About this Manual P/N: 01.54.455420-10 Release Date: November 2011 © Copyright EDAN INSTRUMENTS, INC. 2011. All rights reserved. Statement This manual will help you understand the operation and maintenance of the product better. It is reminded that the product shall be used strictly complying with this manual. User’s operation failing to comply with this manual may result in malfunction or accident for which EDAN INSTRUMENTS, INC. (hereinafter called EDAN) can not be held liable.
Terms Used in this Manual This guide is designed to give key concepts on safety precautions. WARNING A WARNING label advises against certain actions or situations that could result in personal injury or death. CAUTION A CAUTION label advises against actions or situations that could damage equipment, produce inaccurate data, or invalidate a procedure. NOTE A NOTE provides useful information regarding a function or a procedure.
Table of Contents Chapter 1 Safety Guidance ........................................................................................................... 1 1.1 Intended Use ........................................................................................................................... 1 1.2 Warnings and Cautions ........................................................................................................... 1 1.2.1 Safety Warnings .........................................................
7.6 Copy Printing........................................................................................................................ 51 7.7 ECG Reports......................................................................................................................... 52 7.7.1 ECG Reports in the Auto Mode ..................................................................................... 52 7.7.2 ECG Reports in the Rhythm Mode ......................................................................
9.6.3 Selecting Antialising (Only for SE-601B/C).................................................................. 89 9.6.4 Setting Key Volume ....................................................................................................... 89 9.6.5 Setting Hint Volume ....................................................................................................... 90 9.6.6 Setting QRS Volume ......................................................................................................
A1.1 Safety Specifications ....................................................................................................... 115 A1.2 Environment Specifications............................................................................................. 116 A1.3 Physical Specifications .................................................................................................... 116 A1.4 Power Supply Specifications .................................................................................
SE-601 Series Electrocardiograph User Manual Safety Guidance Chapter 1 Safety Guidance This chapter provides important safety information related to the use of SE-601 series electrocardiograph (hereinafter called SE-601). 1.1 Intended Use The intended use of SE-601 is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals.
SE-601 Series Electrocardiograph User Manual Safety Guidance WARNING 3. EXPLOSION HAZARD - Do not use the electrocardiograph in the presence of flammable anesthetic mixtures with oxygen or other flammable agents. 4. SHOCK HAZARD - The power receptacle must be a hospital grade grounded outlet. Never try to adapt the three-prong plug to fit a two-slot outlet. 5. Make sure that the power is turned off and the power cord is disconnected from the AC socket before connecting or disconnecting equipment.
SE-601 Series Electrocardiograph User Manual Safety Guidance WARNING 18.If the wireless AP technology is used, the minimum distance between the radiator and the human body should be not less than 20cm. Use the supplied antenna only. There should be no shield in or around the room where the wireless AP is used. 19.Fix attention on the examination to avoid missing important ECG waves. 20.
SE-601 Series Electrocardiograph User Manual Safety Guidance WARNING 28.Parts and accessories used must meet the requirements of the applicable IEC/EN 601 series safety standards, and/or the system configuration must meet the requirements of the IEC/EN 60601-1-1 medical electrical systems standard. 29.Connecting any accessory (such as external printer) or other device (such as the computer) to this electrocardiograph makes a medical system.
SE-601 Series Electrocardiograph User Manual Safety Guidance WARNING 6. When leakage or foul smell is found, stop using the battery immediately. If your skin or cloth comes into contact with the leakage liquid, cleanse it with clean water at once. If the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with clean water first and go to see a doctor immediately. 7. Properly dispose of or recycle the depleted battery according to local regulations . 8.
SE-601 Series Electrocardiograph User Manual Safety Guidance 1.
SE-601 Series Electrocardiograph User Manual Safety Guidance Fn key Power On/Off key MODE key RESET key 1mV/COPY key START/STOP key Tab key Recycle Part Number Serial Number Manufacturer Date of Manufacture The symbol indicates that the device complies with the European Council Directive 93/42/EEC concerning medical devices.
