F6 and F6 Express Fetal & Maternal Monitor User Manual EMC Information
- 119 -
A5.2 Electromagnetic Immunity – for all EQUIPMENT and
SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The F6/F6 Express Fetal & maternal Monitor is intended for use in the electromagnetic environment
specified below. The customer or the user of F6/F6 Express Fetal & maternal Monitor should assure
that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment
- guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If floor
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 2kV for power
supply lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line to line
± 2 kV line to ground
± 1 kV line to line
± 2 kV line to ground
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency
(50Hz/60Hz)
magnetic field
IEC61000-4-8
3 A/m 3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.