About this Manual P/N: 01.54.455422-10 Release Date: November 2011 © Copyright EDAN INSTRUMENTS, INC. 2011. All rights reserved. Statement This manual will help you understand the operation and maintenance of the product better. It is reminded that the product shall be used strictly complying with this manual. User’s operation failing to comply with this manual may result in malfunction or accident for which EDAN INSTRUMENTS, INC. (hereinafter called EDAN) can not be held liable.
The electrical installation of the relevant room complies with national standards, and The instrument is used in accordance with the instructions for use. Upon request, EDAN may provide, with compensation, necessary circuit diagrams, and other information to help qualified technician to maintain and repair some parts, which EDAN may define as user serviceable. Terms Used in this Manual This guide is designed to give key concepts on safety precautions.
Table of Contents Chapter 1 Safety Guidance ........................................................................................................... 1 1.1 Intended Use ........................................................................................................................... 1 1.2 Warnings and Cautions ........................................................................................................... 1 1.2.1 Safety Warnings .........................................................
Chapter 6 Entering Patient Information ................................................................................... 47 6.1 Entering Patient ID ............................................................................................................... 47 6.2 Entering Other Information .................................................................................................. 48 6.3 Using a Bar Code Reader ......................................................................................
10.7 Printing Files..................................................................................................................... 102 10.8 Previewing a File .............................................................................................................. 102 Chapter 11 System Setup........................................................................................................... 105 11.1 Work Mode Setup .............................................................................
11.6.1 Setting Rhythm Lead1/2/3 ......................................................................................... 125 11.6.2 Setting Lead Sequence ............................................................................................... 126 11.6.3 Selecting Nehb ........................................................................................................... 126 11.7 Display&Sound Setup ..................................................................................................
Chapter 12 Turning off the Electrocardiograph ..................................................................... 143 Chapter 13 Hint Information.................................................................................................... 144 Chapter 14 Cleaning, Care and Maintenance ......................................................................... 145 14.1 Cleaning............................................................................................................................
SE-1200 Series Electrocardiograph User Manual Safety Guidance Chapter 1 Safety Guidance This chapter provides important safety information related to the use of the 12-channel electrocardiograph. 1.1 Intended Use The intended use of the 12-channel electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals.
SE-1200 Series Electrocardiograph User Manual Safety Guidance 1.2.1 Safety Warnings WARNING 1. The electrocardiograph is intended to be used by qualified physicians or personnel professionally trained. They should be familiar with the contents of this user manual before operation. 2. Only qualified service engineers can install this equipment, and only service engineers authorized by the manufacturer can open the shell. 3.
SE-1200 Series Electrocardiograph User Manual Safety Guidance WARNING 14. The disposable electrodes can only be used for one time. 15. No hazard is found when the electrocardiograph is operated with cardiac pacemakers or other stimulators. 16. Do not touch the patient, bed, table or the equipment while using the ECG together with a defibrillator. 17. Do not touch accessible parts of non-medical electrical equipment and the patient simultaneously. 18.
SE-1200 Series Electrocardiograph User Manual Safety Guidance WARNING 26. Multiple portable socket-outlets shall not be placed on the floor. 27. Do not use the additional multiple portable socket-outlet or extension cord in the medical electrical system, unless it’s specified as part of the system by manufacturer. And the multiple portable socket-outlets provided with the system shall only be used for supplying power to equipment which is intended to form part of the system. 28.
SE-1200 Series Electrocardiograph User Manual Safety Guidance 1.2.2 Li-ion Battery Care Warnings WARNING 1. Improper operation may cause the internal li-ion battery (hereinafter called battery) to be hot, ignited or exploded, and it may lead to the decrease of the battery capacity. It is necessary to read the user manual carefully and pay more attention to warning messages. 2.
SE-1200 Series Electrocardiograph User Manual Safety Guidance CAUTION 2. Do not use the equipment in a dusty environment with bad ventilation or in the presence of corrosive. 3. Make sure that there is no intense electromagnetic interference source around the equipment, such as radio transmitters or mobile phones etc. Attention: large medical electrical equipment such as electrosurgical equipment, radiological equipment and magnetic resonance imaging equipment etc.
SE-1200 Series Electrocardiograph User Manual Potential equalization Mains supply USB socket Battery indicator Battery recharging indicator Delete key Enter key Esc key/Reset key Shift key Fn key Power On/Off key COPY key MODE key SLEEP/WAKE UP key -7- Safety Guidance
SE-1200 Series Electrocardiograph User Manual Safety Guidance START/STOP key Tab key/Feed paper key UP/DOWN/LEFT/RIGHT Arrow key Recycle Part Number Serial Number Date of Manufacture Manufacturer Authorized Representative in the European Community The symbol indicates that the device complies with the European Council Directive 93/42/EEC concerning medical devices. With respect to electrical shock, fire and mechanical hazards only in accordance with UL 60601-1and CAN/CSA C22.2 No. 601.
SE-1200 Series Electrocardiograph User Manual Rx only (U.S.) Safety Guidance Federal (US) law restricts this device to sale by or on the order of a physician. It indicates that the device should be sent to the special agencies according to local regulations for separate collection after its useful life.
SE-1200 Series Electrocardiograph User Manual Introduction Chapter 2 Introduction The 12-channel electrocardiograph gathers ECG signals of 12 leads simultaneously. It displays the operation menu, ECG parameters as well as electrocardiograms. The 12-channel ECG waves can be viewed on the LCD screen and printed out by using a high-quality thermal recorder. The sampled ECG data can be saved, transmitted and exported. The manual, auto, rhythm, R-R analysis or off mode can be chosen freely.
SE-1200 Series Electrocardiograph User Manual Introduction 2.
SE-1200 Series Electrocardiograph User Manual Introduction 2.2 Keyboard and Keys A B F C G H D E I J L K M N Figure 2-3 SE-1200 Express/SE-1200 Keyboard Name Explanation A Function Key Selecting menu functions on the screen B Delete Deleting characters C Enter Confirming operation Moving the cursor Pressing Tab can move the cursor forward, and pressing Shift + Tab can move the cursor backward.
SE-1200 Series Electrocardiograph User Manual Introduction Press Shift and a number key to input the special character in the top right corner of the key. If Caps Lock is set to Off, pressing Shift + P can type a capital P. If Caps Lock is set to On, pressing Shift + P can type a lowercase p. Canceling operation H Esc/Reset I Power On/Off NOTE: A large polarization voltage may cause baseline drift.
SE-1200 Series Electrocardiograph User Manual Introduction 2.3 Front Panel Figure 2-4 SE-1200 Express Front Panel Figure 2-5 SE-1200 Front Panel Symbol A B C Name Explanation Mains supply indicator When the device is powered by the mains supply, this indicator is lit. Battery indicator When the device is powered by battery, this indicator is lit. Battery recharging indicator When the battery is being recharged, this indicator is lit.
SE-1200 Series Electrocardiograph User Manual Introduction 2.4 Rear Panel Figure 2-6 SE-1200/SE-1200 Express Rear Panel Name A Explanation Mains Supply Socket AC SOURCE: alternating current supply socket Potential Equalization Conductor Potential equalization conductor provides a connection between the unit and the potential equalization bus bar of the electrical installation. B C Handle Part for people to hold D Heat Emission Hole Path for internal heat emission 2.
SE-1200 Series Electrocardiograph User Manual Introduction Name Explanation A Patient Cable Socket Connecting to the patient cable B Serial Port 1 Connecting to a PC C USB Socket 1 (Optional) Standard USB socket, connecting to a PC USB Socket 2 (Optional) Standard USB socket, connecting to a U disk, a bar code reader or a USB printer recommended by the manufacturer E External Input / Output Socket Connecting to the external signal device F Serial Port 2 Reserved G Net port Standard ne
SE-1200 Series Electrocardiograph User Manual Introduction 2) Serial Port 1 WARNING The isolated intensity of serial port 1 is 1500V AC and the maximum voltage applied should not exceed +15V DC.
