M3A Vital Signs Monitor User Manual NIBP Monitoring (Optional)
- 62 -
WARNING
In order to avoid endangering the patient’s life, the user should use this function
cautiously.
Set High for the higher alarm limit, and set Low for the lower alarm limit. If the measured value
is higher than High or lower than Low, the monitor will give an alarm.
The adjusting range of NIBP alarm limits is: 0 mmHg ~ 300 mmHg.
Default NIBP alarm limits:
ADU (mmHg) PED (mmHg) NEO (mmHg)
Lower
Limit
Upper
Limit
Lower
Limit
Upper
Limit
Lower
Limit
Upper
Limit
SYS 90 160 70 120 40 90
DIA 50 90 40 70 20 60
MAP 60 110 50 90 25 70
The adjusting range of NIBP alarm limits:
Adult Mode
SYS 40 mmHg ~ 270 mmHg
DIA 10 mmHg ~ 215 mmHg
MAP 20 mmHg ~ 235 mmHg
Pediatric Mode
SYS 40 mmHg ~ 200 mmHg
DIA 10 mmHg ~ 150 mmHg
MAP 20 mmHg ~ 165 mmHg
Neonatal Mode
SYS 40 mmHg ~ 135 mmHg
DIA 10 mmHg ~ 100 mmHg
MAP 20 mmHg ~ 110 mmHg
When the monitor is configured to NIBP only measuring mode, the adjusting alarm limits of PR
are displayed in the ALARM SETUP menu.