About this Manual P/N: 01.54.455421-10 Release Date: November 2011 © Copyright EDAN INSTRUMENTS, INC. 2011. All rights reserved. Statement This manual will help you understand the operation and maintenance of the product better. It is reminded that the product shall be used strictly complying with this manual. User’s operation failing to comply with this manual may result in malfunction or accident for which EDAN INSTRUMENTS, INC. (hereinafter called EDAN) can not be held liable.
The electrical installation of the relevant room complies with national standards, and The instrument is used in accordance with the instructions for use. Upon request, EDAN may provide, with compensation, necessary circuit diagrams, and other information to help qualified technician to maintain and repair some parts, which EDAN may define as user serviceable. Terms Used in this Manual This guide is designed to give key concepts on safety precautions.
Table of Contents Chapter 1 Safety Guidance ........................................................................................................... 1 1.1 Intended Use ........................................................................................................................... 1 1.2 Warnings and Cautions ........................................................................................................... 1 1.2.1 Safety Warnings .........................................................
4.7 Inspection before Power-On ................................................................................................. 55 Chapter 5 Turning on the Electrocardiograph.......................................................................... 56 Chapter 6 Entering Patient Information ................................................................................... 57 6.1 Entering Patient ID .............................................................................................................
10.6 Editing Patient Information .............................................................................................. 111 10.7 Printing Files..................................................................................................................... 112 10.8 Previewing a File .............................................................................................................. 112 Chapter 11 System Setup..............................................................................
11.6 Lead Setup ........................................................................................................................ 134 11.6.1 Setting Rhythm Lead1/2/3 ......................................................................................... 135 11.6.2 Setting Lead Sequence ............................................................................................... 136 11.6.3 Selecting Nehb ..............................................................................................
11.11.6 Setting Pacemaker Detection Sensitivity ................................................................. 155 Chapter 12 Operation Instructions for Exercise ECG (Configurable for SE-12 Express) . 156 12.1 Activating Exercise ECG Function................................................................................... 156 12.2 Operation Preparation....................................................................................................... 157 12.3 Exercise Test ..........................
SE-12 Series Electrocardiograph User Manual Safety Guidance Chapter 1 Safety Guidance This chapter provides important safety information related to the use of the 12-channel electrocardiograph. 1.1 Intended Use The intended use of the 12-channel electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals.
SE-12 Series Electrocardiograph User Manual Safety Guidance 1.2.1 Safety Warnings WARNING 1. The electrocardiograph is intended to be used by qualified physicians or personnel professionally trained. They should be familiar with the contents of this user manual before operation. 2. Only qualified service engineers can install this equipment, and only service engineers authorized by the manufacturer can open the shell. 3.
SE-12 Series Electrocardiograph User Manual Safety Guidance WARNING 14. The disposable electrodes can only be used for one time. 15. No hazard is found when the electrocardiograph is operated with cardiac pacemakers or other stimulators. 16. Do not touch the patient, bed, table or the equipment while using the ECG together with a defibrillator. 17. Do not touch accessible parts of non-medical electrical equipment and the patient simultaneously. 18.
SE-12 Series Electrocardiograph User Manual Safety Guidance WARNING 26. Multiple portable socket-outlets shall not be placed on the floor. 27. Do not use the additional multiple portable socket-outlet or extension cord in the medical electrical system, unless it’s specified as part of the system by manufacturer. And the multiple portable socket-outlets provided with the system shall only be used for supplying power to equipment which is intended to form part of the system. 28.
SE-12 Series Electrocardiograph User Manual Safety Guidance 1.2.2 Li-ion Battery Care Warnings WARNING 1. Improper operation may cause the internal li-ion battery (hereinafter called battery) to be hot, ignited or exploded, and it may lead to the decrease of the battery capacity. It is necessary to read the user manual carefully and pay more attention to warning messages. 2.
SE-12 Series Electrocardiograph User Manual Safety Guidance CAUTION 2. Do not use the equipment in a dusty environment with bad ventilation or in the presence of corrosive. 3. Make sure that there is no intense electromagnetic interference source around the equipment, such as radio transmitters or mobile phones etc. Attention: large medical electrical equipment such as electrosurgical equipment, radiological equipment and magnetic resonance imaging equipment etc.
SE-12 Series Electrocardiograph User Manual Safety Guidance WARNING 7. Keep the treadmill/ergometer on a horizontal place and make sure it’s stably working. 8. Make sure that the power cable of the treadmill/ergometer is grounded. Don’t use other electrical equipment with the same power supply line. 9. Keep the power cable far away from the running treadmill area and the heat surface. 10. Examine the treadmill/ergometer carefully before using it. 11.
SE-12 Series Electrocardiograph User Manual Safety Guidance 6. Decompensated symptomatic heart failure 7. Acute pulmonary embolus or pulmonary infarction 8. Acute noncardiac disorder that may affect exercise performance or be aggravated by exercise (eg, infection, renal failure, thyrotoxicosis) 9. Acute myocarditis or pericarditis 10. Physical disability that would preclude safe and adequate test performance 11. The patient opposes the test. Relative Contraindications: 1.
SE-12 Series Electrocardiograph User Manual Safety Guidance Potential equalization Mains supply USB socket Battery indicator Battery recharging indicator Delete key Enter key Esc key/Reset key Shift key Fn key Power On/Off key COPY key MODE key SLEEP/WAKE UP key START/STOP key -9-
SE-12 Series Electrocardiograph User Manual Safety Guidance Tab key/Feed paper key UP/DOWN/LEFT/RIGHT Arrow key Recycle Part Number Serial Number Date of Manufacture Manufacturer Authorized Representative in the European Community The symbol indicates that the device complies with the European Council Directive 93/42/EEC concerning medical devices. With respect to electrical shock, fire and mechanical hazards only in accordance with UL 60601-1and CAN/CSA C22.2 No. 601.1 Rx only (U.S.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application Chapter 2 Introduction The 12-channel electrocardiograph gathers ECG signals of 12 leads simultaneously. It displays the operation menu, ECG parameters as well as electrocardiograms. The 12-channel ECG waves can be viewed on the LCD screen and printed out by using a high-quality thermal recorder. The sampled ECG data can be saved, transmitted and exported.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application 2.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application 2.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application Name When the main screen is displayed, press this key to display the main screen of the exercise test. A NOTE: Only if the stress ECG function is activated, can these keys be available. Explanation Pretest When the main screen of the exercise test is displayed, press this key to enter the pretest phase. The length of the pretest phase is not fixed. Exercise Press this key to enter the exercise phase.
SE-12 Series Electrocardiograph User Manual Start/Stop tmill About the Electrocardiograph Application Before the exercise test, press Start tmill to test the connection between the electrocardiograph and the treadmill, and then press Stop tmill to stop the connection test.
SE-12 Series Electrocardiograph User Manual L SLEEP/WAKE UP M COPY N START/STOP About the Electrocardiograph Application Resting/wakening the electrocardiograph In the auto mode, pressing the Copy key can print the ECG report which was printed out last time. Start/Stop printing reports Moving the cursor (Up, Down, Left, Right) In the manual mode, press the Up or Down arrow to switch the lead groups. O Arrow Keys During the exercise test, pressing the Up or Down arrow can switch the display style.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application A C B Figure 2-6 SE-12 Front Panel Symbol Name A B C Explanation supply When the device is powered by the mains supply, this indicator is lit. Mains indicator Battery indicator Battery indicator When the device is powered by battery, this indicator is lit. recharging When the battery is being recharged, this indicator is lit. 2.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application A B C D Figure 2-8 SE-12 Rear Panel Name Explanation A Mains Supply Socket B Potential Equalization Conductor AC SOURCE: alternating current supply socket Potential equalization conductor a connection between the unit and the potential equalization bus bar of the electrical installation. C Handle Part for people to hold D Heat Emission Hole Path for internal heat emission 2.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application D C G A B E F Figure 2-10 SE-12 Right Panel A Name Explanation Patient Cable Socket Connecting to the patient cable Connecting to a PC In the exercise test, only the treadmill/ergometer recommended by the manufacturer can be connected.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application Definitions of corresponding pins: Pin Signal Pin Signal Pin Signal 1 C2 / V2 6 SH 11 F / LL 2 C3 / V3 7 NC 12 3 C4 / V4 8 NC 13 4 C5 / V5 9 R / RA 14 C1 / V1 or NC C1 / V1 RF (N) /RL or NC C6 / V6 5 L / LA 10 RF (N) / RL 15 NOTE: The left side of “/” is European standard, and the right side is American standard.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application Definitions of corresponding pins: Pin Signal Pin Signal 1 TxD (output) 5 NC 2 RxD (input) 6 NC 3 NC 7 GND 4 NC 8 GND 4) External Input/Output Socket Definitions of corresponding pins: Pin Signal Pin Signal 1 GND 4 GND 2 GND 5 ECG Signal (input) 3 GND 6 ECG Signal (output) 5) USB Socket 1/USB Socket 2 (Optional) CAUTION Only the USB equipment recommended by the manufacturer can be con
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application Definitions of corresponding pins: Pin Signal Pin Signal 1 +5V 3 D+ 2 D- 4 GND WARNING 1. Accessory equipment connected to the analog and digital interfaces must be certified according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing equipment and IEC/EN 60601-1 for medical equipment).