SE-1200 Series Electrocardiograph User Manual Introduction 4) External Input/Output Socket Definitions of corresponding pins: Pin Signal Pin Signal 1 GND 4 GND 2 GND 5 ECG Signal (input) 3 GND 6 ECG Signal (output) 5) USB Socket 1/USB Socket 2 (Optional) CAUTION Only the USB equipment recommended by the manufacturer can be connected to the USB interface. Definitions of corresponding pins: Pin Signal Pin Signal 1 +5V 3 D+ 2 D- 4 GND WARNING 1.
SE-1200 Series Electrocardiograph User Manual Introduction WARNING Furthermore all configuration shall comply with the valid version of the standard IEC/EN 60601-1-1. Therefore anybody, who connects additional equipment to the signal input or output connector to configure a medical system, must make sure that it complies with the requirements of the valid version of the system standard IEC/EN 60601-1-1. If in doubt, consult our technical service department or your local distributor. 2.
SE-1200 Series Electrocardiograph User Manual D Heat Emission Hole Path for internal heat emission E Label Position for product information label Introduction 1) Battery Compartment The rated voltage and the rated capacity of the battery pack are as follows: SE-1200 Express: Rated Voltage: 14.8V; Rated Capacity: 4400mAh SE-1200: Rated Voltage: 14.8V; Rated Capacity: 2200mAh WARNING 1.
SE-1200 Series Electrocardiograph User Manual Introduction 2.
SE-1200 Series Electrocardiograph User Manual About the Electrocardiograph Application Chapter 3 About the Electrocardiograph Application The following sections provide an overview of the main operations and functions. You can operate the electrocardiograph by using the keyboard. For SE-1200 Express, you can also carry out the operations by using the touch screen. CAUTION Do not touch the LCD screen with sharp things such as pencils or pens; otherwise, it will be damaged. 3.
SE-1200 Series Electrocardiograph User Manual About the Electrocardiograph Application 3.2 Entering Data 1. When the main screen is displayed, press the function key F1 below Patient to open the Patient Information window. 2. Press Tab or Shift + Tab to move the cursor to the Name textbox. 3. Press on the keyboard to erase the typed information. 4. Press the letter or number keys on the keyboard to input patient name.
SE-1200 Series Electrocardiograph User Manual About the Electrocardiograph Application 3.3 Selecting an Item When the main screen is displayed, press the function key F6 below Setup to open the System Setup window. The cursor is on Work Mode. Press to open the Work Mode Setup window. 1. Press Tab or Shift + Tab to move the cursor to R-R, press Spacebar to select it, and then a check mark √ appears in the box before R-R. 2. Press Tab or Shift + Tab to move the cursor to the Rhythm Style list box.
SE-1200 Series Electrocardiograph User Manual About the Electrocardiograph Application 3.4 About the main screen After you turn on the device, the main screen pops up.
SE-1200 Series Electrocardiograph User Manual A About the Electrocardiograph Application Name Explanation Name Patient Name: within 60 ASCII characters or the equivalent number of other characters that can be supported by equivalent memory used by 60 ASCII characters When ID Mode is set to Manual, the patient ID is within 30 ASCII characters. B ID When ID Mode is set to Auto, the patient ID is 0~1999, 999, 999. When ID Mode is set to Time, the patient ID is within 10 characters.
SE-1200 Series Electrocardiograph User Manual N O Gain Speed About the Electrocardiograph Application Gain: 2.5 mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV, 10/5 mm/mV or AGC NOTE: This setup can’t be saved if it is modified on the main screen. In the manual mode, you can set Speed to 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s or 50mm/s. In the auto and rhythm modes, only 25mm/s and 50mm/s are available. In the R-R analysis mode, only 25mm/s is available.
SE-1200 Series Electrocardiograph User Manual About the Electrocardiograph Application On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on an item, press to open the setup window of the item. Take the Work Mode Setup window for example: Check Box Textbox Dropdown List Box 1. In the Work Mode Setup window, press Tab or Shift + Tab to move the cursor among different check boxes.
SE-1200 Series Electrocardiograph User Manual About the Electrocardiograph Application 2. In the Work Mode Setup window, press Tab or Shift + Tab to move the cursor to a list box or dropdown. Press the Up or Down arrow to highlight an option. 3. In the Work Mode Setup window, press Tab or Shift + Tab to move the cursor to a textbox. Fill in the selected textbox. 4. Press and then press to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, to confirm. 5.
SE-1200 Series Electrocardiograph User Manual A B C D About the Electrocardiograph Application E F G M N Figure 3-3 File Manager screen1 H I J K L Figure 3-4 File Manager screen2 - 30 -
SE-1200 Series Electrocardiograph User Manual About the Electrocardiograph Application Name Explanation A Trans All Press the function key F1 below Trans All to transmit all the files to the PC. B Export All Press the function key F2 below Export All to export all the files from the electrocardiograph to the U disk. C Del All Press the function key F3 below Del All to delete all the files from the electrocardiograph.
SE-1200 Series Electrocardiograph User Manual Operation Preparations Chapter 4 Operation Preparations WARNING Before use, the equipment, patient cable and electrodes should be checked. Replace them if there is any evident defectiveness or aging which may impair the safety or the performance, and make sure that the equipment is in proper working condition. 4.1 Power and Earthing WARNING If the integrity of the external protective conductor is in doubt, the equipment should be powered by battery.
SE-1200 Series Electrocardiograph User Manual Operation Preparations ♦ Battery When the battery is used, turn on the unit by pressing battery indicator ( on the keyboard. Then the ) will be lit and the battery symbol will be displayed on the LCD screen. Because of the consumption during the storage and transport course, the battery capacity may not be full. If the symbol and the hint information Battery Weak are displayed, which means the battery capacity is low, please recharge the battery first.
SE-1200 Series Electrocardiograph User Manual Operation Preparations 4.2 Loading/Replacing Recorder Paper Two kinds of recorder paper can be used. One is the folded thermal paper, and the other is the rolled thermal paper. NOTE: 1. When the folded thermal paper is used, the paper roller is unnecessary and must be taken out. 2. When using the paper of 216mm in width, the two movable parts should be removed. 3. The exit edge can help you tear the recorder paper.
SE-1200 Series Electrocardiograph User Manual Operation Preparations Loading/Replacing Process of Folded Paper: 1) Press the casing button downwards and remove the casing to open the recorder. 2) Remove the remainder paper from the paper tray if necessary. 3) Take off the wrapper of the new folded paper, and then put it in the paper tray. 4) Pull the paper out with the grid side facing the thermal print head, and replace the casing on the recorder.
SE-1200 Series Electrocardiograph User Manual Operation Preparations 5) Press down the recorder casing firmly. 6) To select Paper Marker. Press on the keyboard to turn on the unit. After the main screen is displayed, press the function key F6 below Setup to display the System Setup screen. Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Record Info on the System Setup screen, press to open the Record Info Setup window.
SE-1200 Series Electrocardiograph User Manual Operation Preparations 4) Place the paper and the roller gently in the recorder with the roller pin clicking into the groove. Groove 5) Pull the paper out with the grid side facing the thermal print head, and replace the casing on the recorder. 6) Press down the recorder casing firmly. 7) To select Paper Marker. Press on the keyboard to turn on the unit.
SE-1200 Series Electrocardiograph User Manual Operation Preparations 8) Advance the recorder paper. When the main screen is displayed, you can press 2.5cm. Press to advance the paper for again to stop advancing the paper. 4.3 Preparing the Patient 4.3.1 Instructing the Patient Before attaching the electrodes, greet the patient and explain the procedure. Explaining the procedure decreases the patient’s anxiety. Reassure the patient that the procedure is painless. Privacy is important for relaxation.
SE-1200 Series Electrocardiograph User Manual 4.4 Connecting the Patient Electrocardiograph and Electrodes Operation Preparations Cable to the WARNING The performance and electric shock protection can be guaranteed only if the original patient cable and electrodes of the manufacturer are used. The patient cable includes the main cable and lead wires which can be connected to electrodes. Connecting to Lead Wires Electrocardiograph Main Cable Connecting to Electrodes 1.
