Frequency:
2400-2483.5MHz
Tolerance Frequency:
≤ 20ppm
RF output power:
≤ 20dBm (EIRP)
Occupied Channel Bandwidth:
≤ 2MHz
Transmitter Unwanted Emissions:
30dBm
Low Output Summary Table
(For systems whose global maximum valuedoes not exceed 1.0)
System: SD2 Ultrasonic Pocket Doppler
Model (MHz)
I
spta.3
(mW/cm
2
)
TI Type
TI Value
MI
I
sppa.3
(W/cm
2
)
SD2 CD3.0
3.25
TIS
0.032
0.01
0.009
TIB
0.11
SD2 CD2.0
4.21
TIS
0.028
0.013
0.012
TIB
0.086
Ordering Information
CAUTION
Only the parts supplied by the manufacturershould be used with the Doppler.
Parts
Part Number
Parts
Part Number
SD2 Doppler
02.06.263378
SD2 Lite Doppler
02.06.263382
SD2 Pro Doppler
02.06.263379
Rechargeable NI-MH
battery
21.21.064180
SD2 Plus Doppler
02.06.263380
AA Alkaline Battery
01.21.064086
SD2 Basic Doppler
02.06.263381
Zipper bag
01.56.466428
Ultrasound Intensity and Safety
Ultrasound in Medicine
The use of diagnostic ultrasound has proved to be a valuable tool in medical practice. Given its known
benefits for non-invasive investigations and medical diagnosis, including investigation of the human fetus,
the question of clinical safety with regards to ultrasound intensity arises.
There is no easy answer to the question of safety surrounding the use of diagnostic ultrasound
equipment. Application of the ALARA (As Low As Reasonably Achievable) principle serves as a
rule-of-thumb that will help you to get reasonable results with the lowest possible ultrasonic output.
The American Institute of Ultrasound in Medicine (AIUM) states that given its track record of over 25
years of use and no confirmed biological effects on patients or instrument operators, the benefits of the
prudent use of diagnostic ultrasound clearly outweigh any risks.
Ultrasound Safety and the ALARA Principle
Ultrasound waves dissipate energy in the form of heat and can therefore cause tissue warming. Although
this effect is extremely low with Doppler, it is important to know how to control and limit patient exposure.
Major governing bodies in ultrasound have issued statements to the effect that there are no known
adverse effects from the use of diagnostic ultrasound, however, exposure levels should always be limited
to As Low As Reasonably Achievable (the ALARA principle).
Explanation of MI/TI
MI (Mechanical Index)
Cavitations will be generated when ultrasound wave passes through and contacts tissues, resulting in
instantaneous local overheating. This phenomenon is determined by acoustic pressure, spectrum, focus,
transmission mode, and factors such as states and properties of the tissue and boundary. This
mechanical bioeffect is a threshold phenomenon that occurs when a certain level of ultrasound output is
exceeded. The threshold is related to the type of tissue. Although no confirmed adverse mechanical
effects on patients or mammals caused by exposure at intensities typical of present diagnostic
ultrasound instruments have ever been reported, the threshold for cavitation is still undetermined.
Generally speaking, the higher the acoustic pressure, the greater the potential for mechanical bioeffects;
the lower the acoustic frequency, the greater the potential for mechanical bioeffects.
The AIUM and NEMA formulate mechanical index (MI) in order to indicate the potential for mechanical
effects. The MI is defined as the ratio of the peak-rarefactional acoustic pressure (should be calculated
by tissue acoustic attenuation coefficient 0.3 dB/cm/MHz) to the acoustic frequency.
MI = Pr, α
fawf ×CMI
CMI = 1 (MPa / MHz )
TI (Thermal Index)
Heating of tissues is caused by absorption of ultrasound when the ultrasound energy is applied. The
temperature rise is determined by the acoustic intensity, exposed area and thermophysical properties of
the tissue.
In order to indicate the potential for temperature rise caused by thermal effects, the AIUM and NEMA
formulate thermal index (TI). It is defined as the ratio of the total acoustic power to the acoustic power
required to raise the tissue temperature by 1ºC (1.8°F).
According to different thermophysical properties of the tissue, TI is divided into three kinds: TIS, TIB and
TIC.
TIS (Soft Tissue Thermal Index): It provides an estimate of potential temperature rise in soft or similar
tissues.
TIB (Bone Thermal Index): It provides an estimate of potential temperature rise when the ultrasound
beam passes through soft tissue and a focal region is in the immediate vicinity of bone.
TIC (Cranial Bone Thermal Index): It provides an estimate of potential temperature rise in the cranial
bones or superficial bones.
Measurement Uncertainties
The uncertainties in the measurements were predominantly systematic in origin; the random
uncertainties were negligible in comparison. The overall systematic uncertainties were determined as
follows:
1. Hydrophone Sensitivity: ± 12 percent for intensity, ± 6 percent for pressure. Based on the
hydrophone calibration report by ONDA. The uncertainty was determined within ±1 dB in frequency
range 1-15 MHz.