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application D C D B E A Figure 2-12 SE-12 Bottom Panel Name Explanation A Speaker Hole Path for sound from speaker B Battery Compartment Compartment for battery C Fuse The specification is: T1AL250VP Ø5×20 D Heat Emission Hole Path for internal heat emission E Label Position for product information label 1) Battery Compartment The rated voltage and the rated capacity of the battery pack are as follows: SE-12 Expr
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application WARNING 2. When leakage or foul smell is found, stop using the battery immediately. If your skin or cloth comes into contact with the leakage liquid, cleanse it with clean water at once. If the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with clean water first and go to see a doctor immediately. 3.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application ♦ Flexible printing formats ♦ The auto, manual, rhythm, R-R analysis and off modes can be chosen freely ♦ Automatic baseline adjustment for optimal printing ♦ Convenient operation of system setup and file management ♦ Multiple file formats: DAT/SCP/FDA-XML/PDF ♦ Measurement function and interpretation function ♦ Hint information of lead off, lack of paper, low battery capacity etc.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application Chapter 3 About the Electrocardiograph Application The following sections provide an overview of the main operations and functions. You can operate the electrocardiograph by using the keyboard. For SE-12 Express, you can also carry out the operations by using the touch screen. CAUTION Do not touch the LCD screen with sharp things such as pencils or pens; otherwise, it will be damaged. 3.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application 3.2 Entering Data 1. When the main screen is displayed, press the function key F1 below Patient to open the Patient Information window. 2. Press Tab or Shift + Tab to move the cursor to the Name textbox. 3. Press on the keyboard to erase the typed information. 4. Press the letter or number keys on the keyboard to input patient name.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application 3.3 Selecting an Item When the main screen is displayed, press the function key F6 below Setup to open the System Setup window. The cursor is on Work Mode. Press to open the Work Mode Setup window. 1. Press Tab or Shift + Tab to move the cursor to R-R, press Spacebar to select it, and then a check mark √ appears in the box before R-R. 2. Press Tab or Shift + Tab to move the cursor to the Rhythm Style list box.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application 3.4 About the main screen After you turn on the device, the main screen pops up.
SE-12 Series Electrocardiograph User Manual A About the Electrocardiograph Application Name Explanation Name Patient Name: within 60 ASCII characters or the equivalent number of other characters that can be supported by equivalent memory used by 60 ASCII characters When ID Mode is set to Manual, the patient ID is within 30 ASCII characters. B ID When ID Mode is set to Auto, the patient ID is 0~1999, 999, 999. When ID Mode is set to Time, the patient ID is within 10 characters.
SE-12 Series Electrocardiograph User Manual N O Gain Speed About the Electrocardiograph Application Gain: 2.5 mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV, 10/5 mm/mV or AGC NOTE: This setup can’t be saved if it is modified on the main screen. In the manual mode, you can set Speed to 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s or 50mm/s. In the auto and rhythm modes, only 25mm/s and 50mm/s are available. In the R-R analysis mode, only 25mm/s is available.
SE-12 Series Electrocardiograph User Manual E D About the Electrocardiograph Application F G I H C J K L B M A U T S R Q P O N Figure 3-3 Main screen of Exercise Test X V Figure 3-4 Main screen of ST Trend - 32 - W
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application Z Y Figure 3-5 Main screen of Template When ID Mode is set to Manual, the patient ID is within 30 characters. A ID When ID Mode is set to Auto, the patient ID is 0~1999, 999, 999. When ID Mode is set to Time, the patient ID is within 10 characters.
SE-12 Series Electrocardiograph User Manual H Total Time and Stage Time About the Electrocardiograph Application The left number is the total time, which is counted from the beginning of the pretest phase to the end of the exercise test. The right number is the stage time, which shows the running time of the current stage.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application Please select a proper option based on the patient’s actual ECG waves. S Hold Press the function key F2 below Hold, and then the system will not follow the previous settings to change the speed and grade of the treadmill, but keep the current speed and grade until this key is pressed again. T Patient Press the function key F1 below Patient to open the Patient Information window. It is invalid in the exercise test.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on an item, press to open the setup window of the item. Take the Work Mode Setup window for example: Check Box Textbox Dropdown List Box 1. In the Work Mode Setup window, press Tab or Shift + Tab to move the cursor among different check boxes.
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application 2. In the Work Mode Setup window, press Tab or Shift + Tab to move the cursor to a list box or dropdown. Press the Up or Down arrow to highlight an option. 3. In the Work Mode Setup window, press Tab or Shift + Tab to move the cursor to a textbox. Fill in the selected textbox. 4. Press and then press to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, to confirm. 5.
SE-12 Series Electrocardiograph User Manual A B C About the Electrocardiograph Application D E F G Figure 3-6 File Manager screen1 H I J K L M Figure 3-7 File Manager screen2 - 38 - N
SE-12 Series Electrocardiograph User Manual About the Electrocardiograph Application Name Explanation A Trans All Press the function key F1 below Trans All to transmit all the files to the PC. B Export All Press the function key F2 below Export All to export all the files from the electrocardiograph to the U disk. C Del All Press the function key F3 below Del All to delete all the files from the electrocardiograph.
SE-12 Series Electrocardiograph User Manual Operation Preparations Chapter 4 Operation Preparations WARNING Before use, the equipment, patient cable and electrodes should be checked. Replace them if there is any evident defectiveness or aging which may impair the safety or the performance, and make sure that the equipment is in proper working condition. 4.1 Power and Earthing WARNING If the integrity of the external protective conductor is in doubt, the equipment should be powered by battery.
SE-12 Series Electrocardiograph User Manual Operation Preparations ♦ Battery When the battery is used, turn on the unit by pressing battery indicator ( on the keyboard. Then the ) will be lit and the battery symbol will be displayed on the LCD screen. Because of the consumption during the storage and transport course, the battery capacity may not be full. If the symbol and the hint information Battery Weak are displayed, which means the battery capacity is low, please recharge the battery first.
SE-12 Series Electrocardiograph User Manual Operation Preparations 4.2 Loading/Replacing Recorder Paper Two kinds of recorder paper can be used. One is the folded thermal paper, and the other is the rolled thermal paper. NOTE: 1. When the folded thermal paper is used, the paper roller is unnecessary and must be taken out. 2. When using the paper of 216mm in width, the two movable parts should be removed. 3. The exit edge can help you tear the recorder paper.
SE-12 Series Electrocardiograph User Manual Operation Preparations Loading/Replacing Process of Folded Paper: 1) Press the casing button downwards and remove the casing to open the recorder. 2) Remove the remainder paper from the paper tray if necessary. 3) Take off the wrapper of the new folded paper, and then put it in the paper tray. 4) Pull the paper out with the grid side facing the thermal print head, and replace the casing on the recorder.
SE-12 Series Electrocardiograph User Manual Operation Preparations 5) Press down the recorder casing firmly. 6) To select Paper Marker. Press on the keyboard to turn on the unit. After the main screen is displayed, press the function key F6 below Setup to display the System Setup screen. Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Record Info on the System Setup screen, press to open the Record Info Setup window.
SE-12 Series Electrocardiograph User Manual Operation Preparations 4) Place the paper and the roller gently in the recorder with the roller pin clicking into the groove. Groove 5) Pull the paper out with the grid side facing the thermal print head, and replace the casing on the recorder. 6) Press down the recorder casing firmly. 7) To select Paper Marker. Press on the keyboard to turn on the unit.
SE-12 Series Electrocardiograph User Manual Operation Preparations 8) Advance the recorder paper. When the main screen is displayed, you can press 2.5cm. Press to advance the paper for again to stop advancing the paper. 4.3 Preparing the Patient 4.3.1 Instructing the Patient Before attaching the electrodes, greet the patient and explain the procedure. Explaining the procedure decreases the patient’s anxiety. Reassure the patient that the procedure is painless. Privacy is important for relaxation.
SE-12 Series Electrocardiograph User Manual Operation Preparations 4.4 Connecting the Patient Electrocardiograph and Electrodes Cable to the WARNING The performance and electric shock protection can be guaranteed only if the original patient cable and electrodes of the manufacturer are used. The patient cable includes the main cable and lead wires which can be connected to electrodes.
SE-12 Series Electrocardiograph User Manual Operation Preparations 1. Connecting the Patient Cable to the Electrocardiograph Connect the patient cable to the patient cable socket on the right side of the main unit, and then secure them with two screws. 2. Connecting the Patient Cable to Electrodes Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to the reusable electrodes or the alligator clips. Firmly attach them. 4.
SE-12 Series Electrocardiograph User Manual Operation Preparations Chest 5 C5 White/Black V5 Brown/Orange Chest 6 C6 White/Violet V6 Brown/Violet As the following figure shows, the positions of chest electrodes on the body surface are C1: Fourth intercostal space at the right border of the sternum C2: Fourth intercostal space at the left border of the sternum C3: Fifth rib between C2 and C4 C4: Fifth intercostal space on the left midclavicular line C5: Left anterior axillary line at the horizont
SE-12 Series Electrocardiograph User Manual Operation Preparations 4.5.1.2 NEHB Placement The electrodes’ positions on the body surface are shown in the following table and figure.