SE-1200 Series Electrocardiograph User Manual Operation Preparations 4.5 Attaching Electrodes to the Patient 4.5.1 Lead Placement 4.5.1.1 Standard 12-Lead Placement The identifiers and color codes of electrode connectors used comply with IEC/EN requirements. In order to avoid incorrect connection, the identifiers and color codes are specified in Table 4-1. Moreover the equivalent codes according to European requirements are given in Table 4-1 too.
SE-1200 Series Electrocardiograph User Manual Operation Preparations 4.5.1.2 NEHB Placement The electrodes’ positions on the body surface are shown in the following table and figure.
SE-1200 Series Electrocardiograph User Manual Operation Preparations 4.5.2 Electrode Placement Two kinds of electrode can be used, one is the reusable electrode, and the other is the disposable electrode. 4.5.2.1 Reusable Electrodes Reusable electrodes include limb electrodes and chest electrodes, as the following figures show.
SE-1200 Series Electrocardiograph User Manual Operation Preparations NOTE: Long-time measurement with a strong negative pressure on the suction bulb may cause reddening of the skin. When using the electrode on small children or patients with delicate skin, squeeze the suction ball lightly.
SE-1200 Series Electrocardiograph User Manual Operation Preparations Disposable Electrode Connection 1) Align all lead wires of the patient cable to avoid twisting, and connect the alligator clips to the lead wires. 2) Clean the electrode areas on the body surface with 75% alcohol. 3) Attach the disposable electrodes to the electrode positions on the body surface. 4) Clip the disposable electrodes with the alligator clips.
SE-1200 Series Electrocardiograph User Manual Operation Preparations 3) Patient Cable: ♦ Make sure that the patient cable is connected to the unit firmly, and keep it far away from the power cord. 4) Electrodes: ♦ Make sure that all electrodes are connected to lead wires of the patient cable correctly. ♦ Ensure that the chest electrodes do not contact. 5) Recorder Paper: ♦ Ensure that there is enough recorder paper loaded correctly.
SE-1200 Series Electrocardiograph User Manual Turning on the Electrocardiograph Chapter 5 Turning on the Electrocardiograph ♦ When the mains supply is used, connect the power cord, and the mains supply indicator ( ) is lit. Then press on the keyboard to turn on the unit. The logo will be displayed on the LCD screen after self-test. Then the electrocardiograph is ready for use. ♦ When battery is used, press battery indicator ( on the keyboard directly to turn on the unit, and then the ) is lit.
SE-1200 Series Electrocardiograph User Manual Entering Patient Information Chapter 6 Entering Patient Information After the electrocardiograph is turned on, the main screen appears. 6.1 Entering Patient ID 1. By default, the system generates the patient ID automatically. The range of the patient ID is 0 ~ 1999, 999, 999. 2. Or the system can generate the patient ID according to the time when you press the START/STOP key to print an ECG report.
SE-1200 Series Electrocardiograph User Manual Entering Patient Information 3) Press Tab or Shift + Tab to move the cursor to ID Mode, and then press the Up or Down arrow to select Manual. 4) Press press , or press Tab or Shift + Tab to move the cursor to the OK button, and then . 5) Press ESC to return to the main screen. 6) Press the function key F1 below Patient to open the Patient Information window. Press Tab or Shift + Tab to move the cursor.
SE-1200 Series Electrocardiograph User Manual Entering Patient Information First Name Within 30 ASCII characters Last Name Within 30 ASCII characters Age Age Unit: Years, Months, Weeks or Days Gender Patient Gender (Male/Female/Empty) Weight (kg) Patient Weight (Range 0~999) Height (cm) Patient Height (Range 0~999) Pacemaker If you select Pacemaker in the Patient Information Setup window, Pacemaker appears in the Patient Information window.
SE-1200 Series Electrocardiograph User Manual 3. After inputting all the information, press Entering Patient Information to confirm; or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. NOTE: If you select D.O.B in the Patient Information Setup window, the D.O.B textbox appears and the Age textbox becomes unavailable in the Patient Information window, you can enter the birthday of the patient, and the system will calculate the patient age automatically. 6.
SE-1200 Series Electrocardiograph User Manual Entering Patient Information 3) Enter the start and end addresses, the male and female codes, the vendor ID and the product ID, and then press to confirm. View the vendor ID and product ID on the PC: a) Connect the bar code reader to the PC b) Click on start, right-click on My Computer, and then select Manage in the pop-up menu to open the Computer Management window.
SE-1200 Series Electrocardiograph User Manual Entering Patient Information c) In the Computer Management window, click on Device Manager. Right-click on HID Keyboard Device, and then select Properties in the pop-up menu. d) Click on the Details tab in the HID Keyboard Device Properties window. View the vector ID (VID) and the product ID (PID). Take the following figure for example, the vector ID is 04B4, and the product ID is 0100.
SE-1200 Series Electrocardiograph User Manual Entering Patient Information 4) Press ESC to return to the main screen. 2. Connect the bar code reader to USB socket 2 on the right panel of the electrocardiograph. 3. When the main screen is displayed, scan the patient’s bar code with the bar code reader, and then the patient information will appear in the corresponding box. NOTE: Only bar code readers recommended by the manufacturer can be used.
SE-1200 Series Electrocardiograph User Manual Freezing ECG Waves Chapter 7 Freezing ECG Waves You can freeze the ECG waves displayed on the main screen. Operation Method: 1) Press the function key F2 below Freeze to display the freezing screen. NOTE: Within ten seconds after returning to the main screen, pressing the function key F2 can not display the freezing screen.
SE-1200 Series Electrocardiograph User Manual Freezing ECG Waves Figure 7-2 Replay screen 4) Press F1 below Pre Page or F4 below Next Page to turn pages. 5) Press F2 below Pre Sec or F3 below Next Sec to view the ECG waves of the previous or next second. 6) Press the function key F5 below Play to display ECG waves continuously. 7) Press the function key F6 below Pause to stop playing ECG waves. 8) Press the function key F7 below Return to return to the main screen.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports Chapter 8 Printing ECG Reports There are four modes to print ECG reports. In the auto mode, the lead groups are switched automatically according to the lead sequence during the printing course. After the ECG waves of one lead group are printed within a certain time, the system switches to print ECG waves of another lead group automatically. A 1mV calibration mark will be printed at the beginning of an ECG report.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports 2) Press Tab or Shift + Tab to move the cursor to the Auto DispStyle list box, and then press the Up or Down arrow to highlight a style. 3) Press Tab or Shift + Tab to move the cursor to the Sampling Mode list box, and then press the Up or Down arrow to highlight a sampling mode. For details of the sampling mode, refer to Section 11.1.5, “Selecting Sampling Mode”.
SE-1200 Series Electrocardiograph User Manual 8) Press Printing ECG Reports , or press Tab or Shift + Tab to move the cursor to the OK button, and then press . 9) On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Lead, press to open the Lead Setup window. 10) Press Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3 list box, and then press the Up or Down arrow to select a lead.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports After setup, press ESC to return to the main screen. 2. When the main screen is displayed, press the MODE key to select the auto mode. Press F3 to select a paper speed. Press F4 to switch the gain. Press F5 to set the EMG filter or the Lowpass filter. 3. Press the START/STOP key to print an ECG report. It will stop automatically after printing a complete ECG report of 12 leads. Or press the START/STOP key again to stop printing the report.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports 1) Press the function key F1 below Waveform to display the Waveform Preview Screen. 2) Press the function key F2 below Template to display the template preview screen. 3) Press the function key F3 below Measure to display the measure preview screen. 4) Press the function key F4 below Diagnosis to open the diagnosis preview screen. 5) Press the function key F5 below Continue to print an ECG report.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports 4) On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Lead, press to open the Lead Setup window. 5) Press Tab or Shift + Tab to move the cursor to the Lead Sequence list box, and then press the Up or Down arrow to select a sequence. 6) Press then press , or press Tab or Shift + Tab to move the cursor to the OK button, and .
