Product Manual

F6 and F6 Express Fetal & Maternal Monitor User Manual Maternal Monitoring
- 78 -
9.2.3 Enabling SpO
2
Trace Printing
The real-time SpO
2
measurement result is displayed in the parameter area of the main interface.
You can choose to print them as a continuous trace on the recorder paper (refer to figure 5-1).
To enable or disable SpO
2
trace printing,
1 Select the setup key on the main interface.
2 Select Recorder > SpO
2
Trace.
3 Select ON or OFF (default).
4 Select OK.
9.2.4 Switching the SpO
2
Alarm On or Off
You can choose to switch the SpO
2
alarm on or off.
1 Select the setup key on the main interface.
2 Select Alarm > SpO
2
> Alarm.
3 Select ON (default) or OFF.
4 Select OK.
9.2.5 Changing SpO
2
Alarm Limits
You can change the SpO
2
alarm limits.
1 Select the setup key on the main interface.
2 Select Alarm > SpO
2
.
3 Select a value from 50 ~ 99 for Low Alarm Limit.
4 Select a value from 51 ~ 100 for High Alarm Limit.
5 Select OK.
9.3 Maternal HR Monitoring
9.3.1 Introduction
Maternal heart rate (HR) monitoring does not need an extra accessory. When you perform ECG
or SpO
2
(Pulse) monitoring, the HR result can be acquired at the same time.
When monitoring ECG and SpO
2
at the same time, you can choose the HR source. If only one of
them is being performed, the source will automatically switch to the available one (the screen
reading should prevail).