User's Manual

SE-301 Series Electrocardiograph User Manual Technical Specifications
- 51 -
Appendix 1 Technical Specifications
A1.1 Safety Specifications
Comply with:
IEC 60601-1:2005/A1:2012
EN 60601-1:2006/A1:2013
IEC 60601-1-2:2007
EN 60601-1-2:2007/AC:2010
IEC 60601-2-25:2011
Anti-electric-shock type: Class І with internal power supply
Anti-electric-shock degree: Type CF
Degree of protection against
harmful ingress of water:
Ordinary equipment (Sealed equipment without liquid
proof)
Disinfection/sterilization
method:
Refer to the user manual for details
Degree of safety of application
in the presence of flammable
gas:
Equipment not suitable for use in the presence of
flammable gas
Working mode: Continuous operation
EMC: CISPR 11 Group 1, Class A
Patient
Leakage
Current:
NC <10μA (AC) / <10μA (DC)
SFC <50μA (AC) / <50μA (DC)
Patient
Auxiliary
Current:
NC <10μA (AC) / <10μA (DC)
SFC <50μA (AC) / <50μA (DC)