SE-601 Series Electrocardiograph User Manual Safety Guidance Authorized Representative in the European Community With respect to electrical shock, fire and mechanical hazards only in accordance with UL 60601-1and CAN/CSA C22.2 No. 601.1 It indicates that the device should be sent to the special agencies according to local regulations for separate collection after its useful life Rx only (U.S.) Federal (U.S.
SE-601 Series Electrocardiograph User Manual Introduction Chapter 2 Introduction The 6-channel electrocardiograph gathers ECG signals of 12 leads simultaneously. It displays the operation menu, ECG parameters as well as electrocardiograms. The 6-channel ECG waves can be viewed on the LCD screen and printed out by using a high-quality thermal recorder. The manual, auto, rhythm or R-R mode can be chosen freely. SE-601 can be powered by the mains supply or a battery.
SE-601 Series Electrocardiograph User Manual Introduction 2.1 Top Panel Recorder LCD Screen Indicators From left to right on the top panel: A, B, C. Keyboard Figure 2-1 SE-601A LCD Screen Recorder Indicators From left to right on the top panel: A, B, C. Keyboard Figure 2-2 SE-601B / SE-601C Indicator Symbol Name A B C Explanation Mains supply indicator When the device is powered by the mains supply, this indicator is lit.
SE-601 Series Electrocardiograph User Manual Introduction 2.
SE-601 Series Electrocardiograph User Manual Introduction In the MANU mode, pressing the 1mV/Copy key can insert a 1mV calibration mark in the printing course. J START/STOP Start/Stop printing reports K Fn Inputting special characters. Press Fn + a to type è. L Shift Inputting a capital letter. Press Shift + p to type a capital P. M Enter Confirming operation N Delete Deleting characters 2.
SE-601 Series Electrocardiograph User Manual Introduction 2.
SE-601 Series Electrocardiograph User Manual Introduction 3 C4 / V4 8 NC 13 C1 / V1 4 C5 / V5 9 R / RA 14 RF (N) /RL or NC 5 C6 / V6 10 L / LA 15 RF (N) / RL NOTE: The left side of “/” is European standard; and the right side is American standard. 2) Serial Port WARNING The isolated intensity of serial port is 1500V AC and the maximum voltage applied should not exceed +15V DC.
SE-601 Series Electrocardiograph User Manual Introduction 4) USB Socket 1 / USB Socket 2 (Optional) WARNING Only the USB equipment recommended by the manufacturer can be connected to the USB interface. Definitions of corresponding pins: Pin Signal Pin Signal 1 VBUS 3 D+ 2 D- 4 GND WARNING 1. Accessory equipment connected to the analog and digital interfaces must be certified according to the respective IEC/EN standards (e.g.
SE-601 Series Electrocardiograph User Manual Introduction 2.5 Bottom Panel C A E D B C Figure 2-6 SE-601 A / SE-601B / SE-601C Bottom Panel Name Explanation A Speaker Hole Path for sound from speaker (Only for SE-601B/C) B Battery Compartment Compartment for battery C Heat Emission Hole Path for internal heat emission D Label Position for product information label E Fuse The fuse specification is T1AL250VP Ø5×20.
SE-601 Series Electrocardiograph User Manual Introduction WARNING 3. Only qualified service engineers authorized by the manufacturer can open the battery compartment and replace the battery, and batteries of the same model and specification must be used. 4. Only when the device is off can the battery be installed or removed. NOTE: If the battery has not been used for two months or more, you should recharge it before using it again. 2) Fuse There is a fuse installed on the bottom of the main unit.
SE-601 Series Electrocardiograph User Manual About SE-601 Application Chapter 3 About SE-601 Application The following sections provide an overview of the main operations and functions in the SE-601 application. 3.1 Selecting Menu Functions To select Patient, press the function key F1 below Patient. To select , press the function key F5 below - 18 - .