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports 2. When the main screen is displayed, press the MODE key to select the manual mode. Press F3 to select a paper speed. Press F4 to switch the gain. Press F5 to set the EMG filter or the Lowpass filter. 3. Press the Up or Down arrow to select the lead group to be displayed and printed. 4. Press the START/STOP key to print an ECG report. 5. Press the START/STOP key to stop printing the ECG report. 8.3 Rhythm Mode Operation Method: 1.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports 4) On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Lead, press to open the Lead Setup window. 5) Press Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3 list box, and then press the Up or Down arrow to select a lead.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports After setup, press ESC to return to the main screen. 2. When the main screen is displayed, press the MODE key to select the rhythm mode. Press F3 to select a paper speed. Press F4 to switch the gain. Press F5 to set the EMG filter or the Lowpass filter. 3. Press the START/STOP key to begin to count the sampling time.
SE-1200 Series Electrocardiograph User Manual 2) Press then press Printing ECG Reports , or press Tab or Shift + Tab to move the cursor to the OK button, and . 3) On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Lead, press to open the Lead Setup window. 4) Press Tab or Shift + Tab to move the cursor to the Rhythm Lead1 list box, and then press the Up or Down arrow to select a lead.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports After setup, press ESC to return to the main screen. 2. When the main screen is displayed, press the MODE key to select the R-R analysis mode. Press F4 to switch the gain. Press F5 to set the EMG filter or the Lowpass filter. 3. Press the START/STOP key to begin to count the sampling time. When the sampling time reaches 180s, it begins to print an ECG report. 4.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports 8.6 ECG Reports 8.6.1 ECG Reports in the Auto Mode 8.6.1.1 Example of 6×2+1R Save Paper Mode The above figure shows an ECG report in the auto mode. The style is 6×2+1R. The ECG report includes: Patient Information, Measure Information, Diagnosis Information, Report Confirmed by Current Date and Current Time 6×2+1R ECG waves 0.67~100Hz (0.
SE-1200 Series Electrocardiograph User Manual Institution Name SE-1200 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports (b) The above figure (a) and (b) show an ECG report in the auto mode. Template is selected, and Record Style is set to 6×2+1R. Figure (a) shows: 02-06-2010 16:52:40 (Current Date & Current Time) ID: 001 (Patient ID) Name: Marry, Johnson (Patient Name) Age: 25 Years (Patient Age) (1mV calibration mark) І, II, Ш, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 (12 standard leads) and ECG waves II on the bottom (Rhythm Lead) and rhythm wave 0.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports 2*5s+1r (12 leads are printed in 2 groups of 6, and every group is printed for about 5s, with one rhythm lead at the bottom simultaneously) Institution Name SE-1200 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.5(Algorithm Version Number) Figure (b) shows: Patient Information, Measure Information, Diagnosis Information, Average Template, Report Confirmed by, Current Date and Current Time, 0.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports Average Template: Average template shows the average value of 10s sampled ECG signals of every lead. The broken lines on the template are position markers. They respectively mark the start and end points of the P and QRS waves, and the end point of the T wave. 8.6.1.2 Example of 3×4+1R The above figure shows an ECG report in the auto mode. The style is 3×4+1R.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports SE-1200 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.5(Algorithm Version Number) NOTE: Please refer to Section 8.6.1.1 for the explanation of the patient information, measure information and diagnosis information. 8.6.2 ECG Reports in the Rhythm Mode The above figure shows an ECG report in the rhythm mode, and Rhythm Style is set to Three Leads.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports 20s rhythm waveform of lead II / V1 / V5 0.67~100Hz (0.67Hz DFT Filter, 100Hz Lowpass Filter) AC50 (50Hz AC Filter) 25mm/s (Paper Speed) 10mm/mV (Gain) Institution Name SE-1200 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.5(Algorithm Version Number) 8.6.3 ECG Reports in the Manual Mode The above figure shows an ECG report in the manual mode, and Manual Style is set to 12 channels.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports ECG waveform of 12 standard leads 0.67~100Hz (0.67Hz DFT Filter, 100Hz Lowpass Filter) AC50 (50Hz AC Filter) 25mm/s (Paper Speed) 60 (Heart Rate) 10mm/mV (Gain) Institution Name SE-1200 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.5(Algorithm Version Number) 8.6.
SE-1200 Series Electrocardiograph User Manual (b) - 75 - Printing ECG Reports
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports (c) The above figure (a), (b) and (c) show an ECG report in the R-R analysis mode.
SE-1200 Series Electrocardiograph User Manual Figure (b) shows: Current Date & Current Time Patient Information (Name, ID, Age) Gain, Speed, Filter (1mV calibration mark) II (Lead name) 180s compressed ECG wave of Lead II Figure (c) shows: All the RR interval values within the measuring time - 77 - Printing ECG Reports
SE-1200 Series Electrocardiograph User Manual 8.6.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports As the above figure shows, the ECG report printed by the USB printer includes: Institution Name Date and Time Patient Information Measurement Information Diagnosis Information; Report Confirmed by; ECG waves of 12 leads DFT Filter, Lowpass Filter, AC Filter, Paper Speed, Gain; SE-1200 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.5(Algorithm Version Number) 8.6.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports NOTE: 1. To the patient with a pacemaker, the EMG filter should be set to Off and the Lowpass filter should be set to 150Hz; or else, the amplitude of the pulses may decrease, the width may increase, and the diagnosis results may not include the pacemaker information. 2. To the patient with a pacemaker, the results given by the equipment may be invalid. 8.6.
SE-1200 Series Electrocardiograph User Manual Printing ECG Reports 0.67~100Hz (0.67Hz DFT Filter, 100Hz Lowpass Filter) AC50 (50Hz AC Filter) 10mm/mV (Gain) 25mm/s (Paper Speed) 60 (Heart Rate) Institution Name SE-1200 Express (Electrocardiograph Model) V1.8 (Software Version Number) SEMIP V1.
SE-1200 Series Electrocardiograph User Manual Chapter 9 Transmitting ECG Data Automatically Transmitting ECG Data Automatically 9.1 Transmitting ECG Data to the PC NOTE: 1. To transmit ECG data in DAT format to the PC, the Smart ECG Viewer software of the manufacturer must be installed in the PC. You should log into the Smart ECG Viewer software before transmission. 2. To transmit ECG data in SCP/FDA-XML/PDF format to the PC, the FTP receiving software must be installed in the PC.
SE-1200 Series Electrocardiograph User Manual Transmitting ECG Data Automatically Figure 9-1 RS232-USB Assembly 3. Set Auto Transmission to On and Transmission Mode to UART in the Transmission Setup window. 1) When the main screen is displayed, press the function key F6 below Setup to open the System Setup screen. 2) Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor to to open the Transmission Setup window.
SE-1200 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 5. In the auto or rhythm mode, ECG data will be transmitted through the serial cable automatically after an ECG report is printed out. In the off mode, ECG data will be transmitted through the serial cable automatically after the START/STOP key is pressed. 9.1.2 Transmitting ECG Data in DAT Format through Net Cable 1.
SE-1200 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 6) Press Tab or Shift + Tab to move the cursor to the Local IP item. For the cross-network transmission, set the first two sections of the Local IP item to the first two sections of the IP of the PC. Set the third section of the Local IP item to the network segment of the electrocardiograph which depends on the configuration of Router. The last section of the Local IP item can be set at random.
SE-1200 Series Electrocardiograph User Manual Transmitting ECG Data Automatically In the Transmission Setup window of the electrocardiograph, set Transmission Mode to Ethernet. Set the Server IP item to the IP of the PC. Set the first three sections of the Local IP item to the first three sections of the IP of the PC. The last section of the Local IP item is not the same as the last section of the IP of the PC.
SE-1200 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 9.1.3 Transmitting ECG Data in SCP/FDA-XML/PDF Format through Net Cable 1. Activate SCP/FDA-XML/PDF function on the Advanced Setup screen 1) When the main screen is displayed, press the function key F6 below Setup to open the System Setup screen. 2) Press F1 to display the Service Password window.
SE-1200 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 3) Press Tab or Shift + Tab to move the cursor to the Activate button after the function to be activated. Press password and press to open the Activate Password window. Enter the to activate the function. NOTE: Please contact local distributors to get the service password and the activating password. 2.