SE-12 Series Electrocardiograph User Manual Operation Preparations Connecting to a Lead Wire Reed Clamp Limb Electrode Chest Electrode Connection: 1) Ensure that the electrodes are clean; 2) Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to the corresponding electrodes according to the colors and identifiers; 3) Clean the electrode area on the chest surface with 75% alcohol; 4) Daub the round area of 25mm in diameter on each electrode site with gel evenly;
SE-12 Series Electrocardiograph User Manual Operation Preparations 4.5.2.2 Disposable Electrodes CAUTION The disposable electrodes can only be used for one time. Disposable Electrode Alligator Clip Disposable electrodes must be used together with alligator clips. Disposable Electrode Connection 1) Align all lead wires of the patient cable to avoid twisting, and connect the alligator clips to the lead wires. 2) Clean the electrode areas on the body surface with 75% alcohol.
SE-12 Series Electrocardiograph User Manual Operation Preparations 4.6 Attaching Electrodes to the Patient (for Exercise ECG) The identifiers and color codes of electrodes used comply with IEC/EN requirements. In order to avoid incorrect connections, the electrode identifiers and color codes are specified in Table 4-2. Moreover the equivalent codes according to American requirements are given in Table 4-2 too.
SE-12 Series Electrocardiograph User Manual Operation Preparations The Precordial Electrodes’ Positions on Body Surface: C1: Fourth intercostal space at the right border of the sternum C2: Fourth intercostal space at the left border of the sternum C3: Fifth rib between C2 and C4 C4: Fifth intercostal space on the left midclavicular line C5: Left anterior axillary line at the horizontal level of C4 C6: Left midaxillary line at the horizontal level of C4 The Extremity Electrodes’ Positions on Body Surface:
SE-12 Series Electrocardiograph User Manual Operation Preparations 4.7 Inspection before Power-On In order to avoid safety hazards and get good ECG records, the following inspection procedure is recommended before power-on and operation. 1) Environment: ♦ Make sure that there is no electromagnetic interference source around the equipment, especially large medical electrical equipment such as electrosurgical equipment, radiological equipment, magnetic resonance imaging equipment etc.
SE-12 Series Electrocardiograph User Manual Turning on the Electrocardiograph Chapter 5 Turning on the Electrocardiograph ♦ When the mains supply is used, connect the power cord, and the mains supply indicator ( ) is lit. Then press on the keyboard to turn on the unit. The logo will be displayed on the LCD screen after self-test. Then the electrocardiograph is ready for use. ♦ When battery is used, press battery indicator ( on the keyboard directly to turn on the unit, and then the ) is lit.
SE-12 Series Electrocardiograph User Manual Entering Patient Information Chapter 6 Entering Patient Information After the electrocardiograph is turned on, the main screen appears. For SE-12 Express, pressing the Pretest key can open the main screen of the exercise test. 6.1 Entering Patient ID 1. By default, the system generates the patient ID automatically. The range of the patient ID is 0 ~ 1999, 999, 999. 2.
SE-12 Series Electrocardiograph User Manual Entering Patient Information 3) Press Tab or Shift + Tab to move the cursor to ID Mode, and then press the Up or Down arrow to select Manual. 4) Press , or press Tab or Shift + Tab to move the cursor to the OK button, and then press . 5) Press ESC to return to the main screen. 6) Press the function key F1 below Patient to open the Patient Information window. Press Tab or Shift + Tab to move the cursor.
SE-12 Series Electrocardiograph User Manual Entering Patient Information First Name Within 30 ASCII characters Last Name Within 30 ASCII characters Age Age Unit: Years, Months, Weeks or Days Gender Patient Gender (Male/Female/Empty) Weight (kg) Patient Weight (Range 0~999) Height (cm) Patient Height (Range 0~999) Pacemaker If you select Pacemaker in the Patient Information Setup window, Pacemaker appears in the Patient Information window.
SE-12 Series Electrocardiograph User Manual Entering Patient Information 3. After inputting all the information, press to confirm; or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. NOTE: If you select D.O.B in the Patient Information Setup window, the D.O.B textbox appears and the Age textbox becomes unavailable in the Patient Information window, you can enter the birthday of the patient, and the system will calculate the patient age automatically. 6.
SE-12 Series Electrocardiograph User Manual Entering Patient Information 3) Enter the start and end addresses, the male and female codes, the vendor ID and the product ID, and then press to confirm. View the vendor ID and product ID on the PC: a) Connect the bar code reader to the PC b) Click on start, right-click on My Computer, and then select Manage in the pop-up menu to open the Computer Management window.
SE-12 Series Electrocardiograph User Manual Entering Patient Information c) In the Computer Management window, click on Device Manager. Right-click on HID Keyboard Device, and then select Properties in the pop-up menu. d) Click on the Details tab in the HID Keyboard Device Properties window. View the vector ID (VID) and the product ID (PID). Take the following figure for example, the vector ID is 04B4, and the product ID is 0100.
SE-12 Series Electrocardiograph User Manual Entering Patient Information 4) Press ESC to return to the main screen. 2. Connect the bar code reader to USB socket 2 on the right panel of the electrocardiograph. 3. When the main screen is displayed, scan the patient’s bar code with the bar code reader, and then the patient information will appear in the corresponding box. NOTE: Only bar code readers recommended by the manufacturer can be used.
SE-12 Series Electrocardiograph User Manual Freezing ECG Waves Chapter 7 Freezing ECG Waves You can freeze the ECG waves displayed on the main screen. Operation Method: 1) Press the function key F2 below Freeze to display the freezing screen. NOTE: Within ten seconds after returning to the main screen, pressing the function key F2 can not display the freezing screen.
SE-12 Series Electrocardiograph User Manual Freezing ECG Waves Figure 7-2 Replay screen 4) Press F1 below Pre Page or F4 below Next Page to turn pages. 5) Press F2 below Pre Sec or F3 below Next Sec to view the ECG waves of the previous or next second. 6) Press the function key F5 below Play to display ECG waves continuously. 7) Press the function key F6 below Pause to stop playing ECG waves. 8) Press the function key F7 below Return to return to the main screen.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports Chapter 8 Printing ECG Reports There are four modes to print ECG reports. In the auto mode, the lead groups are switched automatically according to the lead sequence during the printing course. After the ECG waves of one lead group are printed within a certain time, the system switches to print ECG waves of another lead group automatically. A 1mV calibration mark will be printed at the beginning of an ECG report.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports 3) Press Tab or Shift + Tab to move the cursor to the Sampling Mode list box, and then press the Up or Down arrow to highlight a sampling mode. For details of the sampling mode, refer to Section 11.1.5, “Selecting Sampling Mode”. 4) Press Tab or Shift + Tab to move the cursor to Preview, and then press the Up or Down arrow to highlight Off or On. 5) Press then press , or press Tab or Shift + Tab to move the cursor to the OK button, and .
SE-12 Series Electrocardiograph User Manual 8) Press Printing ECG Reports , or press Tab or Shift + Tab to move the cursor to the OK button, and then press . 9) On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Lead, press to open the Lead Setup window. 10) Press Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3 list box, and then press the Up or Down arrow to select a lead.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports After setup, press ESC to return to the main screen. 2. When the main screen is displayed, press the MODE key to select the auto mode. Press F3 to select a paper speed. Press F4 to switch the gain. Press F5 to set the EMG filter or the Lowpass filter. 3. Press the START/STOP key to print an ECG report. It will stop automatically after printing a complete ECG report of 12 leads. Or press the START/STOP key again to stop printing the report.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports 1) Press the function key F1 below Waveform to display the Waveform Preview Screen. 2) Press the function key F2 below Template to display the template preview screen. 3) Press the function key F3 below Measure to display the measure preview screen. 4) Press the function key F4 below Diagnosis to open the diagnosis preview screen. 5) Press the function key F5 below Continue to print an ECG report.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports 4) On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Lead, press to open the Lead Setup window. 5) Press Tab or Shift + Tab to move the cursor to the Lead Sequence list box, and then press the Up or Down arrow to select a sequence. 6) Press then press , or press Tab or Shift + Tab to move the cursor to the OK button, and .
SE-12 Series Electrocardiograph User Manual Printing ECG Reports 2. When the main screen is displayed, press the MODE key to select the manual mode. Press F3 to select a paper speed. Press F4 to switch the gain. Press F5 to set the EMG filter or the Lowpass filter. 3. Press the Up or Down arrow to select the lead group to be displayed and printed. 4. Press the START/STOP key to print an ECG report. 5. Press the START/STOP key to stop printing the ECG report. 8.3 Rhythm Mode Operation Method: 1.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports 4) On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Lead, press to open the Lead Setup window. 5) Press Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3 list box, and then press the Up or Down arrow to select a lead.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports After setup, press ESC to return to the main screen. 2. When the main screen is displayed, press the MODE key to select the rhythm mode. Press F3 to select a paper speed. Press F4 to switch the gain. Press F5 to set the EMG filter or the Lowpass filter. 3. Press the START/STOP key to begin to count the sampling time. When the sampling time reaches 60s in the Single Lead style or 20s in the Three Leads style, it begins to print an ECG report.