SE-601 Series Electrocardiograph User Manual About SE-601 Application 3.2 Entering Data 1. When the main screen1 is displayed, press the function key F1 below Patient to open the Patient Information window. 2. Press Tab or Shift + Tab to move the cursor to the Name textbox. 3. Press on the keyboard to erase the typed information. 4. Press the letter or number keys on the keyboard to input patient name. Press Fn and a letter key to input the special character in the top right corner of the key.
SE-601 Series Electrocardiograph User Manual About SE-601 Application 3.3 Selecting an Option from a List For SE-601B/C 1. Press Tab or Shift + Tab to move the cursor to the Work Mode item. 2. Press F1 or F2 to select MANU, AUTO, RHYT or R-R. 3. Press Enter to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press Enter to confirm. 4.
SE-601 Series Electrocardiograph User Manual About SE-601 Application D C B C E F A G M L J K I H Figure 3-1 SE-601B / SE-601C Main Screen1 O P N Figure 3-2 SE-601B / SE-601C Main Screen2 E B F D G M L K J I H Figure 3-3 SE-601A Main Screen1 - 21 -
SE-601 Series Electrocardiograph User Manual P About SE-601 Application O N Figure 3-4 SE-601A Main Screen2 Name A B Name ID Explanation Patient Name: within 12 characters (for SE-601B/C) within 10 characters (for SE-601A) When ID Mode is set to Auto or Time, the patient ID is within 10 characters. When ID Mode is set to Manu, the patient ID is within 18 characters.
SE-601 Series Electrocardiograph User Manual About SE-601 Application L Patient Press the function key F1 below Patient to display the Patient Information window M ECG waveform Display ECG waveform Press the function key F5 below screen1 N O P to return to the main File Press the function key F2 below File to open the File Manage screen. For details, please refer to Chapter 8, “Managing Files”. Setup Press the function key F1 below Setup to display the System Setup screen.
SE-601 Series Electrocardiograph User Manual About SE-601 Application On the main screen3, you can print the previous ECG report again according to the set style, paper speed and gain which are shown on the bottom of the screen. Pressing the function key F3 can switch the auto style, pressing the function key F4 can switch the paper speed, and pressing the function key F5 can switch the gain.
SE-601 Series Electrocardiograph User Manual About SE-601 Application Take the Work Mode Setup window for example: On the System Setup screen, press F1, F2, F3, F4, Tab or Shift + Tab to move the cursor. When the cursor is on Work Mode, press Enter to display the Work Mode Setup window. Item Option Figure 3-9 SE-601B / SE-601C Work Mode Setup Window Figure 3-10 SE-601A Work Mode Setup Window For SE-601B/C 1.
SE-601 Series Electrocardiograph User Manual About SE-601 Application 3.6 About the File Manage Screen When the main screen1 is displayed, press the function key F5 below to display the main screen2. Press the function key F2 below File to open the File Manage screen1. 1. Press the function key F5 below the function key F5 below to display the File Manage screen2. Then press to return to the File Manage screen1. 2.
SE-601 Series Electrocardiograph User Manual J K About SE-601 Application L M N Figure 3-13 SE-601B / SE-601C File Manage Screen3 A B C D E Figure 3-14 SE-601A File Manage Screen1 F G H I Figure 3-15 SE-601A File Manage Screen2 J K L M O Figure 3-16 SE-601A File Manage Screen3 Name Explanation A Press the function key F1 below to highlight a file. B Press the function key F2 below to highlight a file.
SE-601 Series Electrocardiograph User Manual About SE-601 Application Press the function key F5 below Manage screen2. E to display the File F Del All Press the function key F1 below Del All to delete all the files. G AllToUSB Press the function key F2 below AllToUSB to copy all the files into the ECGDATA folder of the U disk. H USBToECG Press the function key F3 below USBToECG to copy files in the ECGDATA folder of the U disk to SE-601. I Press the function key F5 below Manage screen1.