SE-1200 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 3) Log into the FTP receiving software. The FTP port is 21. 4) In the auto or rhythm mode, ECG data will be transmitted through the net cable automatically after an ECG report is printed out. In the off mode, ECG data will be transmitted through the net cable automatically after the START/STOP key is pressed. 9.2 Exporting ECG Data to the U Disk Automatically 1.
SE-1200 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 4. In the auto or rhythm mode, ECG data will be automatically exported to the directory of ECGDATA\ECG-X\Store\Examination Date of the U disk after an ECG report is printed out. In the off mode, ECG data will be automatically exported to the directory of ECGDATA\ECG-X\Store\Examination Date of the U disk after the START/STOP key is pressed. NOTE: X can be set in the Device No. textbox in the Transmission Setup window. 9.
SE-1200 Series Electrocardiograph User Manual Transmitting ECG Data Automatically 2. Connect USB socket 1 of the electrocardiograph to the USB socket of the PC by using the high-speed USB cable 3. Start the real-time transmission The electrocardiograph transmits to the PC the ECG signals acquired from the patient. Acquisition and transmission are simultaneous. The ECG signals are displayed on the PC monitor and eventually analyzed. For more details, refer to the user manual of the PC ECG software.
SE-1200 Series Electrocardiograph User Manual Managing Files Chapter 10 Managing Files If you want to save the ECG data in the electrocardiograph, you should set Auto Save to On and Saving Path to ECG. By default, Auto Save is On and Saving Path is ECG. Then the ECG data in the auto, off or rhythm mode will be saved on the File Manager screen automatically. To Set Auto Save to On and Saving Path to ECG 1.
SE-1200 Series Electrocardiograph User Manual Managing Files Figure 10-1 File Manager screen1 Figure 10-2 File Manager screen2 On the File Manager screen, files can be printed, transmitted, exported, displayed, edited searched or deleted. SE-1200 can accommodate 100 files, and SE-1200 Express can accommodate 200 files. If there is no file on the File Manager screen, the following dialog box will pop up when you press function keys.
SE-1200 Series Electrocardiograph User Manual Managing Files CAUTION 1. When files are being printed, transmitted, deleted or exported, you can not turn off the electrocardiograph. 2. Do not cut off the mains supply directly when no battery is installed in the device, or else, the stored data may be lost. 10.1 Transmitting Files to the PC NOTE: 1. To transmit DAT files to the PC, the Smart ECG Viewer software of the manufacturer must be installed in the PC.
SE-1200 Series Electrocardiograph User Manual Managing Files 10.1.1 Transmitting DAT Files through Serial Cable 1. Connect the RS232 socket of the PC to the RS232 socket of the electrocardiograph with an RS232 cable. 2. Or if the PC has no RS232 socket, connect the USB socket of the PC to the RS232 socket of the electrocardiograph by using the RS232-USB assembly. For details of the RS232-USB assembly, please refer to Section 9.1.1, “Transmitting ECG Data in DAT Format Through Serial Cable”. 3.
SE-1200 Series Electrocardiograph User Manual Managing Files 10.1.2 Transmitting DAT Files through Net Cable 1. Connect the electrocardiograph to the network of the PC with an Ethernet cable recommended by the manufacturer. If the wireless AP transmission is used, connect the PC to the server wireless AP, and connect the electrocardiograph to the client wireless AP. Only the wireless AP recommended by the manufacturer can be used. NOTE: 1.
SE-1200 Series Electrocardiograph User Manual Managing Files 6. Or press the Up or Down arrow to highlight a file on the File Manager screen1, and then press the function key F4 below Select to select the file and display the File Manager screen2. Press the function key F3 below Trans to transmit the selected file to the PC. 10.1.3 Transmitting SCP/FDA-XML/PDF Files through Net Cable 1. Activate SCP/FDA-XML/PDF function on the Advanced Setup screen For details, refer to Section 9.1.
SE-1200 Series Electrocardiograph User Manual Managing Files If the wireless AP transmission is used, connect the PC to the server wireless AP, and connect the electrocardiograph to the client wireless AP. Only the wireless AP recommended by the manufacturer can be used. 2) Log into the FTP receiving software. The FTP port is 21. 3) Input the correct FTP path, FTP user name and password. Set Transmission Mode to Ethernet. Set the IP addresses. For details, please see Section 9.1.
SE-1200 Series Electrocardiograph User Manual Managing Files If the U disk is not connected well, the hint U disk is not ready! will be displayed on the LCD screen of the electrocardiograph, and then you should connect the U disk again. If the electrocardiograph fails to export files, the hint Exporting fails! will pop up. After the electrocardiograph successfully exports files, the hint Exporting succeeds! will be displayed. NOTE: 1.
SE-1200 Series Electrocardiograph User Manual Managing Files 10.4 Searching Files To Open the SearchInfo Setup Window 1. When the main screen is displayed, press the function key F7 below File to open the File Manager screen1. 2. Press the function key F5 below Search to open the SearchInfo Setup window. To Search Files 1. Press Tab, Shift + Tab, or the Up or Down arrow to select Name, and then input the patient name. Press , or press Tab or Shift + Tab to move the cursor to the OK button, and .
SE-1200 Series Electrocardiograph User Manual Managing Files 3. Press Tab, Shift + Tab, or the Up or Down arrow to select Time, and then input the examination time range. Press , or press Tab or Shift + Tab to move the cursor to the OK button, and then press . Then all the corresponding files from the start time to the end time will be displayed on the File Manager screen1. NOTE: The time mode is year-month-day in the SearchInfo Setup window. 10.5 Importing Files 1.
SE-1200 Series Electrocardiograph User Manual Managing Files NOTE: The Address item is customized. 1. Press Tab or Shift + Tab to move the cursor to a textbox. Fill in the selected textbox. 2. Press Tab or Shift + Tab to move the cursor to a list box. Press the Up or Down arrow to select an option. 3. After inputting all the information, press to confirm; or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm.
SE-1200 Series Electrocardiograph User Manual Managing Files To Preview the File 1. Press the function key F1 below Waveform to display the waveform screen. 2. Press the function key F2 below Template to display the template screen. 3. Press the function key F3 below Measure to display the measure screen which shows Patient Name, Patient ID, Heart Rate, P Duration, PR interval, QRS Duration, QT/QTC interval, P/QRS/T axis, RV5/SV1 amplitude, RV5+SV1 amplitude, and RV6/SV2 amplitude.
SE-1200 Series Electrocardiograph User Manual Managing Files 4. Press the function key F4 below Diagnosis to show Minnesota Code and Diagnosis Information. 5. Press START/STOP or the function key F5 below Record to print the file. Pressing START/STOP or the function key F5 below Stop can stop printing the file. 6. Press the function key F7 below Return to return to the File Manager screen2. NOTE: Rhythm data can not be previewed.
SE-1200 Series Electrocardiograph User Manual System Setup Chapter 11 System Setup After you turn on the device, the main screen pops up. Press the function key F6 below Setup to display the System Setup screen. Figure 11-1 System Setup screen After setup, 1. Press ESC to return to the main screen. 2. Or press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor to Return, and then press to return to the main screen. 11.
SE-1200 Series Electrocardiograph User Manual System Setup Figure 11-2 Work Mode Setup window After setup, press OK button, and then press to confirm, or press Tab or Shift + Tab to move the cursor to the to confirm. Then the System Setup screen appears. 11.1.1 Selecting Work Mode Press Tab or Shift + Tab to move the cursor to a work mode, press Spacebar to select it, and then a check mark √ appears in the box before the work mode.
SE-1200 Series Electrocardiograph User Manual System Setup R-R: In the R-R analysis mode, you can select a lead to print its R-R histogram, R-R trend chart, 180s compressed ECG waveform and all the R-R interval values. OFF: In the off mode, the lead groups are switched automatically according to the lead sequence. When the main screen is displayed, after pressing the START/STOP key, the sampled ECG data can be saved and transmitted, but can not be printed. 11.1.