SE-12 Series Electrocardiograph User Manual 2) Press then press Printing ECG Reports , or press Tab or Shift + Tab to move the cursor to the OK button, and . 3) On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Lead, press to open the Lead Setup window. 4) Press Tab or Shift + Tab to move the cursor to the Rhythm Lead1 list box, and then press the Up or Down arrow to select a lead.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports After setup, press ESC to return to the main screen. 2. When the main screen is displayed, press the MODE key to select the R-R analysis mode. Press F4 to switch the gain. Press F5 to set the EMG filter or the Lowpass filter. 3. Press the START/STOP key to begin to count the sampling time. When the sampling time reaches 180s, it begins to print an ECG report. 4.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports 8.6 ECG Reports 8.6.1 ECG Reports in the Auto Mode 8.6.1.1 Example of 6×2+1R Save Paper Mode The above figure shows an ECG report in the auto mode. The style is 6×2+1R. The ECG report includes: Patient Information, Measure Information, Diagnosis Information, Report Confirmed by Current Date and Current Time 6×2+1R ECG waves 0.67~100Hz (0.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports 2*5s+1r (12 leads are printed in 2 groups of 6, and every group is printed for about 5s, with one rhythm lead at the bottom simultaneously) Institution Name SE-12 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports (b) The above figure (a) and (b) show an ECG report in the auto mode. Template is selected, and Record Style is set to 6×2+1R. Figure (a) shows: 02-06-2010 16:52:40 (Current Date & Current Time) ID: 001 (Patient ID) Name: Marry, Johnson (Patient Name) Age: 25 Years (Patient Age) (1mV calibration mark) І, II, Ш, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 (12 standard leads) and ECG waves II on the bottom (Rhythm Lead) and rhythm wave 0.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports 2*5s+1r (12 leads are printed in 2 groups of 6, and every group is printed for about 5s, with one rhythm lead at the bottom simultaneously) Institution Name SE-12 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.5(Algorithm Version Number) Figure (b) shows: Patient Information, Measure Information, Diagnosis Information, Average Template, Report Confirmed by, Current Date and Current Time, 0.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports Average Template: Average template shows the average value of 10s sampled ECG signals of every lead. The broken lines on the template are position markers. They respectively mark the start and end points of the P and QRS waves, and the end point of the T wave. 8.6.1.2 Example of 3×4+1R The above figure shows an ECG report in the auto mode. The style is 3×4+1R.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports Institution Name SE-12 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.5(Algorithm Version Number) NOTE: Please refer to Section 8.6.1.1 for the explanation of the patient information, measure information and diagnosis information. 8.6.2 ECG Reports in the Rhythm Mode The above figure shows an ECG report in the rhythm mode, and Rhythm Style is set to Three Leads.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports 60 (Heart Rate) (1mV calibration mark) II, V1, V5 (Rhythm Lead Name) 20s rhythm waveform of lead II / V1 / V5 0.67~100Hz (0.67Hz DFT Filter, 100Hz Lowpass Filter) AC50 (50Hz AC Filter) 25mm/s (Paper Speed) 10mm/mV (Gain) Institution Name SE-12 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.5(Algorithm Version Number) 8.6.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports (1mV calibration mark) І, II, Ш, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 (12 standard leads) ECG waveform of 12 standard leads 0.67~100Hz (0.67Hz DFT Filter, 100Hz Lowpass Filter) AC50 (50Hz AC Filter) 25mm/s (Paper Speed) 10mm/mV (Gain) 60 (Heart Rate) Institution Name SE-12 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.5(Algorithm Version Number) 8.6.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports (b) - 85 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports (c) The above figure (a), (b) and (c) show an ECG report in the R-R analysis mode.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports SDNN (Standard Deviation of Normal to Normal Intervals) RMSSD (Root Mean Square Successive Difference) RR Histogram RR Trend Figure (b) shows: Current Date & Current Time Patient Information (Name, ID, Age) Gain, Speed, Filter (1mV calibration mark) II (Lead name) 180s compressed ECG wave of Lead II Figure (c) shows: All the RR interval values within the measuring time - 87 -
SE-12 Series Electrocardiograph User Manual Printing ECG Reports 8.6.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports As the above figure shows, the ECG report printed by the USB printer includes: Institution Name Date and Time Patient Information Measurement Information Diagnosis Information; Report Confirmed by; ECG waves of 12 leads DFT Filter, Lowpass Filter, AC Filter, Paper Speed, Gain; SE-12 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.5(Algorithm Version Number) 8.6.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports NOTE: 1. To the patient with a pacemaker, the EMG filter should be set to Off and the Lowpass filter should be set to 150Hz; or else, the amplitude of the pulses may decrease, the width may increase, and the diagnosis results may not include the pacemaker information. 2. To the patient with a pacemaker, the results given by the equipment may be invalid. 8.6.
SE-12 Series Electrocardiograph User Manual Printing ECG Reports 0.67~100Hz (0.67Hz DFT Filter, 100Hz Lowpass Filter) AC50 (50Hz AC Filter) 10mm/mV (Gain) 25mm/s (Paper Speed) 60 (Heart Rate) Institution Name SE-12 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.
SE-12 Series Electrocardiograph User Manual Chapter 9 Transmitting ECG Data Automatically Transmitting ECG Data Automatically 9.1 Transmitting ECG Data to the PC NOTE: 1. To transmit ECG data in DAT format to the PC, the Smart ECG Viewer software of the manufacturer must be installed in the PC. You should log into the Smart ECG Viewer software before transmission. 2. To transmit ECG data in SCP/FDA-XML/PDF format to the PC, the FTP receiving software must be installed in the PC.
SE-12 Series Electrocardiograph User Manual Transmitting ECG Data Automatically Figure 9-1 RS232-USB Assembly 3. Set Auto Transmission to On and Transmission Mode to UART in the Transmission Setup window. 1) When the main screen is displayed, press the function key F6 below Setup to open the System Setup screen. 2) Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor to open the Transmission Setup window.
SE-12 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 5. In the auto or rhythm mode, ECG data will be transmitted through the serial cable automatically after an ECG report is printed out. In the off mode, ECG data will be transmitted through the serial cable automatically after the START/STOP key is pressed. 9.1.2 Transmitting ECG Data in DAT Format through Net Cable 1.
SE-12 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 6) Press Tab or Shift + Tab to move the cursor to the Local IP item. For the cross-network transmission, set the first two sections of the Local IP item to the first two sections of the IP of the PC. Set the third section of the Local IP item to the network segment of the electrocardiograph which depends on the configuration of Router. The last section of the Local IP item can be set at random.
SE-12 Series Electrocardiograph User Manual Transmitting ECG Data Automatically In the Transmission Setup window of the electrocardiograph, set Transmission Mode to Ethernet. Set the Server IP item to the IP of the PC. Set the first three sections of the Local IP item to the first three sections of the IP of the PC. The last section of the Local IP item is not the same as the last section of the IP of the PC.
SE-12 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 9.1.3 Transmitting ECG Data in SCP/FDA-XML/PDF Format through Net Cable 1. Activate SCP/FDA-XML/PDF function on the Advanced Setup screen 1) When the main screen is displayed, press the function key F6 below Setup to open the System Setup screen. 2) Press F1 to display the Service Password window.
SE-12 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 3) Press Tab or Shift + Tab to move the cursor to the Activate button after the function to be activated. Press password and press to open the Activate Password window. Enter the to activate the function. NOTE: Please contact local distributors to get the service password and the activating password. 2.
SE-12 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 3) Log into the FTP receiving software. The FTP port is 21. 4) In the auto or rhythm mode, ECG data will be transmitted through the net cable automatically after an ECG report is printed out. In the off mode, ECG data will be transmitted through the net cable automatically after the START/STOP key is pressed. 9.2 Exporting ECG Data to the U Disk Automatically 1.
SE-12 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 4. In the auto or rhythm mode, ECG data will be automatically exported to the directory of ECGDATA\ECG-X\Store\Examination Date of the U disk after an ECG report is printed out. In the off mode, ECG data will be automatically exported to the directory of ECGDATA\ECG-X\Store\Examination Date of the U disk after the START/STOP key is pressed. NOTE: X can be set in the Device No. textbox in the Transmission Setup window. 9.
SE-12 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 2. Connect USB socket 1 of the electrocardiograph to the USB socket of the PC by using the high-speed USB cable 3. Start the real-time transmission The electrocardiograph transmits to the PC the ECG signals acquired from the patient. Acquisition and transmission are simultaneous. The ECG signals are displayed on the PC monitor and eventually analyzed. For more details, refer to the user manual of the PC ECG software.
SE-12 Series Electrocardiograph User Manual Managing Files Chapter 10 Managing Files If you want to save the ECG data in the electrocardiograph, you should set Auto Save to On and Saving Path to ECG. By default, Auto Save is On and Saving Path is ECG. Then the ECG data in the auto, off or rhythm mode will be saved on the File Manager screen automatically. To Set Auto Save to On and Saving Path to ECG 1.