SE-601 Series Electrocardiograph User Manual Operation Preparations Chapter 4 Operation Preparations WARNING Before use, the equipment, patient cable and electrodes should be checked. Replace them if there is any evident defectiveness or aging which may impair the safety or the performance, and make sure that the equipment is in proper working condition. 4.1 Power and Earthing WARNING If the integrity of the external protective conductor is in doubt, the equipment should be powered by the battery.
SE-601 Series Electrocardiograph User Manual Operation Preparations Please refer to the maintenance section for how to recharge the battery. When the battery is being recharged, the electrocardiograph can be powered by the mains supply. WARNING Potential equalization bar of the unit should be connected to the potential equalization terminal of the electrical installation when necessary. CAUTION 1.
SE-601 Series Electrocardiograph User Manual Operation Preparations Casing Paper Tray Casing Button Loading/Replacing Process of Folded Paper: 1. Press the casing button downwards with one hand and pull the casing upwards with the other hand to open the recorder. 2. Remove remainder paper from the paper tray if necessary. 3. Take off the wrapper of the new folded paper, and then put it in the paper tray. NOTE: If the paper with black markers is used, make sure that the markers are on the bottom.
SE-601 Series Electrocardiograph User Manual Operation Preparations 4. Pull the paper out with the grid side facing the thermal print head, and shut the recorder casing. 5. Press down the recorder casing firmly. 6. To select Paper Marker. 1) When the 6-channel electrocardiograph is powered by the mains supply or a battery, press on the keyboard to turn on the unit. 2) After the main screen1 is displayed, press the function key F5 below to open the main screen2.
SE-601 Series Electrocardiograph User Manual Operation Preparations 4.3 Preparing the Patient 4.3.1 Instructing the Patient Before attaching the electrodes, greet the patient and explain the procedure. Explaining the procedure decreases the patient’s anxiety. Reassure the patient that the procedure is painless. Privacy is important for relaxation. When possible, prepare the patient in a quiet room or area where others can’t see the patient. Make sure that the patient is comfortable.
SE-601 Series Electrocardiograph User Manual Operation Preparations Main Cable Lead Wires Screw Connecting to Electrocardiograph Connecting to Electrodes 1. Connecting the Patient Cable to the Electrocardiograph Connect the patient cable to the patient cable socket on the right side of the main unit, and then secure them with two screws. 2.
SE-601 Series Electrocardiograph User Manual Operation Preparations Two types of electrode can be used, one is the reusable electrodes, and the other is the disposable electrodes. 4.5.1 Reusable Electrodes Reusable electrodes include limb electrodes and chest electrodes, as the following figures show.
SE-601 Series Electrocardiograph User Manual Operation Preparations Chest Electrode Connection: 1) Ensure that the electrodes are clean; 2) Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to the corresponding electrodes according to the colors and identifiers; 3) Clean the electrode area on the chest surface with 75% alcohol; 4) Daub the round area of 25mm in diameter on each electrode site with gel evenly; 5) Place a small amount of gel on the brim of chest electro
SE-601 Series Electrocardiograph User Manual Operation Preparations 4.5.2 Disposable Electrodes Disposable Electrode Alligator Clip Disposable electrodes must be used together with alligator clips. The electrodes’ positions on the body surface are shown in the following table and figures.
SE-601 Series Electrocardiograph User Manual Operation Preparations V1 V4 V2 V6 V3 V5 Disposable Electrode Connection 1) 2) 3) 4) Align all lead wires of the patient cable to avoid twisting, and connect the alligator clips to the lead wires. Clean the electrode areas on the body surface with 75% alcohol. Attach the disposable electrodes to the electrode positions on the body surface. Clip the disposable electrodes with the alligator clips.
