3.2.5 Care of Batteries
WARNING
1 Replacement or charging of batteries shall be done at least 1.5 meters away from
the patient.
2 Close the battery compartment before using the main unit or the probe.
3 Do not expose the batteries to high temperature, heat them or dispose of them in
fire.
4 Do not disassemble the batteries.
5 Short circuit should be avoided.
CAUTION
1 If the Doppler is not used for an extended period, remove the batteries and keep the
batteries in a cool and dry environment. The environment temperature must not
exceed the range of -20°C ~ +40°C (-4°F ~ +104°F).
2 If the Doppler is not used for an extended period, charge the batteries at least once
a month.
3 The batteries must be properly disposed according to local regulations after their
useful life.
3.3 Connecting the Power Cable
Apply the power cable provided with the Doppler. Put the connector of the power cord to the
power socket on the Doppler (figure 2-4). Put the plug of the power cord into a grounded 3-slot
power output special for hospital usage.
NOTE:
Make sure the AC power supply of the Doppler complies with the following specification:
100 V-240 V~, 50 Hz/60 Hz.
WARNING
If the protective grounding (protective earth) system is doubtful, the Doppler must be
supplied only by inner power.
3.4 Using Wired Probe
3.4.1 Probe Socket
SD5 adopts wired probes. The probe cable is a telephone cable with a standard RJ11 plug on each
terminal.