To export the current monitor’s congifuration:
1. Connect USB device to the monitor’s USB port.
2. Select USB Configure in the Menu > Maintenance > User Maintain, enter the required
password ABC.
3. In the USB Configure menu, select Export to export configuration. A status message will
display after the operation.
To import the configuration on the USB drive to the monitor:
1. Connect the USB drive to the monitor’s USB port.
2. Select USB Configure in the Menu > Maintenance > User Maintain, enter the required
password ABC.
3. In the USB Configure menu, select Import to import configuration. A status message will
display after the operation.
NOTE:
1. The U disk only supports two kinds of format: FAT and FAT 32.
2. The exported configuration files are saved into the folder named USERCONFIG, and
the user can’t modify files the in the folder.
3. Up to three configuration files can be saved and recognized.
7.10 Default Configuration
To set default configuration, please select Menu > Default and choose a configuration (adult,
pediatric or neonate) based on your patient category. This configuration is factory configuration.