SE-1200 Series Electrocardiograph User Manual System Setup When the auto display style is set to 3×4+1R, ECG waves of 12 leads are displayed in 4 groups of 3, with the ECG wave of one rhythm lead on the bottom. When the auto display style is set to 3×4+3R, ECG waves of 12 leads are displayed in 4 groups of 3, with ECG waves of three rhythm leads on the bottom. When the auto display style is set to 6×2, ECG waves of 12 leads are displayed in 2 groups of 6.
SE-1200 Series Electrocardiograph User Manual System Setup 11.1.6 Setting Interval and Duration Press Tab or Shift + Tab to move the cursor to the Duration and Interval textboxes in the Work Mode Setup window, and then input the duration and the interval manually. In the auto mode, when Sampling Mode is set to Periodic Sample, if Interval is set to 2 min, Duration is set to 24 min, after pressing the START/STOP key, the printing will be performed every two minutes and come to 12 times. 11.1.
SE-1200 Series Electrocardiograph User Manual System Setup Figure 11-3 Filter Setup window After setup, press to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. Then the System Setup screen appears. 11.2.1 Setting AC Filter Press Tab or Shift + Tab to move the cursor to the AC Filter list box in the Filter Setup window. Press the Up or Down arrow to highlight Off or On.
SE-1200 Series Electrocardiograph User Manual System Setup DFT Filter greatly reduces the baseline fluctuations without affecting the ECG signals. The purpose of this filter is to keep the ECG signals on the baseline of the printout. The set value is the low limit of the frequency range, including 0.05Hz, 0.15Hz, 0.25Hz, 0.32Hz, 0.5Hz and 0.67Hz. 11.2.3 Setting EMG Filter Press Tab or Shift + Tab to move the cursor to the EMG Filter list box in the Filter Setup window.
SE-1200 Series Electrocardiograph User Manual System Setup 11.3 Record Info Setup Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Record Info on the System Setup screen, press to open the Record Info Setup window. You can press F1 or F2 to switch between Setup 1 and Setup 2. After setup, press OK button, and then press to confirm, or press Tab or Shift + Tab to move the cursor to the to confirm. Then the System Setup screen appears. 11.3.
SE-1200 Series Electrocardiograph User Manual System Setup In the auto mode, when Record Style is set to 3×4, ECG waves of 12 leads are printed in 4 groups of 3. In the auto mode, when Record Style is set to 3×4+1R, ECG waves of 12 leads are printed in 4 groups of 3, with the ECG wave of one rhythm lead on the bottom of the ECG reports.
SE-1200 Series Electrocardiograph User Manual System Setup When Record Sequence is set to Synchronous, all leads are printed simultaneously. The start time of each group is the same. 11.3.1.4 Setting Gain Press Tab or Shift + Tab to move the cursor to the Gain list box on the Setup 1 tab, and then press the Up or Down arrow to highlight an option. You can set the indicated height of 1mV ECG on the paper. You can set Gain to 10mm/mV, 20mm/mV, 10/5mm/mV, AGC, 2.5mm/mV or 5mm/mV.
SE-1200 Series Electrocardiograph User Manual Options HP-DJ-D2468 HP-DJ-D5568 HP-LJ-P2035 System Setup Corresponding USB Printers HP Deskjet 3938 HP Deskjet D2368 HP Deskjet D2468 HP Deskjet D2568 HP Deskjet D5568 HP Laserjet P2015 HP Laserjet P2035 NOTE: 1. During the USB printing course,pressing the START/STOP key again can not stop printing ECG reports. 2. For details of the ECG report printed by the USB printer, please refer to Section 8.6.5, “ECG Reports Printed by the USB Printer”. 3.
SE-1200 Series Electrocardiograph User Manual System Setup When the paper with black markers on the bottom is used and Paper Marker is set to Yes, the device can identify the start point of each page of the recorder paper while printing ECG reports. When Paper Marker is set to No, the device can not identify the start point of each page of the recorder paper while printing ECG reports. 11.3.1.8 Setting Sampling Time If Record Style is set to 12x1, you can set the time period.
SE-1200 Series Electrocardiograph User Manual System Setup Auto Record Info includes Patient Info, Position Marker, Analysis, Template, Measure, Minnesota Code and Baseline Adjustment. When Patient Info, Position Marker, Analysis, Template, Measure or Minnesota Code is selected, it will be printed in the ECG reports. When Baseline Adjustment is set to Auto, the baselines of the lead groups are adjusted respectively.
SE-1200 Series Electrocardiograph User Manual System Setup Figure 11-6 Patient Information Setup window After setup, press to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. Then the System Setup screen appears. 11.4.1 Selecting First/Last Name / Gender / D.O.B / Height / Weight / BP / Race / Pacemaker / Medication / Room No.
SE-1200 Series Electrocardiograph User Manual System Setup You can set ID Mode to Auto, Time or Manual. When ID Mode is set to Auto, the patient ID can be automatically generated. The patient ID range is 0~1999, 999, 999. When ID Mode is set to Time, the patient ID can be automatically generated according to the time when you press the START/STOP key to print an ECG report. When ID Mode is set to Manual, you can enter the patient ID manually in the Patient Information window.
SE-1200 Series Electrocardiograph User Manual System Setup When BP Unit is set to kpa, two extra edit boxes will be displayed in the Patient Information window for inputting decimal fraction. 11.4.7 Selecting Next Patient Press Tab or Shift + Tab to move the cursor to Next Patient in the Patient Information Setup window. Press the Up or Down arrow to highlight Off or On.
SE-1200 Series Electrocardiograph User Manual System Setup 11.4.8 Setting the Time to Clear Patient Information Press Tab or Shift + Tab to move the cursor to Patient Info Cleared in the Patient Information Setup window. Press the Up or Down arrow to highlight an option. If Patient Info Cleared is set to After Recording, the patient information will be cleared after the ECG report is printed out.
SE-1200 Series Electrocardiograph User Manual System Setup Figure 11-7 Transmission Setup window After setup, press to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. Then the System Setup screen appears. 11.5.1 Setting Device No. Press Tab or Shift + Tab to move the cursor to the Device No. textbox, and then fill in the textbox. If you input 1 in the Device No.
SE-1200 Series Electrocardiograph User Manual System Setup 11.5.2 Selecting Auto Transmission Press Tab or Shift + Tab to move the cursor to Auto Transmission, and then press the Up or Down arrow to select On or Off. When Auto Transmission is set to On, in the auto or rhythm mode, ECG data will be transmitted automatically after an ECG report is printed out; in the off mode, ECG data will be transmitted automatically after the START/STOP key is pressed. 11.5.
SE-1200 Series Electrocardiograph User Manual System Setup To Set Local IP For the cross-network transmission, set the first two sections of the Local IP item to the first two sections of the IP of the PC. Set the third section of the Local IP item to the network segment of the electrocardiograph which depends on the configuration of Router. The last section of the Local IP item can be set at random.
SE-1200 Series Electrocardiograph User Manual System Setup Figure 11-8 Lead Setup window After setup, press OK button, and then press to confirm, or press Tab or Shift + Tab to move the cursor to the to confirm. Then the System Setup screen appears. 11.6.1 Setting Rhythm Lead1/2/3 Press Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3 list box in the Lead Setup window, and then press the Up or Down arrow to highlight an option.
SE-1200 Series Electrocardiograph User Manual System Setup In the R-R analysis mode, the R-R analysis report of the rhythm lead selected in the Rhythm Lead1 list box will be printed. 11.6.2 Setting Lead Sequence Press Tab or Shift + Tab to move the cursor to the Lead Sequence list box in the Lead Setup window, and then press the Up or Down arrow to highlight Standard or Cabrera.
SE-1200 Series Electrocardiograph User Manual System Setup Figure 11-9 Display&Sound Setup window 11.7.1 Setting Brightness (Optional) Press Tab or Shift + Tab to move the cursor to the Brightness progress bar in the Display&Sound Setup window, and then press the Left or Right arrow to adjust the brightness of the LCD screen. You can set the brightness within 0~20. 11.7.