SE-12 Series Electrocardiograph User Manual Managing Files Figure 10-1 File Manager screen1 Figure 10-2 File Manager screen2 On the File Manager screen, files can be printed, transmitted, exported, displayed, edited searched or deleted. SE-12 can accommodate 100 files, and SE-12 Express can accommodate 200 files. If there is no file on the File Manager screen, the following dialog box will pop up when you press function keys.
SE-12 Series Electrocardiograph User Manual Managing Files CAUTION 1. When files are being printed, transmitted, deleted or exported, you can not turn off the electrocardiograph. 2. Do not cut off the mains supply directly when no battery is installed in the device, or else, the stored data may be lost. 10.1 Transmitting Files to the PC NOTE: 1. To transmit DAT files to the PC, the Smart ECG Viewer software of the manufacturer must be installed in the PC.
SE-12 Series Electrocardiograph User Manual Managing Files 10.1.1 Transmitting DAT Files through Serial Cable 1. Connect the RS232 socket of the PC to the RS232 socket of the electrocardiograph with an RS232 cable. 2. Or if the PC has no RS232 socket, connect the USB socket of the PC to the RS232 socket of the electrocardiograph by using the RS232-USB assembly. For details of the RS232-USB assembly, please refer to Section 9.1.1, “Transmitting ECG Data in DAT Format Through Serial Cable”. 3.
SE-12 Series Electrocardiograph User Manual Managing Files 10.1.2 Transmitting DAT Files through Net Cable 1. Connect the electrocardiograph to the network of the PC with an Ethernet cable recommended by the manufacturer. If the wireless AP transmission is used, connect the PC to the server wireless AP, and connect the electrocardiograph to the client wireless AP. Only the wireless AP recommended by the manufacturer can be used. NOTE: 1.
SE-12 Series Electrocardiograph User Manual Managing Files 6. Or press the Up or Down arrow to highlight a file on the File Manager screen1, and then press the function key F4 below Select to select the file and display the File Manager screen2. Press the function key F3 below Trans to transmit the selected file to the PC. 10.1.3 Transmitting SCP/FDA-XML/PDF Files through Net Cable 1. Activate SCP/FDA-XML/PDF function on the Advanced Setup screen For details, refer to Section 9.1.
SE-12 Series Electrocardiograph User Manual Managing Files 2) Log into the FTP receiving software. The FTP port is 21. 3) Input the correct FTP path, FTP user name and password. Set Transmission Mode to Ethernet. Set the IP addresses. For details, please see Section 9.1.3, “Transmitting ECG Data in SCP/FDA-XML/PDF Format to the PC Through Net Cable”. 4) Press the function key F7 below File to display the File Manager screen1.
SE-12 Series Electrocardiograph User Manual Managing Files If the U disk is not connected well, the hint U disk is not ready! will be displayed on the LCD screen of the electrocardiograph, and then you should connect the U disk again. If the electrocardiograph fails to export files, the hint Exporting fails! will pop up. After the electrocardiograph successfully exports files, the hint Exporting succeeds! will be displayed. NOTE: 1.
SE-12 Series Electrocardiograph User Manual Managing Files 10.4 Searching Files To Open the SearchInfo Setup Window 1. When the main screen is displayed, press the function key F7 below File to open the File Manager screen1. 2. Press the function key F5 below Search to open the SearchInfo Setup window. To Search Files 1. Press Tab, Shift + Tab, or the Up or Down arrow to select Name, and then input the patient name. Press , or press Tab or Shift + Tab to move the cursor to the OK button, and .
SE-12 Series Electrocardiograph User Manual Managing Files 3. Press Tab, Shift + Tab, or the Up or Down arrow to select Time, and then input the examination time range. Press , or press Tab or Shift + Tab to move the cursor to the OK button, and then press . Then all the corresponding files from the start time to the end time will be displayed on the File Manager screen1. NOTE: The time mode is year-month-day in the SearchInfo Setup window. 10.5 Importing Files 1.
SE-12 Series Electrocardiograph User Manual Managing Files NOTE: The Address item is customized. 1. Press Tab or Shift + Tab to move the cursor to a textbox. Fill in the selected textbox. 2. Press Tab or Shift + Tab to move the cursor to a list box. Press the Up or Down arrow to select an option. 3. After inputting all the information, press to confirm; or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. For details on inputting data, please refer to Section 3.
SE-12 Series Electrocardiograph User Manual Managing Files 3. Press the function key F6 below Preview to open the file preview screen. To Preview the File 1. Press the function key F1 below Waveform to display the waveform screen. 2. Press the function key F2 below Template to display the template screen. 3.
SE-12 Series Electrocardiograph User Manual Managing Files 4. Press the function key F4 below Diagnosis to show Minnesota Code and Diagnosis Information. 5. Press START/STOP or the function key F5 below Record to print the file. Pressing START/STOP or the function key F5 below Stop can stop printing the file. 6. Press the function key F7 below Return to return to the File Manager screen2. NOTE: Rhythm data can not be previewed.
SE-12 Series Electrocardiograph User Manual System Setup Chapter 11 System Setup After you turn on the device, the main screen pops up. Press the function key F6 below Setup to display the System Setup screen. Figure 11-1 System Setup screen After setup, 1. Press ESC to return to the main screen. 2. Or press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor to Return, and then press to return to the main screen. 11.
SE-12 Series Electrocardiograph User Manual System Setup Figure 11-2 Work Mode Setup window After setup, press OK button, and then press to confirm, or press Tab or Shift + Tab to move the cursor to the to confirm. Then the System Setup screen appears. 11.1.1 Selecting Work Mode Press Tab or Shift + Tab to move the cursor to a work mode, press Spacebar to select it, and then a check mark √ appears in the box before the work mode.
SE-12 Series Electrocardiograph User Manual System Setup R-R: In the R-R analysis mode, you can select a lead to print its R-R histogram, R-R trend chart, 180s compressed ECG waveform and all the R-R interval values. OFF: In the off mode, the lead groups are switched automatically according to the lead sequence. When the main screen is displayed, after pressing the START/STOP key, the sampled ECG data can be saved and transmitted, but can not be printed. 11.1.
SE-12 Series Electrocardiograph User Manual System Setup When the auto display style is set to 3×4+1R, ECG waves of 12 leads are displayed in 4 groups of 3, with the ECG wave of one rhythm lead on the bottom. When the auto display style is set to 3×4+3R, ECG waves of 12 leads are displayed in 4 groups of 3, with ECG waves of three rhythm leads on the bottom. When the auto display style is set to 6×2, ECG waves of 12 leads are displayed in 2 groups of 6.
SE-12 Series Electrocardiograph User Manual System Setup 11.1.6 Setting Interval and Duration Press Tab or Shift + Tab to move the cursor to the Duration and Interval textboxes in the Work Mode Setup window, and then input the duration and the interval manually. In the auto mode, when Sampling Mode is set to Periodic Sample, if Interval is set to 2 min, Duration is set to 24 min, after pressing the START/STOP key, the printing will be performed every two minutes and come to 12 times. 11.1.
SE-12 Series Electrocardiograph User Manual System Setup Figure 11-3 Filter Setup window After setup, press to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. Then the System Setup screen appears. 11.2.1 Setting AC Filter Press Tab or Shift + Tab to move the cursor to the AC Filter list box in the Filter Setup window. Press the Up or Down arrow to highlight Off or On.
SE-12 Series Electrocardiograph User Manual System Setup DFT Filter greatly reduces the baseline fluctuations without affecting the ECG signals. The purpose of this filter is to keep the ECG signals on the baseline of the printout. The set value is the low limit of the frequency range, including 0.05Hz, 0.15Hz, 0.25Hz, 0.32Hz, 0.5Hz and 0.67Hz. 11.2.3 Setting EMG Filter Press Tab or Shift + Tab to move the cursor to the EMG Filter list box in the Filter Setup window.
SE-12 Series Electrocardiograph User Manual System Setup 11.3 Record Info Setup Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Record Info on the System Setup screen, press to open the Record Info Setup window. You can press F1 or F2 to switch between Setup 1 and Setup 2. After setup, press OK button, and then press to confirm, or press Tab or Shift + Tab to move the cursor to the to confirm. Then the System Setup screen appears. 11.3.
SE-12 Series Electrocardiograph User Manual System Setup In the auto mode, when Record Style is set to 3×4+1R, ECG waves of 12 leads are printed in 4 groups of 3, with the ECG wave of one rhythm lead on the bottom of the ECG reports. In the auto mode, when Record Style is set to 3×4+3R, ECG waves of 12 leads are printed in 4 groups of 3, with the ECG waves of three rhythm leads on the bottom of the ECG reports.
SE-12 Series Electrocardiograph User Manual System Setup When Record Sequence is set to Synchronous, all leads are printed simultaneously. The start time of each group is the same. 11.3.1.4 Setting Gain Press Tab or Shift + Tab to move the cursor to the Gain list box on the Setup 1 tab, and then press the Up or Down arrow to highlight an option. You can set the indicated height of 1mV ECG on the paper. You can set Gain to 10mm/mV, 20mm/mV, 10/5mm/mV, AGC, 2.5mm/mV or 5mm/mV. AGC means auto gain control.