2. Digitizer: ±0.3 percent for intensity. ± 0.15 percent for pressure.
Based on the stated accuracy of the 8-bit resolution of the Agilent DSO6012 Digital Oscilloscope
and the signal-to-noise ratio of the measurement.
3. Temperature:±2.4 percent for intensity uncertainty, ±1.2 percent for pressure uncertainty.
Based on the temperature variation of the water bath of ±C (1.8°F).
4. Spatial Averaging: ± 3.5 percent for intensity, ± 1.75 percent for pressure.
5. Non-linear Distortion: N/A.
No effects of nonlinear propagation were observed.
Since all the above error sources are independent, they may be added on an RMS basis, giving a
total uncertainty of ± 12.73 percent for all intensity values reported, ± 6.37 percent for all the pressure
values,,± 12.6 percent for the Mechanical Index, uncertainty of ±12.73% percent for power0.15 percent
for center frequency, ±6 .87% for the MI.
Prudent Use Statement
Although no confirmed bioeffects on patients caused by exposure from present diagnostic ultrasound
equipment have ever been reported, the potential exists that such bioeffects may be identified in the
future. Therefore, the ultrasound should be used prudently. High levels of acoustic output and long
exposure time should be avoided while acquiring necessary clinical information.
Acoustic Output Reporting Table for Track 1 Acoustic output reporting table for
IEC60601-2-37 (IEC60601-2-37, Edition 2.1, 2015-0, table 201.103)
Transducer Model: SD2, Operating Mode: PW mode
Index label
MI
TIS
TIB
TIC
At
Surface
Below
Surface
At
surface
Below
Surface
Maximum index value
0.013
0.028
0.086
N/A
Index component value
N/A
0.028
NA
0.086
Acoustic
Parameters
pr.αat zMI
(MPa)
0.019
P (mW)
8.62
8.62
N/A
P1x1 (mW)
N/A
N/A
zs (cm)
5.12
zb (cm)
4.97
zMI (cm)
0.50
zPII.α(cm)
.α
0.50
fawf (MHz)
2.00
3.00
3.00
N/A
Other
information
prr (Hz)
4999.00
srr (Hz)
N/A
npps
N/A
Ipa.α at zPII.α
(W/cm2)
0.012
Ispta.α at zPII.α
or
zSII.α(mW/cm2)
4.21
Ispta at zPII or
zSII (mW/cm2)
4.67
pr. at zPII
(MPa)
0.020
Operating control conditions
Fixed
N/A
Fixed
N/A
Fixed
N/A
Fixed
N/A
Fixed
N/A
Fixed
N/A
2.00
N/A
2.00
N/A
2.00
N/A
Transducer Model: SD2, Operating Mode: PW mode
Index label
MI
TIS
TIB
TIC
At
Surface
Below
Surface
At
surface
Below
Surface
Maximum index value
0.011
0.032
0.11
N/A
Index component value
N/A
0.032
NA
0.11
Acoustic
Parameters
pr.αat zMI
(MPa)
0.019
P (mW)
10.96
10.96
N/A
P1x1 (mW)
N/A
N/A
zs (cm)
4.65
zb (cm)
4.55
zMI (cm)
0.50
zPII.α (cm)
.α
0.50
fawf (MHz)
3.00
3.00
3.00
N/A
Other
information
prr (Hz)
4999.00
srr (Hz)
N/A
npps
N/A
Ipa.α at zPII.α
(W/cm2)
0.0090
Ispta.α at zPII.α
or
zSII.α(mW/cm2)
3.25
Ispta at zPII or
zSII (mW/cm2)
3.72
pr. at zPII
(MPa)
0.021
Operating control conditions
Fixed
N/A
Fixed
N/A
Fixed
N/A
Fixed
N/A
Fixed
N/A
Fixed
N/A
3.00
N/A
3.00
N/A
3.00
N/A
Standard Parameter Equal Contrast List
IEC60601-2-37 Standard Parameters
Parameter
Note
Parameter
Note
p
r.α
Attenuated
Peak-rare-factional
Acoustic Pressure
f
awf
Center Frequency, Acoustic
Working Frequency
p
r
Peak-rare-factional
Acoustic Pressure
X
-12dB Output Beam Dimensions
P
Output Power
Y
z
s
Depth for Soft Tissue
Thermal Index
t
d
Pulse Duration
P
α
(Z
s
)
Attenuated Output Power
prr
Pulse Repetition Frequency
(Pulse Repetition Rate)
I
ta.α
(Z
s
)
Attenuated
Temporal-average Intensity
d
eq
Equivalent Beam Diameter
z
bp
Break-point Depth
I
pi.α
at max
MI
Attenuated Pulse-average
Intensity at the point of Maximum
MI
z
b
Depth for Bone Thermal
Index
A
aprt
-12dB Output Beam Area
I
pi.α
Attenuated Pulse-intensity
Integral
MI
Mechanical Index
I
pi
Pulse-intensity Integral
TIS
Soft Tissue Thermal Index
d
eq
(Z
b
)
Equivalent Beam Diameter
at the point of Z
sp
TIB
Bone Thermal Index
TIC
Cranial-bone Thermal Index
EMC Information
Electromagnetic Emissions
Guidance and manufacturer’s declaration – electromagnetic emission
The SD2 Ultrasonic Pocket Doppler is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should assure that it is used in such an
environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The SD2Ultrasonic Pocket Doppler uses RF energy only
for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF emission
CISPR 11
Class B
The SD2 Ultrasonic Pocket Doppler is suitable for use in
all establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC/EN61000-3-2
Not
applicable
Voltage fluctuations
/flicker emissions
IEC/EN61000-3-3
Not
applicable
Electromagnetic Immunity
Guidance and manufacture’s declaration–electromagnetic immunity
The SD2 Ultrasonic Pocket Doppler is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should assure that it is used in such an
environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic
environment-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
8 kV contact
2 kV4 kV8 kV
15 kV air
8 kV
contact
2 kV4
kV8 kV
15 kV air
Floors should be wood,
concrete or ceramic tile.
