SE-301 Series Electrocardiograph User Manual EMC Information
- 55 -
Appendix 2 EMC Information
Electromagnetic emissions
Guidance and manufacture’s declaration – electromagnetic emission
The Electrocardiograph is intended for use in the electromagnetic environment specified
below. The user of the Electrocardiograph should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The Electrocardiograph uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class A
The Electrocardiograph is suitable for use in
all establishments other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies building used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Electromagnetic immunity
Guidance and manufacture’s declaration – electromagnetic immunity
The Electrocardiograph is intended for use in the electromagnetic environment specified
below. The user of Electrocardiograph should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
It is recommended the use
of antistatic materials. If
floor are covered with
synthetic material, the
relative humidity should
be at least 50%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for power
supply lines
It is recommended the use
of filters on power input
lines and enough
separation between signal
lines and power lines.