Product Manual

F6 and F6 Express Fetal & Maternal Monitor User Manual Product Specifications
- 105 -
Technique: Peak-peak detection technique
DFHR Measurement
Range:
30bpm ~ 240bpm
Resolution: 1bpm
Accuracy: ±1bpm
Input Impedance: > 10M (Differential, DC50/60Hz)
Input Impedance: > 20M (Common Mode)
CMRR: > 110dB
Noise: < 4µVp
Skin Voltage Tolerance: ±500mV
DECG
Fetal Input Voltage
Current:
20µVp-3mVp
TOCO Range: 0% ~ 100%
Non-linear Error: ±10%
Resolution: 1%
Zero Mode: Automatic/ Manual
TOCO
Dielectric Strength: > 4000Vrms
Pressure Range: 0 ~ 100mmHg
Sensitivity: 5µV/V/mmHg
Non-linear Error: ± 3mmHg
Resolution: 1%
IUP
Zero Mode: Automatic / Manual
Technique: Pulsed Doppler ultrasound
Range: 0% ~ 100%,
AFM
Resolution: 1%
Marking
Manual fetal movement mark
HR Measurement Range: 30 bpm ~ 240 bpm
HR Measuring Accuracy: ±2 bpm
ECG Waveform: Manual control ECG waveform display
ECG falls off: Detect automatically
ECG
Patient Leakage Current
(Limit)
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 10µA 50µA