Product Manual
F6 and F6 Express Fetal & Maternal Monitor User Manual EMC Information
- 121 -
A5.3 Electromagnetic Immunity – for EQUIPMENT and
SYSTEM that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
The F6/F6 Express Fetal & maternal Monitor is intended for use in the electromagnetic environment
specified below. The customer or the user of F6/F6 Express Fetal & maternal Monitor should assure
that it is used in such an environment.
Immunity test IEC 60601 test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
rms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of the F6/F6 Express Fetal &
maternal Monitor including cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
P
V
d
⎥
⎦
⎤
⎢
⎣
⎡
=
1
5.3
P
E
d
⎥
⎦
⎤
⎢
⎣
⎡
=
1
5.3
80 MHz to 800 MHz
P
E
d
⎥
⎦
⎤
⎢
⎣
⎡
=
1
7
800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a)
should be less than the
compliance level in each frequency
range.
b)
Interference may occur in the vicinity of
equipment marked with the following
symbol: