Product Manual
F6 and F6 Express Fetal & Maternal Monitor User Manual Safety Guide
- 7 -
WARNING
47 Use the battery only in the F6 or F6 Express monitor.
48 If the liquid leak from the battery gets into eyes, do not rub the eyes. Wash them well
with clean water and see a doctor immediately.
49 If the liquid leak from the battery spills onto your skin or clothes, wash well with fresh
water immediately.
50 Keep the battery away from fire immediately when leakage or foul odor is detected.
51 Stop using the battery if abnormal heat, odor, discoloration, deformation or abnormal
condition is detected during use, charge, or storage. Keep it away from the monitor.
52 Do not use a battery with serious scar or deformation.
53 Batteries have life cycles. If the time that the monitor using battery becomes much
shorter than usual, the battery life is at an end. Replace the battery with a new one
of the same specification as the one provided or recommended by the manufacturer.
54 Remove the battery and store it at a cool and dry environment if the monitor is not
used for a long time.
55 If the battery is stored alone and not used for a long time, we recommend that the
battery should be charged at least once every 6 months to prevent overdischarge.
CAUTION
1 Refer servicing to qualified personnel.
2 Keep the environment clean. Avoid vibration. Keep it far from corrosive medicine,
dust area, high-temperature and humid environment.
3 When installing the unit into a cabinet, allow for adequate ventilation, accessibility for
servicing, and room for adequate visualization and operation.
4 Do not operate the unit if it is damp or wet because of condensation or spills. Avoid
using the equipment immediately after moving it from a cold environment to a warm,
humid location.
5 Only the sensor and cable of US/TOCO transducers are watertight. Pay attention not
let any liquid enter the transducer plug.
6 Sterility can not be guaranteed if package of the fetal spiral electrode is broken or
opened.
7 The fetal spiral electrode has been sterilized by gamma radiation. Do not re-sterilize.
8 Do not sterilize the monitor or any accessory with autoclave or gas.
9 Switch off the system power before cleaning. Cleaning consists of removing all dust
from the exterior surface of the equipment with a soft brush or cloth.
10 Electromagnetic Interference - Ensure that the environment in which the monitor is
installed is not subject to any source of strong electromagnetic interference, such as
CT, radio transmitters, mobile phone base stations, etc.