Product Manual
F6 and F6 Express Fetal & Maternal Monitor User Manual Troubleshooting
- 115 -
Power socket has no standard ground
wire.
Check if power socket has
standard ground wire.
The special ground wire connecting with
monitor is not properly earthed.
Check if the special ground
wire connecting with
monitor is earthed.
A3.9 NIBP and SpO2 No Results
Phenomenon Possible Cause Solution
The NIBP cuff is not properly wrapped to
the position of patient's arm.
Check if the NIBP cuff is
properly wrapped to the
position of patient's arm.
The NIBP can not be inflated.
Extend catheter, and check
the connection.
Hose connector plug is not connected well
with the NIBP socket.
Check if the hose connector
plug is connected well with
the NIBP socket.
SpO
2
transducer is not connected well
with the SpO
2
socket.
Check if the SpO
2
transducer is connected
well with the SpO
2
socket.
NIBP and SpO
2
have
no results
Abnormal working condition.
Shut off the power, then
switch it on again.
A3.10 Blown Fuses
WARNING
Switch off the monitor and unplug it before changing the fuse.
Replace the fuse when it is blown.
The two fuses of the monitor are located on the bottom panel, their specifications are:
Size: ะค5mm*20mm; Model: T2AH250V.
To replace a fuse:
1 Fold the LCD display completely flat.
2 Carefully place the monitor upside down on a flat surface covered with cloth or other
protecting pad.
3 With a flat-head screw driver, push the fuse in for about 1 mm and then unscrew it
anticlockwise.
4 Remove the old fuse and replace it with a new fuse that is supplied by the manufacturer or of
the same specifications.