9.4.6 Correcting the Measurement
To correct the measurement if the limb is not at heart level,
add 0.75 mmHg (0.10 kPa) for each inch higher.
deduct 0.75 mmHg (0.10 kPa) for each inch lower.
9.4.7 Changing NIBP Unit
You can change the NIBP unit.
1 Select the setup key on the main interface.
2 Select Mother > Unit (NIBP Setup).
3 Select mmHg (default) or kPa.
4 Select OK.
9.4.8 Switching the NIBP Alarm On or Off
You can choose to switch the NIBP alarm on or off. The SYS alarm and DIA alarm are related.
Once one of them is switched off, the rest will be switched off as well.
1 Select the setup key on the main interface.
2 Select Alarm > SYS or DIA or MAP.
3 Select ON (default) or OFF.
4 Select OK.
9.4.9 Changing SYS Alarm Limits
You can change the SYS alarm limits.
1 Select the setup key on the main interface.
2 Select Alarm > SYS.
3 Select a value from 30 ~ 269 for Low Alarm Limit.
4 Select a value from 31 ~ 270 for High Alarm Limit.
5 Select OK.
9.4.10 Changing DIA Alarm Limits
You can change the DIA alarm limits.
1 Select the setup key on the main interface.
2 Select Alarm > DIA.
3 Select a value from 10 ~ 244 for Low Alarm Limit.