Appendix 3 Troubleshooting
A3.1 No Display
Phenomenon Possible Cause Solution
Power cable is loose. Tighten the power cable.
The fuse is blown. Change the fuse.
Power indicator is off.
The battery runs out of power. Connect to AC power supply.
A3.2 Noise
Phenomenon Possible Cause Solution
Too high volume setup. Turn down the volume.
Turn off or move the
interference source.
Noise
Interfered by mobile phone or
other electromagnetic interference
source.
Move the monitor to a place
with less interference.
A3.3 Recorder Error
Phenomenon Possible Cause Solution
Paper jam
Wrong loading paper or paper is
dampened.
Load paper correctly and keep
paper from moist.
The recorder is not started. Press the PRINT key.
Run out of paper. Load paper.
Recorder does not work.
The paper drawer is not locked.
Slide the paper drawer in until
both latches are locked in
position.