SE-601 Series Electrocardiograph User Manual Operation Preparations 1) Environment: ♦ ♦ Make sure that there is no electromagnetic interference source around the equipment, especially large medical electrical equipment such as electrosurgical equipment, radiological equipment, magnetic resonance imaging equipment etc. Switch off these devices when necessary. Keep the examination room warm to avoid muscle tremor voltages in ECG signals caused by cold.
SE-601 Series Electrocardiograph User Manual Switching On the Electrocardiograph Chapter 5 Switching Electrocardiograph On the ♦ When the mains supply is used, connect the power cord, and the mains supply indicator ( ) is lit. Then press on the keyboard to turn on the unit. The equipment information such as the device name, the version number will be displayed on the LCD screen after self-test. Then the electrocardiograph is ready for use.
SE-601 Series Electrocardiograph User Manual Entering Patient Information Chapter 6 Entering Patient Information After the electrocardiograph is switched on, the main screen1 appears. 6.1 Entering Patient ID 1. By default, the system generates the patient ID automatically. The range of the patient ID is 0 ~ 1999, 999, 999. 2. Or the system can generate the patient ID according to the current time. The procedures are as follows: to display the main screen2.
SE-601 Series Electrocardiograph User Manual Entering Patient Information NOTE: In the AUTO or RHYT mode, when ID Mode is set to Manu and ID Hint is set to On, if you do not input the patient ID, the hint will pop up to remind you to input the patient ID when you press the START/STOP key. 6.2 Entering Other Information When the main screen1 is displayed, press the function key F1 below Patient to open the Patient Information window. 1. Press Tab or Shift + Tab to move the cursor to a textbox. 2.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports Chapter 7 Printing ECG Reports There are four modes to print ECG reports. In the AUTO mode, the lead groups are switched automatically according to the lead sequence during the printing course. After the ECG signals of one lead group are printed within a certain time, the system switches to print ECG signals of another lead group automatically. A 1mV calibration mark will be printed at the beginning of ECG reports.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports For details of the recording sequence, please see Section 9.1.6, “Specifying Recording Sequence”. 5) If you set Auto Style to 6×2 adjust, press Tab or Shift + Tab to move the cursor to the Sample Time textbox, and then enter the sampling time within 10-24s. 6) Press Enter to confirm.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports SE-601B / SE-601C Lead Setup Window SE-601A Lead Setup Window After setup, press Esc or the function key F5 below to exit the System Setup screen. 2. When the main screen1 is displayed, press the MODE key to select the AUTO mode. Press F2 to switch the gain. Press F3 to select a paper speed. Press F4 to set the EMG filter and the Lowpass filter. 3. Press the START/STOP key to print an ECG report.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports SE-601B / SE-601C Work Mode Setup Window SE-601A Work Mode Setup Window 4) On the System Setup screen, press F1, F2, F3, F4, Tab or Shift + Tab to move the cursor. When the cursor is on Lead, press Enter to open the Lead Setup window. 5) For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Lead Sequence item, and then press F1 or F2 to select a sequence.
SE-601 Series Electrocardiograph User Manual 2. 3. 4. 5. Printing ECG Reports After setup, press Esc or the function key F5 below to exit the System Setup screen. When the main screen1 is displayed, press the MODE key to select the MANU mode. Press F2 to switch the gain. Press F3 to select a paper speed. Press F4 to set the EMG filter and the Lowpass filter. Press Tab to select the lead group to be printed. Press the START/STOP key to print an ECG report.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports 5) For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3 item, and then press F1 or F2 to select a lead. For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3 item, and then press F3 or F4 to select a lead. 6) For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Lead Sequence item, and then press F1 or F2 to select a sequence.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports 7.4 R-R Mode Operation Method: to open the 1. When the main screen1 is displayed, press the function key F5 below main screen2. Press the function key F1 below Setup to display the System Setup screen and the cursor is on Work Mode. Press Enter to open the Work Mode Setup window. 1) For SE-601B/C, press F1 or F2 to select R-R. For SE-601A, press F3 or F4 to select R-R. 2) Press Enter to confirm.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports SE-601B / SE-601C Lead Setup Window SE-601A Lead Setup Window After setup, press Esc or the function key F5 below to exit the System Setup screen. 2. When the main screen1 is displayed, press F2 to switch the gain. Press F4 to set the EMG filter and the Lowpass filter. 3. Press the START/STOP key to begin to count the sampling time. When the sampling time reaches 180s, it begins to print an ECG report. 4.
