Product Manual

M3B Vital Signs Monitor Service Manual
- 33 -
Record paper goes out
skewwhiffly.
Bad recorder installing or
positioning.
Adjust the installation
of recorder.
Network linking wire is
damaged.
Check and repair
network linking wire or
HUB.
Network bed No. conflicts Change bed No.
Can not be linked into
network
Main control board failure
Replace Main control
board
6.4 Power Board Failures
Failure
Possible cause Solution
Fuse is burned upon power-on
Short-circuit occurs in
power supply or other part.
Check after power on
Fuse is burned although all loads
are disconnected.
Power failure
Replace power supply
Fuse is burned after connecting a
part.
This part occurs
short-circuit.
Replace this part
Power indicator lights on,
however, the fan does not run
and the indicator of keyset does
not light.
+12V DC power supply
has failure.
Replace the power
Power indicator does not light
on, however, the fan runs
normally and the indicator of
keyset lights on.
+5V DC power supply
has failure.
Replace the power
6.5 Parameter Failures
Failure Possible cause Solution
No SpO
2
waveform
Sensor or SpO
2
module is
damaged
Replace the sensor and confirm
the failure
Patient is moving K eep the patient quiet
SpO
2
waveform has
strong interference.
Environment light is very
intensive
Weaken the light intensity in
the environment
SpO
2
value is
inaccurate
Coloring agent has been
injected into patient body
Remove the coloring agent
before perform measurement