SE-1200 Series Electrocardiograph User Manual System Setup 11.7.4 Setting Key Volume Press Tab or Shift + Tab to move the cursor to the Key Volume list box in the Display&Sound Setup window, and then press the Up or Down arrow to highlight an option. When you press keys on the keyboard, the electrocardiograph gives a short sound. You can set Key Volume to Low, Medium, High or Off. When it is set to Off, there is no sound when you press keys. 11.7.
SE-1200 Series Electrocardiograph User Manual System Setup Figure 11-10 Date&Time Setup window After setup, press to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. Then the System Setup screen appears. 11.8.1 Setting Date and Time Press Tab or Shift + Tab to move the cursor to the Date or Time textbox in the Date&Time Setup window, and then input the date or the time manually.
SE-1200 Series Electrocardiograph User Manual System Setup 11.8.3 Setting Time Mode Press Tab or Shift + Tab to move the cursor to the Time Mode list box, and then press the Up or Down arrow to highlight 24 Hours or 12 Hours. 11.8.4 Setting Power-Off Time Press Tab or Shift + Tab to move the cursor to the Power Off textbox in the Date&Time Setup window, and then input the power-off time manually. When Power Off is not set or is set to 0 Minutes, this function will not be effective. NOTE: 1.
SE-1200 Series Electrocardiograph User Manual System Setup Figure 11-11 File Setup window After setup, press to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press to confirm. Then the System Setup screen appears. 11.9.1 Selecting Auto Save Press Tab or Shift + Tab to move the cursor to the Auto Save list box in the File Setup window, and then press the Up or Down arrow to highlight Off or On.
SE-1200 Series Electrocardiograph User Manual System Setup 11.9.3 Setting File Format Press Tab or Shift + Tab to move the cursor to File Format in the File Setup window, and then press the Up or Down arrow to highlight an option. If you want to set File Format to SCP/FDA-XML/PDF, first activate the SCP/FDA-XML/PDF function on the Advanced Setup screen For details on activating the SCP/FDA-XML/PDF function, please refer to Section 9.1.3, “Transmitting ECG Data in SCP/FDA-XML/PDF Format Through Net Cable”.
SE-1200 Series Electrocardiograph User Manual System Setup 11.10 System Maintenance Setup Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on Maintenance on the System Setup screen, press to open the System Maintenance window. Figure 11-12 System Maintenance window After setup, press OK button, and then press to confirm, or press Tab or Shift + Tab to move the cursor to the to confirm. Then the System Setup screen appears. 11.10.
SE-1200 Series Electrocardiograph User Manual System Setup 11.10.2 Loading System Settings from U Disk 1. Connect the U disk containing the system settings to the electrocardiograph. 2. Press Tab or Shift + Tab to move the cursor to the Load Settings From U disk button in the System Maintenance window, and then press to load the system settings from the U disk to the electrocardiograph. 11.10.
SE-1200 Series Electrocardiograph User Manual System Setup Table 11-1 System Setup Defaults Setup Default 1 Mode Options Auto, Manual, Rhythm 2 Manual Style 6 channels 3 Auto Display Style 6×2 4 Rhythm Style Three Leads 5 Sampling Mode Real-time Sample 6 Duration (Periodic Sample) 60 min 7 Interval (Periodic Sample) 1 min 8 Preview Off 9 Auto Extend Record Off 10 AC filter On 11 EMG filter Off 12 DFT filter 0.
SE-1200 Series Electrocardiograph User Manual System Setup 27 Minnesota Code Off 28 Baseline Adjustment Horizontal 29 R-R Histogram On 30 R-R Interval List On 31 R-R Waveform On 32 Grid of Thermal Report Off 33 Grid of USB Report On 34 Gender\Height\Weight On 35 First/Last Name \ D.O.B \ BP \ Race \ Pacemaker \ Medication \ Room No. \ Department \ Physician \ Technician \ Ref-Physician Off 36 ID Hint On 37 Next Patient Off 38 ID Mode Auto 39 Patient Information Cleared
SE-1200 Series Electrocardiograph User Manual System Setup 53 Antialiasing Off 54 QRS Volume Off 55 Hint Volume Medium 56 Key Volume Medium 57 Date Mode DD-MM-YYYY 58 Time Mode 24 Hours 59 Power Off 0 60 LCD Off 0 61 Auto Save On 62 Saving Path ECG 63 File Format DAT 64 Delete After Trans.
SE-1200 Series Electrocardiograph User Manual System Setup 10 AC filter On 11 EMG filter Off 12 DFT filter 0.
SE-1200 Series Electrocardiograph User Manual System Setup 37 Next Patient Off 38 ID Mode Manual 39 Patient Information Cleared After Recording 40 H/W Unit inch/lb 41 BP Unit mmHg 42 Prompt Confirmed By 43 User Defined Cleared 44 Auto Transmission Off 45 Transmission Mode Ethernet 46 FTP User Name/FTP Password/FTP Path Cleared 47 Lead Sequence Standard 48 Nehb Off 49 Rhythm Lead 1 II 50 Rhythm Lead 2 V1 51 Rhythm Lead 3 V5 52 Screen Colors(Only for SE-1200 Exp
SE-1200 Series Electrocardiograph User Manual System Setup 65 Replace When Memory Full Off 66 SCP File Compression On 67 System Password Cleared 68 Caps Lock Off 69 Pacemaker Detection Sensitivity Low 11.10.7 Setting System Password Press Tab or Shift + Tab to move the cursor to the System Password textbox in the System Maintenance window, and then enter the password. If you set the system password, after you press the function key F6 below Setup, the System Password window will pop up.
SE-1200 Series Electrocardiograph User Manual System Setup 11.11.1 Entering Institution Press Tab or Shift + Tab to move the cursor to the Institution textbox on the Other Setup Window, and then manually input the institution name within 40 ASCII characters. NOTE: The total number of supported characters may be fewer if either special Latin characters or Chinese characters are entered. 11.11.
SE-1200 Series Electrocardiograph User Manual System Setup 11.11.5 Setting External Output Press Tab or Shift + Tab to move the cursor to the External Output list box on the Other Setup Window, and then press the Up or Down arrow to highlight an option. The extern output socket is equipped in the electrocardiograph, through which the electrocardiograph can send signals to the external equipment.
SE-1200 Series Electrocardiograph User Manual Turning off the Electrocardiograph Chapter 12 Turning off the Electrocardiograph When the battery is used, hold down the key to display the hint System is shutting down… on the screen. Then the device will be off a few seconds later. When the mains supply is used, hold down the key to display the hint System is shutting down… on the screen. Then the device will be off a few seconds later. Remove the plug from the outlet. NOTE: 1.
SE-1200 Series Electrocardiograph User Manual Hint Information Chapter 13 Hint Information Hint information and the corresponding causes provided by the electrocardiograph are listed in Table 13-1. Table 13-1 Hint Information and Causes Hint Information Causes Lead off Electrodes fall off the patient or the patient cable falls off the unit, or a high polarization voltage occurs. Battery Weak The battery is weak. No Paper Recorder paper runs out or is not loaded.
SE-1200 Series Electrocardiograph User Manual Cleaning, Care and Maintenance Chapter 14 Cleaning, Care and Maintenance CAUTION 1. Turn off the power before cleaning and disinfection. The mains supply must be switched off if it is used. 2. Prevent the detergent from seeping into the equipment. 14.1 Cleaning CAUTION Any remainder of detergent should be removed from the unit and the patient cable after cleaning. 14.1.1 Cleaning the Main Unit and the Patient Cable To clean the electrocardiograph: 1.
SE-1200 Series Electrocardiograph User Manual Cleaning, Care and Maintenance 3. Clean them in warm water and make sure there is no remainder gel. 4. Dry the electrodes with a clean dry cloth or air dry naturally. CAUTION Do not clean the unit and accessories with abrasive fabric and avoid scratching the electrodes. 14.1.3 Cleaning the Print Head To clean the print head: 1. Open the recorder casing and remove the paper. 2. Wipe the print head gently with a clean soft cloth damped in 75% alcohol.
SE-1200 Series Electrocardiograph User Manual Cleaning, Care and Maintenance 14.3 Care and Maintenance CAUTION Operate the electrocardiograph, charge the battery, and store the battery at a temperature of 40℃(104ºF) or lower. Exposure to higher or lower temperature may reduce battery life, damage the battery, and degrade overall electrocardiograph performance. 14.3.
