F6 and F6 Express Fetal & Maternal Monitor User Manual Product Specifications
- 107 -
PR Measuring Accuracy: ±2 bpm
Data update period: 2 s
Wave length
Red light
(660±3) nm
Infrared light
(905±5) nm
SpO2
Emitted light energy Not more than 15 mW
Measurement Range:
Systolic pressure: 40mmHg ~ 270mmHg
Diastolic pressure: 10mmHg ~ 215mmHg
Mean pressure: 20mmHg ~ 235mmHg
Measuring Time: ≤ 120 seconds
Software Over Voltage Limit: ≤ 297mmHg
Hardware Over Voltage Protection:
(325 ± 10) mmHg
Resolution: 1 mmHg
Measuring Accuracy
Max. average deviation:
≤ ±5mmHg
NIBP
(for adult)
Max. standard deviation: ≤ 8mmHg
Channel: 1
Measurement Range: 0 ºC ~ +50 ºC
Accuracy:
(transducer error excluded)
0 ºC ~ +25 ºC: ± 0.2 ºC
+25 ºC ~ +45 ºC: ± 0.1 ºC
+45 ºC ~ +50 ºC: ± 0.2 ºC
Responding Time: ≤ 3min
TEMP
Accessory: TEMP transducer
A1.4 Recorder Specifications
Paper:
Z-fold, thermosensitive
(compatible with GE and PHILIPS recorder paper)
Paper width: 152mm (GE), 150mm (PHILIPS)
Effective printing width:
110mm (American Standard)
120mm (International Standard)