User's Manual

SE-301 Series Electrocardiograph User Manual Settings
- 35 -
6.7 File Setup
Item Description
Auto Save
Choose from: Off, To ECG or Ext. Memory
Select Off, ECG data will not be saved.
Select To ECG, ECG data in the AUTO or RHYT mode will be saved in
the ECG automatically.
Select Ext. Memory, ECG data in the AUTO or RHYT mode will be
automatically saved to the directory of
ECGDATA\ECG-X\Store\Examination Date of the external memory
after an ECG report is printed out.
NOTE:
1. Please insert the external memory recommended by the
manufacturer. Please set the format to FAT or FAT32 when
formatting the external memory.
2. X in the directory of ECGDATA\ECG-X\Store\Examination
Date can be set in the Device No. textbox in the Transmission
Setup window.
File Format
Choose from: DAT, PDF, SCP, FDA-XML and DICOM
To select SCP/FDA-XML/DICOM, you should first activate the
SCP/FDA-XML/DICOM function on the Advanced Setup screen. For
details, please contact the manufacturer or the local distributor.
Del. After Trans.
Or Export
Choose from: On or Off
Select On, the files will be automatically deleted from the File Manager
screen after they are transmitted to the PC or exported to the external
memory.
Replace When
Memory Full
Choose from: On or Off
Select On, if the stored files reaches 200, the files will replace the earliest
one automatically.
SCP File
Compression
Choose from: On or Off
Select On, the SCP file will be compressed.