SE-12 Series Electrocardiograph User Manual Options HP-DJ-D2468 HP-DJ-D5568 HP-LJ-P2035 System Setup Corresponding USB Printers HP Deskjet 3938 HP Deskjet D2368 HP Deskjet D2468 HP Deskjet D2568 HP Deskjet D5568 HP Laserjet P2015 HP Laserjet P2035 NOTE: 1. During the USB printing course,pressing the START/STOP key again can not stop printing ECG reports. 2. For details of the ECG report printed by the USB printer, please refer to Section 8.6.5, “ECG Reports Printed by the USB Printer”. 3.
SE-12 Series Electrocardiograph User Manual System Setup When the paper with black markers on the bottom is used and Paper Marker is set to Yes, the device can identify the start point of each page of the recorder paper while printing ECG reports. When Paper Marker is set to No, the device can not identify the start point of each page of the recorder paper while printing ECG reports. 11.3.1.8 Setting Sampling Time If Record Style is set to 12x1, you can set the time period.
SE-12 Series Electrocardiograph User Manual System Setup Auto Record Info includes Patient Info, Position Marker, Analysis, Template, Measure, Minnesota Code and Baseline Adjustment. When Patient Info, Position Marker, Analysis, Template, Measure or Minnesota Code is selected, it will be printed in the ECG reports. When Baseline Adjustment is set to Auto, the baselines of the lead groups are adjusted respectively.
SE-12 Series Electrocardiograph User Manual System Setup Figure 11-6 Patient Information Setup window After setup, press to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. Then the System Setup screen appears. 11.4.1 Selecting First/Last Name / Gender / D.O.B / Height / Weight / BP / Race / Pacemaker / Medication / Room No.
SE-12 Series Electrocardiograph User Manual System Setup You can set ID Mode to Auto, Time or Manual. When ID Mode is set to Auto, the patient ID can be automatically generated. The patient ID range is 0~1999, 999, 999. When ID Mode is set to Time, the patient ID can be automatically generated according to the time when you press the START/STOP key to print an ECG report. When ID Mode is set to Manual, you can enter the patient ID manually in the Patient Information window.
SE-12 Series Electrocardiograph User Manual System Setup 11.4.6 Selecting BP Unit Press Tab or Shift + Tab to move the cursor to BP Unit in the Patient Information Setup window, and then press the Up or Down arrow to select mmHg or kpa. When BP Unit is set to kpa, two extra edit boxes will be displayed in the Patient Information window for inputting decimal fraction. 11.4.7 Selecting Next Patient Press Tab or Shift + Tab to move the cursor to Next Patient in the Patient Information Setup window.
SE-12 Series Electrocardiograph User Manual System Setup When Next Patient is set to Off, in the auto mode, the system will not automatically display the next patient screen after a complete ECG report is printed. 11.4.8 Setting the Time to Clear Patient Information Press Tab or Shift + Tab to move the cursor to Patient Info Cleared in the Patient Information Setup window. Press the Up or Down arrow to highlight an option.
SE-12 Series Electrocardiograph User Manual System Setup Figure 11-7 Transmission Setup window After setup, press to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. Then the System Setup screen appears. 11.5.1 Setting Device No. Press Tab or Shift + Tab to move the cursor to the Device No. textbox, and then fill in the textbox. If you input 1 in the Device No. textbox, after you save data to the U disk, the data will be in the directory of ECGDATA\
SE-12 Series Electrocardiograph User Manual System Setup 11.5.2 Selecting Auto Transmission Press Tab or Shift + Tab to move the cursor to Auto Transmission, and then press the Up or Down arrow to select On or Off. When Auto Transmission is set to On, in the auto or rhythm mode, ECG data will be transmitted automatically after an ECG report is printed out; in the off mode, ECG data will be transmitted automatically after the START/STOP key is pressed. 11.5.
SE-12 Series Electrocardiograph User Manual System Setup To Set Local IP For the cross-network transmission, set the first two sections of the Local IP item to the first two sections of the IP of the PC. Set the third section of the Local IP item to the network segment of the electrocardiograph which depends on the configuration of Router. The last section of the Local IP item can be set at random.
SE-12 Series Electrocardiograph User Manual System Setup Figure 11-8 Lead Setup window After setup, press OK button, and then press to confirm, or press Tab or Shift + Tab to move the cursor to the to confirm. Then the System Setup screen appears. 11.6.1 Setting Rhythm Lead1/2/3 Press Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3 list box in the Lead Setup window, and then press the Up or Down arrow to highlight an option.
SE-12 Series Electrocardiograph User Manual System Setup In the R-R analysis mode, the R-R analysis report of the rhythm lead selected in the Rhythm Lead1 list box will be printed. 11.6.2 Setting Lead Sequence Press Tab or Shift + Tab to move the cursor to the Lead Sequence list box in the Lead Setup window, and then press the Up or Down arrow to highlight Standard or Cabrera.
SE-12 Series Electrocardiograph User Manual System Setup Figure 11-9 Display&Sound Setup window 11.7.1 Setting Brightness Press Tab or Shift + Tab to move the cursor to the Brightness progress bar in the Display&Sound Setup window, and then press the Left or Right arrow to adjust the brightness of the LCD screen. You can set the brightness within 0~20. 11.7.
SE-12 Series Electrocardiograph User Manual System Setup 11.7.4 Setting Key Volume Press Tab or Shift + Tab to move the cursor to the Key Volume list box in the Display&Sound Setup window, and then press the Up or Down arrow to highlight an option. When you press keys on the keyboard, the electrocardiograph gives a short sound. You can set Key Volume to Low, Medium, High or Off. When it is set to Off, there is no sound when you press keys. 11.7.
SE-12 Series Electrocardiograph User Manual System Setup Figure 11-10 Date&Time Setup window After setup, press to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. Then the System Setup screen appears. 11.8.1 Setting Date and Time Press Tab or Shift + Tab to move the cursor to the Date or Time textbox in the Date&Time Setup window, and then input the date or the time manually.
SE-12 Series Electrocardiograph User Manual System Setup 11.8.3 Setting Time Mode Press Tab or Shift + Tab to move the cursor to the Time Mode list box, and then press the Up or Down arrow to highlight 24 Hours or 12 Hours. 11.8.4 Setting Power-Off Time Press Tab or Shift + Tab to move the cursor to the Power Off textbox in the Date&Time Setup window, and then input the power-off time manually. When Power Off is not set or is set to 0 Minutes, this function will not be effective. NOTE: 1.
SE-12 Series Electrocardiograph User Manual System Setup Figure 11-11 File Setup window After setup, press to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. Then the System Setup screen appears. 11.9.1 Selecting Auto Save Press Tab or Shift + Tab to move the cursor to the Auto Save list box in the File Setup window, and then press the Up or Down arrow to highlight Off or On.
SE-12 Series Electrocardiograph User Manual System Setup 11.9.3 Setting File Format Press Tab or Shift + Tab to move the cursor to File Format in the File Setup window, and then press the Up or Down arrow to highlight an option. If you want to set File Format to SCP/FDA-XML/PDF, first activate the SCP/FDA-XML/PDF function on the Advanced Setup screen For details on activating the SCP/FDA-XML/PDF function, please refer to Section 9.1.3, “Transmitting ECG Data in SCP/FDA-XML/PDF Format Through Net Cable”.
SE-12 Series Electrocardiograph User Manual System Setup 11.10 System Maintenance Setup Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Maintenance on the System Setup screen, press to open the System Maintenance window. Figure 11-12 System Maintenance window After setup, press OK button, and then press to confirm, or press Tab or Shift + Tab to move the cursor to the to confirm. Then the System Setup screen appears. 11.10.
SE-12 Series Electrocardiograph User Manual System Setup 11.10.2 Loading System Settings from U Disk 1. Connect the U disk containing the system settings to the electrocardiograph. 2. Press Tab or Shift + Tab to move the cursor to the Load Settings From U disk button in the System Maintenance window, and then press to load the system settings from the U disk to the electrocardiograph. 11.10.
SE-12 Series Electrocardiograph User Manual System Setup Table 11-1 System Setup Defaults Resting ECG Setup Default 1 Mode Options Auto, Manual, Rhythm 2 Manual Style 6 channels 3 Auto Display Style 6×2 4 Rhythm Style Three Leads 5 Sampling Mode Real-time Sample 6 Duration (Periodic Sample) 60 min 7 Interval (Periodic Sample) 1 min 8 Preview Off 9 Auto Extend Record Off 10 AC filter On 11 EMG filter Off 12 DFT filter 0.
