Product Manual
F6 and F6 Express Fetal & Maternal Monitor User Manual Maternal Monitoring
- 85 -
4 Select a value from 11 ~ 245 for High Alarm Limit.
5 Select OK.
9.4.11 Changing MAP Alarm Limits
You can change the MAP alarm limits.
1 Select the setup key on the main interface.
2 Select Alarm > MAP.
3 Select a value from 20 ~ 254 for Low Alarm Limit.
4 Select a value from 21 ~ 255 for High Alarm Limit.
5 Select OK.
9.4.12 Choosing NIBP Printing Mode
When the recorder is printing real-time fetal traces, the NIBP result is also recorded on the paper
whenever NIBP measurement is performed. After the paper stops advancing, you can choose to
keep recording NIBP results on the paper.
To enable or disable NIBP printing after paper advancing stops,
1 Select the setup key on the main interface.
2 Select Recorder > NIBP.
3 Select ON or OFF (default).
4 Select OK.
NOTE:
The NIBP measurement results during the period of paper lacking and fast printing after
new paper is loaded will not be saved or printed. Therefore, do not perform NIBP
measurements during this period.
9.5 Maternal TEMP Monitoring
9.5.1 TEMP Monitoring Procedure
1) Insert the TEMP plug into the TEMP socket on the monitor.
2) Apply the sensor firmly underneath the patient’s axilla.
WARNING
1 Check if the TEMP transducer functions properly prior to use.
2 Do not apply the TEMP sensor to the mouth or the rectum.