If floor are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical Fast
Transient/Burst
IEC/EN61000-4-4
±2kV forpower
supplylines
±1kV forinput/outputlines
Not
applicable
Not applicable
Surge
IEC/EN61000-4-5
± 1 kV line(s) toline(s)
± 2 kV line(s) to earth
Not
applicable
Not applicable
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC/EN61000-4-11
<5%UT(>95% dip inUT)
for 0.5cycle
40%UT(60%dip in UT)
for5 cycles
70%UT(30%dip in UT)
for25 cycles
<5%UT(>95% dip inUT)
for 5s
Not
applicable
Not applicable
Power frequency
(50Hz/60Hz)
magnetic field
IEC61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
The SD2 Ultrasonic Pocket Doppler is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should assure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment-guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
Immunity to
Proximity fields
from RF
wireless
communications
equipment IEC
61000-4-3
3 V
rms
150 kHz ~ 80
MHz
6Vrmsc)in ISM
bands between
0,15 MHz and
80 MHz
10V/m
80 MHz ~ 2.7
GHz
See Table -Test
specifications
for
ENCLOSURE
PORT
IMMUNITY to
RF wireless
communications
equipment
for test level
Not Applicable
10 V/m
80 MHz to 2.7
GHz
See Table-Test
specifications
for
ENCLOSURE
PORT
IMMUNITY to
RF wireless
communications
equipment
for test level
Portable and mobile RF
communications equipment should
be used no closer to any part of the
SD2 Ultrasonic Pocket Doppler,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance:
d = 0.35 P
80 MHz to 800
MHz
d = 0.7 P
800 MHz to 2.7
GHz
d = 6 P/E
at RF wireless
communications equipment bands
(Portable RF communications
equipment (including peripherals
such as antenna cables and
external antennas) should be used
no closer than 30 cm (12 inches) to
any part of the SD2 Ultrasonic
Pocket Doppler, including cables
specified by the manufacturer).
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the
vicinity of equipment marked with
the following symbol:
NOTE 1:At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the SD2 Ultrasonic Pocket Doppler is used
exceeds the applicable RF compliance level above, the SD2Ultrasonic Pocket Doppler should
be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the SD2 Ultrasonic Pocket
Doppler.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
c
The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz
to6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70
MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5
MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to
14,2 MHz, 18,07 MHz to 18,17 MHz,21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz
to 29,7 MHz and 50,0 MHz to 54,0 MHz.