SE-12 Series Electrocardiograph User Manual System Setup 27 Minnesota Code Off 28 Baseline Adjustment Horizontal 29 R-R Histogram On 30 R-R Interval List On 31 R-R Waveform On 32 Grid of Thermal Report Off 33 Grid of USB Report On 34 Gender\Height\Weight On 35 First/Last Name \ D.O.B \ BP \ Race \ Pacemaker \ Medication \ Room No. \ Department \ Physician \ Technician \ Ref-Physician Off 36 ID Hint On 37 Next Patient Off 38 ID Mode Auto 39 Patient Information Cleared
SE-12 Series Electrocardiograph User Manual System Setup 53 Antialiasing Off 54 QRS Volume Off 55 Hint Volume Medium 56 Key Volume Medium 57 Date Mode DD-MM-YYYY 58 Time Mode 24 Hours 59 Power Off 0 60 LCD Off 0 61 Auto Save On 62 Saving Path ECG 63 File Format DAT 64 Delete After Trans.
SE-12 Series Electrocardiograph User Manual System Setup 12 Max Predicted HR 220 13 The proportion of Target HR in Max Predicted HR 85 14 AC Filter On 15 DFT Filter 0.
SE-12 Series Electrocardiograph User Manual System Setup 5 Sampling Mode Real-time Sample 6 Duration (Periodic Sample) 60 min 7 Interval (Periodic Sample) 1 min 8 Preview Off 9 Auto Extend Record Off 10 AC filter On 11 EMG filter Off 12 DFT filter 0.
SE-12 Series Electrocardiograph User Manual System Setup 34 Gender\Height\Weight\D.O.B On 35 First/Last Name \ BP \ Race \ Pacemaker \ Medication \ Room No. \ Department \ Physician \ Technician \ Ref-Physician Off 36 ID Hint On 37 Next Patient Off 38 ID Mode Manual 39 Patient Information Cleared After Recording 40 H/W Unit inch/lb 41 BP Unit mmHg 42 Prompt Confirmed By 43 User Defined Cleared 44 Auto Transmission Off 45 Transmission Mode Ethernet 46 FTP User Name/FTP
SE-12 Series Electrocardiograph User Manual System Setup 60 LCD Off 0 61 Auto Save On 62 Saving Path ECG 63 File Format PDF 64 Delete After Trans.
SE-12 Series Electrocardiograph User Manual System Setup 18 Rhythm Lead 1 II 19 Rhythm Lead 2 V1 20 Rhythm Lead 3 V5 21 Lead Sequence Standard 22 Gain 10mm/mV 23 Paper Marker Yes 24 Exercise Report Late Stage 25 Report Style 3×4+1 26 Recovery Report On 27 Start time of Recovery Report 10s 28 Interval of Recovery Report 2min 29 Summary Report On 30 ST Scope Report On 31 Trend Graph Report On 32 Edit Conclusion On 11.10.
SE-12 Series Electrocardiograph User Manual System Setup 11.11 Other Setup Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Other on the System Setup screen, press to open the Other Setup window. Figure 11-13 Other Setup Window After setup, press OK button, and then press to confirm, or press Tab or Shift + Tab to move the cursor to the to confirm. Then the System Setup screen appears. 11.11.
SE-12 Series Electrocardiograph User Manual System Setup 11.11.3 Selecting Caps Lock Press Tab or Shift + Tab to move the cursor to Caps Lock on the Other Setup Window, and then press the Up or Down arrow to highlight Off or On. When Caps Lock is set to On, the letters entered will be capital. Pressing Shift and a letter key can input a lowercase letter. When Caps Lock is set to Off, the letters entered will be lowercase. Pressing Shift and a letter key can input a capital letter. 11.11.
SE-12 Series Electrocardiograph User Manual System Setup 11.11.6 Setting Pacemaker Detection Sensitivity Press Tab or Shift + Tab to move the cursor to Pacemaker Detection Sens. on the Other Setup Window, and then press the Up or Down arrow to highlight Low or High. When Pacemaker Detection Sens. is set to High, the pacemaker signals are easy to be detected. When Pacemaker Detection Sens. is set to Low, the pacemaker signals are not easy to be detected.
SE-12 Series Electrocardiograph User Manual Chapter 12 Exercise ECG Operation Instructions for Exercise ECG (Optional) Operation Instructions (Configurable for for SE-12 Express) 12.1 Activating Exercise ECG Function 1. When the main screen is displayed, press the function key F6 below Setup to open the System Setup screen. 2. Press F1 to display the Service Password window.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) 3. Press Tab or Shift + Tab to move the cursor to the Activate button after Stress Test. Press to open the Activate Password window. Enter the password and press to activate the exercise ECG function. NOTE: Please contact local distributors to get the service password and the activating password. 12.2 Operation Preparation 1.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) CAUTION Only the treadmill/ergometer recommended by the manufacturer can be connected to Serial Port 1 and Serial Port 2. 1) Connect the electrocardiograph to the treadmill RS232 port RS232 port Connect the RS232 port of the treadmill to the RS232 port of the electrocardiograph with an RS232 cable. NOTE: The position of the RS232 port may be different on different treadmills.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) 2) Connect the electrocardiograph to the ergometer RS232 port Connect the RS232 in port of the ergometer to the RS232 port of the electrocardiograph with an RS232 cable. NOTE: The position of the RS232 in port may be different on different ergometers. 3. Connect the electrocardiograph to the stress BP monitor 4. Connect the power cords, and connect the earth wire (when necessary) 5.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) 6. Turn on the stress BP monitor and treadmill/ergometer 7. Set the electrocardiograph 1) After turning on the electrocardiograph, press Pretest on the keyboard to open the main screen of the exercise test. 2) Press the function key F6 below Setup to open the System Setup screen. 3) Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) 5) Press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. Then the System Setup screen appears. 6) Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Protocol on the System Setup screen, press to open the Protocol Manager screen. Select a protocol by pressing the Up or Down arrow.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) 7) Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor to Return, and then press to return to the main screen of the exercise test. 8. Setting the treadmill or the ergometer for external control 1) For TM-400 treadmill, it is ready for external control and needn’t be set.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) Select Prot id and select the second RS232 protocol which is the Trackmaster communication protocol Select Baudrate and set this to 4800 Baud Press the red button to leave the service menu Turn off the Valiant 3) For Lode ergometer Turn on the ergometer By default, the Corival ergometer is shipped with the LODE RS232 communication protocol. It should be set to the ERGOLINE P4 protocol.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) If you turn off the ergometer, it will start up in the ANALOG mode every time the ergometer is turned on again. The Lode ergometer is now ready for external control For details on setting other models for external control, please refer to the connection instruction of the models. 9.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) 12.4 System Setup for Exercise Test After turning on the electrocardiograph, press Pretest on the keyboard to open the main screen of the exercise test. Press the function key F6 below Setup to open the System Setup screen. Figure 12-1 System Setup screen for Exercise Test After setup, 1. Press ESC to return to the main screen of the exercise test. 2.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) Figure 12-2 General Information Setup window Press Tab or Shift + Tab to move the cursor to a list box, and then press the Up or Down arrow to highlight an option. Or press Tab or Shift + Tab to move the cursor to a textbox, and then fill in the textbox. After setup, press to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) NOTE: When the main screen of the exercise test is displayed, pressing the Up or Down arrow on the keyboard can switch the display styles. When the display style is 3×1, pressing the Left or Right arrow on the keyboard can switch the lead groups. 2. Speed Unit You can set Speed Unit to mph or km/h. 3. Post J Post J is the length after J point of ST segment. You can set Post J to 0ms, 20ms, 40ms, 60ms or 80ms.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) 6. Max Predicted HR and Target HR The following formulas are applied to calculate the target heart rate. Max Predicted HR = 220 - Age Target HR = Max Predicted HR * 85% NOTE: 220 and 85 are default values, and you can modify them in different situations. 12.4.2 Record Information Setup Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor.
SE-12 Series Electrocardiograph User Manual After setup, press Operation Instructions for Exercise ECG (Optional) to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. Then the System Setup screen appears. 1. Exercise Report When Close is highlighted, 12-lead ECG report of the exercise phase will not be printed out automatically during the exercise test.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) 4. Final Report You can select Summary Report, ST Scope Report, Trend Graph Report and Edit Conclusion as the final reports. After an exercise test is terminated, you can press the START/STOP key on the keyboard to print out the selected final reports. NOTE: Pressing the START/STOP key during the printing course can stop printing. 12.4.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) In the Add Protocol dialog box, firstly you should enter the new protocol name in the Protocol Name textbox, and then enter the stage information of every stage, including the time, speed and grade. After that, press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) NOTE: For the Bruce or Modified Bruce protocol, only the stage information of the pretest and recovery phases can be edited. 3. Select Select a protocol on the Protocol Manager screen by pressing the Up or Down arrow, and press the function key F3 below Select to set the selected protocol as the default protocol. 4.
SE-12 Series Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) 5. Del All To delete all the protocols from the Protocol Manager screen, you can press the function key F5 below Del All to display the following dialog box. Press Tab or Shift + Tab to move the cursor to the OK button, and then press to delete all the protocols. 6. Restore Press the function key F6 below Restore to display the following dialog box.
SE-12 Series Electrocardiograph User Manual Turning off the Electrocardiograph Chapter 13 Turning off the Electrocardiograph When the battery is used, hold down the key to display the hint System is shutting down… on the screen. Then the device will be off a few seconds later. When the mains supply is used, hold down the key to display the hint System is shutting down… on the screen. Then the device will be off a few seconds later. Remove the plug from the outlet. NOTE: 1.
