Product Manual

F6 and F6 Express Fetal & Maternal Monitor User Manual Maternal Monitoring
- 86 -
CAUTION
Be cautious when taking and putting the TEMP transducer. Do not pull the cable too
tight or it might cause mechanical damage.
9.5.2 Changing TEMP Unit
You can change the TEMP unit.
1 Select the setup key on the main interface.
2 Select Mother > Unit (TEMP Setup).
3 Select ºC (default) or ºF.
4 Select OK.
9.5.3 Switching the TEMP Alarm On or Off
You can choose to switch the TEMP alarm on or off.
1 Select the setup key on the main interface.
2 Select Alarm > TEMP > Alarm.
3 Select ON (default) or OFF.
4 Select OK.
9.5.4 Changing TEMP Alarm Limits
You can change the TEMP alarm limits.
1 Select the setup key on the main interface.
2 Select Alarm > TEMP.
3 Select a value from 0.0 ~ 49.9 for Low Alarm Limit.
4 Select a value from 0.1 ~ 50.0 for High Alarm Limit.
5 Select OK.