Table-Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications
equipment
Test
Frequenc
y (MHz)
Brand
a)
(MHz)
Service
a)
Modulatio
n
b)
Maximu
m
Power(W
)
Dista
nce
(m)
IMMUN
ITY
TEST
LEVEL
(V/m)
385
380-390
TETRA 400
Pulse
modulation
b)
18Hz
1.8
0.3
27
450
430-470
GMRS 460,
FRS 460
FM
C)
±5 kHz
deviation
1kHz sine
2
0.3
28
710
704-787
LTE Brand 13,
17
Pulse
modulation
b)
217 Hz
0.2
0.3
9
745
780
810
800-960
GSM
800/900,TETR
A 800, iDEN
820, CDMA
850, LTE Band
5
Pulse
modulation
b)
18 Hz
2
0.3
28
870
930
1720
1700-199
0
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3,
4,25;UMTS
Pulse
modulation
b)
217 Hz
2
0.3
28
1845
1970
2450
2400-257
0
Bluetooth,
WLAN,802.11
b/g/n, RFID
2450, LTE
Brand 7
Pulse
modulation
b)
217 Hz
2
0.3
28
5240
5100-580
0
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0.2
0.3
9
5500
5785
Note: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM maybe reduce to 1m. The 1 m test distance is
permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative FM modulation, 50% pulse modulation at 18 Hz may be used because while it
does not represent actual modulation, it would be worst case
Recommended Separation Distances
Recommended separation distances between portable and mobile RF
communications equipment and theSD2 Ultrasonic Pocket Doppler
The SD2 Ultrasonic Pocket Doppleris intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of theSD2 Ultrasonic
Pocket Doppler can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the SD2
Ultrasonic Pocket Doppleras recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter
(W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2 P
80 MHz to 800 MHz
d = 0.35 P
800 MHz to
2.7 GHz
d = 0.7 P
0.01
/
0.035
0.07
0.1
/
0.11
0.22
1
/
0.35
0.7
10
/
1.11
2.21
100
/
3.5
7
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Troubleshooting
Problem
Possible Cause
Solution
Fail to power
on, or shut
down shortly
after switching
on
Battery level is very low.
Replace the batteries or charge
the rechargeable NI-MH batteries
Battery is not installed properly.
Re-install the battery.
Fail to switch on the Doppler as
instructed.
Touch the On/Off touch key for a
while to power on the Doppler.
The Doppler has malfunctions.
Contact service personnel.
Loudspeaker
does not work.
Sound volume has been turned down
to the lowest level.
Adjust sound volume to
appropriate level.
If the Doppler is configured with
Bluetooth, fetal heart sound can be
played by mobile phone.
Set to play fetal heart sound by
mobile phone or the Doppler on
the APP.
The Doppler has malfunctions.
Contact service personnel.
FHR cannot be
displayed
stably.
There is strong interference source
such as high frequency machines and
mobile phones nearby.
Use the Doppler away from strong
interference sources.
The fetal heart position has changed
because of fetal movement.
Relocate the Doppler to the best
fetal heart rate detection position.
Friction between the Doppler and
patient’s abdomen causes false
displaying.
Find the best fetal heart rate
detection position.
Sensitivity is
low and noise
is too much.
There is strong interference source
such as high frequency machines and
mobile phones nearby.
Use the Doppler away from strong
interference sources.
The Doppler is not applied with
coupling gel.
Apply coupling gel to the Doppler.
The Doppler is not placed at the best
detection position.
Relocate the Doppler to the best
fetal heart rate detection position.
The Doppler malfunctions.
Contact service personnel.
-?- displayed
on the screen
The Doppler malfunctions.
Contact service personnel.
SpO
2
or
PRcannot be
displayed
normally
SpO2 sensor is not fully covered by
finger
Put the finger on the sensor again
Strong ambient light
Do not use the device in an
environment with high ambient
light
Patient is in low perfusion or Patients
oxyhemoglobin is too low to be
measured
See a doctor.
Finger is trembling or patient is moving
Please keep still.
The Doppler malfunctions.
Contact service personnel.
Accuracy Results in Clinical Studies
The table below shows Arms values measured with the investigational devicein a clinicalstudy.
SaO
2
Range
Arms
90%-100%
1.01
80%-90%
2.01
70%-80%
2.01
70%-100%
1.68
The figure below shows the Bland-Altman Plot of SaO
2
vs SpO
2
measured withthe investigational device.
In the plots, the upper and lower dotted lines represent the upper and inferior limits of the 95%
consistency, and the middle dotted line represents the average of the bias.
Warranty and Service
Warranty
EDAN warrants that EDAN’s products meet the labeled specifications of the products and will be free
from defects in materials and workmanship that occur within warranty period.
The warranty is void in cases of:
A. damage caused by mishandling during shipping.
B. subsequent damage caused by improper use or maintenance.
C.damage caused by alteration or repair by anyone not authorized by EDAN.
D.damage caused by accidents.
E. replacement or removal of serial number label and manufacture label.
If a product covered by this warranty is determined to be defective because of defective materials,
components, or workmanship, and the warranty claim is made within the warranty period, EDAN will, at
its discretion, repair or replace the defective part(s) free of charge. EDAN will not provide a substitute
product for use when the defective product is being repaired.
Contact Information
If you have any question about maintenance, technical specifications or malfunctions of devices, contact
your local distributor.
Alternatively, you can send an email to EDAN service department at: support@edan.com.
EDAN INSTRUMENTS, INC.
Address: #15 Jinhui Road, Jinsha Community, Kengzi Sub-District, PingshanDistric,
518122 Shenzhen, P.R. China
Email: info@edan.com.cn
Tel: +86-755-2689 8326
Fax: +86-755-2689 8330
www.edan.com.cn