Product Manual

SE-2003&SE-2012 Holter System Recorder User Manual
- 16 -
3.4 Attaching the Electrodes to the Patient
Use qualified and valid silver-silver chloride disposable monitoring electrodes specially used for
the recorder. It is recommended to choose disposable electrodes complying with YY/T 0196-94
or ANIS/AAMI EC12-1991 requirements.
3.4.1 Electrode Placement
The application of electrodes should be performed by trained medical personnel.
It’s better to place electrodes on skin over ribs or cartilage to avoid interference resulting from
movement of soft tissues.
3.4.1.1 10 Electrodes 12 Leads
10-electrode standard lead wires are utilized by the recorder (12-channel) to create a 12-lead ECG
signal. Different placement positions of 10 electrodes will form different lead modes, and it is
recommended to place electrodes as below:
IEC AHA Electrode Placement
Red R White RA Below clavicle, close to right shoulder.
Yellow L Black LA Below clavicle, close to left shoulder.
Black N Green RL Lower right rib margin over bone.
Green F Red LL Lower left rib margin over bone.
Red C1 Red V1 Right of Mid-Clavicular line 4
th
rib
Yellow C2 Yellow V2 Left of Mid-Clavicular line 4
th
rib