SE-12 Series Electrocardiograph User Manual Hint Information Chapter 14 Hint Information Hint information and the corresponding causes provided by the electrocardiograph are listed in the following tables. Table 14-1 Hint Information and Causes Hint Information Causes Lead off Electrodes fall off the patient or the patient cable falls off the unit, or a high polarization voltage occurs.. Battery Weak The battery is weak. No Paper Recorder paper runs out or is not loaded.
SE-12 Series Electrocardiograph User Manual Hint Information Table 14-2 Hint Information and Causes for Exercise ECG Hint Information Causes Battery Weak The battery is weak. No Paper Recorder paper runs out or is not loaded. Paper Error When Paper Marker is set to Yes, the electrocardiograph advances the recorder paper to the next black marker. If it advances the paper for 300mm and can not find the next black marker, the hint Paper Error is displayed.
SE-12 Series Electrocardiograph User Manual Cleaning, Care and Maintenance Chapter 15 Cleaning, Care and Maintenance CAUTION 1. Turn off the power before cleaning and disinfection. The mains supply must be switched off if it is used. 2. Prevent the detergent from seeping into the equipment. 15.1 Cleaning CAUTION Any remainder of detergent should be removed from the unit and the patient cable after cleaning. 15.1.1 Cleaning the Main Unit and the Patient Cable To clean the electrocardiograph: 1.
SE-12 Series Electrocardiograph User Manual Cleaning, Care and Maintenance 3. Clean them in warm water and make sure there is no remainder gel. 4. Dry the electrodes with a clean dry cloth or air dry naturally. CAUTION Do not clean the unit and accessories with abrasive fabric and avoid scratching the electrodes. 15.1.3 Cleaning the Print Head To clean the print head: 1. Open the recorder casing and remove the paper. 2. Wipe the print head gently with a clean soft cloth damped in 75% alcohol.
SE-12 Series Electrocardiograph User Manual Cleaning, Care and Maintenance 15.3 Care and Maintenance CAUTION Operate the electrocardiograph, charge the battery, and store the battery at a temperature of 40℃(104ºF) or lower. Exposure to higher or lower temperature may reduce battery life, damage the battery, and degrade overall electrocardiograph performance. 15.3.
SE-12 Series Electrocardiograph User Manual Cleaning, Care and Maintenance 3) Replacement When the useful life of the battery is over, or foul smell and leakage are found, please contact the manufacturer or the local distributor for replacement. WARNING 1. Only qualified service engineers authorized by the manufacturer can open the battery compartment and replace the battery, and the battery of the same model and specification provided by the manufacturer must be used. 2.
SE-12 Series Electrocardiograph User Manual ♦ Cleaning, Care and Maintenance Do not overlap the recorder paper for a long time, or else the ECG reports may trans-print each other. 15.3.3 Visual inspection Perform a visual inspection of all equipment and peripheral devices daily. If you notice any items that need repair, contact a qualified service engineer to make the repairs. ♦ Check the case and display screen for cracks or other damage.
SE-12 Series Electrocardiograph User Manual Cleaning, Care and Maintenance 8. Test the patient leakage current according to IEC/EN 60601-1: Limit: NC a.c. 10μA, d.c. 10μA; SFC a.c. 50μA, d.c. 50μA. 9. Test the patient auxiliary current according to IEC/EN 60601-1: Limit: NC a.c. 10μA, d.c. 10μA; SFC a.c. 50μA, d.c. 50μA. 10. Test the patient leakage current under single fault condition with mains voltage on the applied part according to IEC/EN 60601-1: Limit: 50μA (CF).
SE-12 Series Electrocardiograph User Manual Cleaning, Care and Maintenance 3) Reusable Electrodes ♦ Electrodes must be cleansed after use and make sure there is no remainder gel on them. ♦ Keep suction bulbs of chest electrodes away from sunshine and excessive temperature. ♦ After long-term use, the surfaces of electrodes will be oxidized because of erosion and other causes. By this time, electrodes should be replaced to achieve high-quality ECG records.
SE-12 Series Electrocardiograph User Manual Accessories List Chapter 16 Accessory List WARNING Only the patient cable and other accessories supplied by the manufacturer can be used. Or else, the performance and electric shock protection can not be guaranteed. Accessory Part Number 01.57.107581 (Snap Style) ECG Cable (European) 01.57.107583 (Grabber Style) 01.57.107582 (Snap Style) ECG Cable (American) 01.57.107584 (Grabber Style) Patient Cable (American) 01.57.107048 Patient Cable (European) 01.
SE-12 Series Electrocardiograph User Manual Accessories List Input/Output Connector 11.13.19907 Grounding Wire 11.13.114214 03.24.38952 Smart ECG Viewer Software 03.24.38953 Snap Clip 01.57.040172 Electrode Gel 11.25.78047 RS232 Cable 11.13.20117 Ethernet Cable 11.13.20096 Smart ECG Key/TINY-SPRO viewer Software 01.18.47116 Smart ECG viewer Software Key/USB 12.01.47194 DeskJet/HP2368 11.18.52168 U Disk 11.18.78105 ECG Bag 11.56.78118 SD Card Reader 11.17.
SE-12 Series Electrocardiograph User Manual Warranty & Service Chapter 17 Warranty & Service 17.1 Warranty EDAN warrants that EDAN’s products meet the labeled specifications of the products and will be free from defects in materials and workmanship that occur within warranty period. The warranty is void in cases of: a) damage caused by mishandling during shipping. b) subsequent damage caused by improper use or maintenance. c) damage caused by alteration or repair by anyone not authorized by EDAN.
SE-12 Series Electrocardiograph User Manual Technical Specifications Appendix 1 Technical Specifications A1.1 Safety Specifications Comply with: IEC/EN 60601-1+A1+A2, IEC/EN 60601-1-2+A1, IEC/EN 60601-2-25, ANSI/AAMI EC11, IEC/EN 60601-2-51, UL 60601-1, CAN/CSA C22.2 No.601.
SE-12 Series Electrocardiograph User Manual Technical Specifications A1.3 Physical Specifications Dimensions Weight Display SE-12: 420mm×330mm×105mm (16.5in×13.0in×4.1in) SE-12 Express: 420mm×330mm×120mm (16.5in×13.0in×4.7in) SE-12: approx. 5.0kg (11.0 lbs); SE-12 Express: approx. 6.5kg (14.3 lbs) (Excluding recorder paper and battery) 320×240 dot single color LCD Screen 800×600 multicolor LCD Screen A1.
SE-12 Series Electrocardiograph User Manual Technical Specifications Cycle life ≥ 300 times Fuse: T1AL250VP Ø5×20 A1.
SE-12 Series Electrocardiograph User Manual Technical Specifications Calibration Voltage: 1mV±2% DC Offset Voltage: ±600mV Noise: ≤12.5 μVp-p Multichannel crosstalk ≤0.5mm AC Filter: On/Off Filter DFT Filter: 0.05Hz/0.15Hz/0.25Hz/0.32Hz/0.5Hz/0.67Hz EMG Filter: 25Hz/35Hz/45Hz/OFF LOWPASS Filter:150Hz/100Hz/75Hz CMRR ≥115dB Sampling Frequency 1000 Hz Pacemaker Detection Amplitude ±2 to ±700 mV Width 0.1 to 2.
SE-12 Series Electrocardiograph User Manual EMC Information Appendix 2 EMC Information Guidance and manufacture’s declaration - electromagnetic emissions - for all EQUIPMENT and SYSTEMS Guidance and manufacture’s declaration – electromagnetic emission The 12-channel electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or the user of the 12-channel electrocardiograph should assure that it is used in such an environment.
SE-12 Series Electrocardiograph User Manual EMC Information Guidance and manufacture’s declaration – electromagnetic immunity - for all EQUIPMENT and SYSTEMS Guidance and manufacture’s declaration – electromagnetic immunity The 12-channel electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or the user of 12-channel electrocardiograph should assure that it is used in such an environment.
SE-12 Series Electrocardiograph User Manual (30% dip in UT) for 25 cycles NOTE EMC Information (30% dip in UT) for 25 cycles 12-channel electrocardiograph be powered from an uninterruptible power supply or a battery. <5% UT <5% UT (>95% dip in UT) (>95% dip in UT) for 5 sec for 5 sec UT is the a.c. mains voltage prior to application of the test level.
SE-12 Series Electrocardiograph User Manual EMC Information less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
SE-12 Series Electrocardiograph User Manual EMC Information Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM - for EQUIPMENT or SYSTEM that is not LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and the 12-channel Electrocardiograph The 12-channel electrocardiograph is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
SE-12 Series Electrocardiograph User Manual Abbreviation Appendix 3 Abbreviation Abbr English LCD Liquid Crystal Display BP Blood Pressure ECG Electrocardiogram/Electrocardiograph HR Heart Rate aVF Left Foot Augmented Lead aVL Left Arm Augmented Lead aVR Right Arm Augmented Lead LA Left Arm LL Left Leg RA Right Arm RL Right Leg ID Identification AC Alternating Current USB Universal Serial Bus AGC Auto Gain Control - 196 -