SE-1200 Series Electrocardiograph User Manual Cleaning, Care and Maintenance WARNING 1. Only qualified service engineers authorized by the manufacturer can open the battery compartment and replace the battery, and the battery of the same model and specification provided by the manufacturer must be used. 2. Danger of explosion -- Do not reverse the anode and the cathode when installing the battery. 3. Remove the battery from the electrocardiograph when the electrocardiograph is not used for a long time. 4.
SE-1200 Series Electrocardiograph User Manual Cleaning, Care and Maintenance 14.3.3 Visual inspection Perform a visual inspection of all equipment and peripheral devices daily. If you notice any items that need repair, contact a qualified service engineer to make the repairs. ♦ Check the case and display screen for cracks or other damage. ♦ Regularly inspect all plugs, cords, cables, and connectors for fraying or other damage. ♦ Verify that all cords and connectors are securely seated.
SE-1200 Series Electrocardiograph User Manual Cleaning, Care and Maintenance i) Test the patient auxiliary current according to IEC/EN 60601-1: Limit: NC a.c. 10μA, d.c. 10μA; SFC a.c. 50μA, d.c. 50μA. j) Test the patient leakage current under single fault condition with mains voltage on the applied part according to IEC/EN 60601-1: Limit: 50μA (CF). The leakage current should never exceed the limit. The data should be recorded in an equipment log.
SE-1200 Series Electrocardiograph User Manual Cleaning, Care and Maintenance 3) Reusable Electrodes ♦ Electrodes must be cleansed after use and make sure there is no remainder gel on them. ♦ Keep suction bulbs of chest electrodes away from sunshine and excessive temperature. ♦ After long-term use, the surfaces of electrodes will be oxidized because of erosion and other causes. By this time, electrodes should be replaced to achieve high-quality ECG records.
SE-1200 Series Electrocardiograph User Manual Accessories Chapter 15 Accessory List WARNING Only the patient cable and other accessories supplied by the manufacturer can be used. Or else, the performance and electric shock protection can not be guaranteed. Accessory Part Number 01.57.107581 (Snap Style) ECG Cable (European) 01.57.107583 (Grabber Style) 01.57.107582 (Snap Style) ECG Cable (American) 01.57.107584 (Grabber Style) Patient Cable (American) 01.57.107048 Patient Cable (European) 01.57.
SE-1200 Series Electrocardiograph User Manual Accessories Clip/Snap/Banana Socket Adapter 01.57.040172 Alligator Clip/Banana Socket Adapters 01.57.040173 Paper Roller 01.51.30340-01 Thermal Recorder Paper 01.57.107371 Fuse 11.21.64073 Input/Output Connector 11.13.19907 Grounding Wire 11.13.114214 03.24.38952 Smart ECG Viewer Software 03.24.38953 Snap Clip 01.57.040172 Electrode Gel 11.25.78047 RS232 Cable 11.13.20117 Ethernet Cable 11.13.
SE-1200 Series Electrocardiograph User Manual Warranty & Service Chapter 16 Warranty & Service 16.1 Warranty EDAN warrants that EDAN’s products meet the labeled specifications of the products and will be free from defects in materials and workmanship that occur within warranty period. The warranty is void in cases of: a) Damage caused by mishandling during shipping. b) Subsequent damage caused by improper use or maintenance. c) Damage caused by alteration or repair by anyone not authorized by EDAN.
SE-1200 Series Electrocardiograph User Manual Technical Specifications Appendix 1 Technical Specifications A1.1 Safety Specifications Comply with: IEC/EN 60601-1+A1+A2, IEC/EN 60601-1-2+A1, IEC/EN 60601-2-25, ANSI/AAMI EC11, IEC/EN 60601-2-51, UL 60601-1, CAN/CSA C22.2 No.601.
SE-1200 Series Electrocardiograph User Manual Technical Specifications A1.2 Environment Specifications Temperature: Transport & Storage Working -20℃ (-4ºF) ~ +55℃ (+131ºF) +5ºC (+41ºF) ~ +40ºC (+104ºF) 25%~93% 25%~80% Non-Condensing Non-Condensing 700hPa ~1060hPa 860hPa ~1060hPa Relative Humidity: Atmospheric Pressure: A1.3 Physical Specifications Dimensions Weight Display 420mm×330mm×105mm (16.5in×13.0in×4.1in) Approx. 5.0kg (11.
SE-1200 Series Electrocardiograph User Manual Technical Specifications Charge mode: Constant current/voltage SE-1200 Express: Charge current (standard) = 0.14 C5A (600mA) SE-1200: Charge current (standard) = 0.28 C5A (600mA) Charge voltage (standard) = (16.8-0.1V) Necessary Charge time: SE-1200 Express: 9.5 hours SE-1200: 5 hours Cycle life ≥ 300 times Fuse: T1AL250VP Ø5×20 A1.
SE-1200 Series Electrocardiograph User Manual Acquisition Mode: simultaneously 12 leads A/D Resolution: 24 bits Time Constant: ≥3.2s Frequency Response: 0.05Hz ~ 150Hz (-3dB) Gain: 2.5, 5, 10, 20, 10/5, AGC (mm/mV) Input Impedance: ≥50MΩ (10Hz) Input Circuit Current: ≤0.01μA Input Voltage Range ≤±5 mVpp Calibration Voltage: 1mV±2% DC Offset Voltage: ±600mV Noise: ≤12.5 μVp-p Multichannel crosstalk ≤0.5mm Technical Specifications AC Filter: On/Off Filter DFT Filter: 0.05Hz/0.
SE-1200 Series Electrocardiograph User Manual Technical Specifications NOTE: 1. Test the accuracy of input signal reproduction according to the methods described in clause 4.2.7.2 in ANSI/AAMI EC11:1991/(R) 2001, and the result complies with the clause 3.2.7.2 in ANSI/AAMI EC11:1991/(R) 2001. 2. Overall System Error is tested using the method described in AAMI EC11 3.2.7.1. Overall System Error is not more than ±5%. 3. Frequency Response is tested using the method described in AAMI EC11 3.2.7.
SE-1200 Series Electrocardiograph User Manual EMC Information Appendix 2 EMC Information Guidance and manufacture’s declaration - electromagnetic emissions - for all EQUIPMENT and SYSTEMS Guidance and manufacture’s declaration - electromagnetic emission The 12-channel electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or the user of the 12-channel electrocardiograph should assure that it is used in such an environment.
SE-1200 Series Electrocardiograph User Manual EMC Information Guidance and manufacture’s declaration - electromagnetic immunity - for all EQUIPMENT and SYSTEMS Guidance and manufacture’s declaration - electromagnetic immunity The 12-channel electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or the user of 12-channel electrocardiograph should assure that it is used in such an environment.
SE-1200 Series Electrocardiograph User Manual NOTE EMC Information (>95% dip in UT) (>95% dip in UT) supply or a battery. for 5 sec for 5 sec UT is the a.c. mains voltage prior to application of the test level. Guidance and manufacture’s declaration - electromagnetic immunity - for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING Guidance and manufacture’s declaration - electromagnetic immunity The 12-channel electrocardiograph is intended for use in the electromagnetic environment specified below.
SE-1200 Series Electrocardiograph User Manual EMC Information NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
SE-1200 Series Electrocardiograph User Manual EMC Information Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM - for EQUIPMENT or SYSTEM that is not LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and the 12-channel Electrocardiograph The 12-channel electrocardiograph is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
SE-1200 Series Electrocardiograph User Manual Appendix 3 Abbreviation Abbr English LCD Liquid Crystal Display BP Blood Pressure ECG Electrocardiogram/Electrocardiograph HR Heart Rate aVF Left Foot Augmented Lead aVL Left Arm Augmented Lead aVR Right Arm Augmented Lead LA Left Arm LL Left Leg RA Right Arm RL Right Leg ID Identification AC Alternating Current USB Universal Serial Bus AGC Auto Gain Control NC Normal Condition SFC Single Fault Condition - 165 - Abbreviation