Patient Monitor User Manual Care and Cleaning
- 133 -
below. Never use strong solvents such as acetone or trichloroethylene.
Approved Cleaning Substances
Mild Soaps
Tenside (as active cleaning agent)
Cables can be cleaned with a warm, damp cloth and mild soap, or isopropyl alcohol.
25.2.2.5 Cleaning TEMP/Quick TEMP Sensor
CAUTION
Do not allow a cleaning or disinfecting agent to leave residues on any equipment surface.
After allowing the appropriate time for the agent to work (as indicated by the
manufacturer), wipe off residues with a cloth dampened with water.
Clean with a lint-free cloth, moistened with warm water (40/104) and substances listed
below. Never use strong solvents such as acetone or trichloroethylene.
Approved Cleaning Substances are Hospital-Grade Ethanol.
25.2.2.6 Other Accessories
For cleaning other accessories, please contact the manufactures for details.
25.3 Disinfection
WARNING
Do not mix disinfecting solutions (such as bleach and ammonia), or it may produce
hazardous gases.
To avoid extended damage to the equipment, disinfection is only recommended when stipulated
as necessary in the Hospital Maintenance Schedule. Disinfection facilities should be cleaned first.
Recommended types of disinfecting agents are:
Alcohol
Aldehyde
CAUTION
Do not use EtO gas or formaldehyde to disinfect the monitor.