Product Manual
F6 and F6 Express Fetal & Maternal Monitor User Manual EMC Information
- 118 -
Appendix 5 EMC Information – Guidance and
Manufacture’s Declaration
A5.1 Electromagnetic Emissions – for all EQUIPMENT and
SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The F6/F6 Express Fetal & maternal Monitor is intended for use in the electromagnetic environment
specified below. The customer of the user of the F6/F6 Express Fetal & maternal Monitor should
assure that it is used in such and environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The F6/F6 Express Fetal & maternal Monitor
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Class A
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
The F6/F6 Express Fetal & maternal Monitor is
suitable for use in all establishments, other than
domestic and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.