Product Manual
M3B Vital Signs Monitor Service Manual
- 33 -
Record paper goes out
skewwhiffly.
Bad recorder installing or ①
positioning.
Adjust the installation ①
of recorder.
Network linking wire is ①
damaged.
① Check and repair
network linking wire or
HUB.
② Network bed No. conflicts ② Change bed No.
Can not be linked into
network
③
Main control board failure
③
Replace Main control
board
6.4 Power Board Failures
Failure
Possible cause Solution
Fuse is burned upon power-on
① Short-circuit occurs in
power supply or other part.
① Check after power on
Fuse is burned although all loads
are disconnected.
Power failure①
① Replace power supply
Fuse is burned after connecting a
part.
① This part occurs
short-circuit.
① Replace this part
Power indicator lights on,
however, the fan does not run
and the indicator of keyset does
not light.
① +12V DC power supply
has failure.
① Replace the power
Power indicator does not light
on, however, the fan runs
normally and the indicator of
keyset lights on.
① +5V DC power supply
has failure.
① Replace the power
6.5 Parameter Failures
Failure Possible cause Solution
No SpO
2
waveform
Sensor or SpO①
2
module is
damaged
① Replace the sensor and confirm
the failure
① Patient is moving K① eep the patient quiet
SpO
2
waveform has
strong interference.
② Environment light is very
intensive
② Weaken the light intensity in
the environment
SpO
2
value is
inaccurate
① Coloring agent has been
injected into patient body
① Remove the coloring agent
before perform measurement