SE-601 Series Electrocardiograph User Manual 3. 4. 5. Printing ECG Reports Select the transmission mode. For details on selecting the transmission mode, please see Section 8.2.1, “Transmitting Files Through the Serial Port”. Log into the Smart ECG Viewer software. In the AUTO or RHYT mode, ECG data will be transmitted through the serial port automatically after an ECG report is printed out. 7.5.2 Transmitting ECG Data through the Net Port 1. 2. 3. 4.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports 7.7 ECG Reports 7.7.1 ECG Reports in the Auto Mode 7.7.1.1 Examples of 6×2+1rhy (a) (b) The above figures (a) and (b) show the ECG reports in the AUTO mode. Template is set to On, and Auto Style is set to 6×2+1rhy.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports 25mm/s (Paper Speed) 10mm/mV (Gain) 60 (Heart Rate) V1.0 (Version number) NOTE: The version number in the ECG reports is just for reference. The real version number can be found on the screen of the electrocardiograph when you turn on the device. Figure (b) shows: Patient Information, Measure Information, Minnesota Code, Diagnosis Information, Average Template, Report Reviewed by, Doctor Name, Current Date and Current Time, 0.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports 7.7.1.2 Example of 3×4+1rhy The above figure shows the ECG report in the AUTO mode. Auto Style is set to 3×4+1rhy. The ECG report includes: 10-10-2008 15: 55: 25 (Current Date and Current Time) 3×4+1rhy ECG wave 0.15~100Hz (0.15Hz DFT Filter, 100Hz Lowpass Filter) 25mm/s (Paper Speed) 60 (Heart Rate) Carefit Hospital (Institution Name) V1.0 (Version number) AC50 (50Hz AC Filter) 10mm/mV (Gain) 7.7.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports Gender: Male, Age: 25 years, Height: 167cm, Weight: 64kg, BP: 112/78mmHg 00:00, 00:10 (Timer) 60 (Heart Rate) (1mV calibration mark) II、V1、V5 (Rhythm Lead name) 20s rhythm waveform of lead II / V1 / V5 0.15~100Hz (0.15Hz DFT Filter, 100Hz Lowpass Filter) AC50 (50Hz AC Filter) 25mm/s (Paper Speed) 10mm/mV (Gain) Carefit Hospital (Institution Name) V1.0 (Version number) 7.7.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports 7.7.4 ECG Reports in the R-R Mode (a) (b) The above figures (a) and (b) show the ECG reports in the R-R mode.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports RR Histogram RR Trend Chart Figure (b) shows: Current Date & Current Time Patient Information (Name, ID, Gender, Age, Height, Weight, BP) Gain, Speed, Filter (1mV calibration mark) II (Lead name) 180s ECG wave of Rhythm Lead 1 - 57 -
SE-601 Series Electrocardiograph User Manual Printing ECG Reports 7.7.
SE-601 Series Electrocardiograph User Manual Printing ECG Reports As the above figure shows, the ECG report printed by the USB printer includes: ID, Paper Speed, Gain, Date and Time; Name, BP, Age, Gender, Weight, Height, Race, Medications, Ward No, Technician; Heart Rate, P duration, PR interval, QRS duration, QT/QTC interval, P/QRS/T axis, RV5/SV1 amplitude, RV5+SV1 amplitude, RV6/SV2 amplitude; Minnesota Code; Diagnosis Information; Report Reviewed by; Doctor Name; DFT Filter, Lowpass Filter, AC Filter
SE-601 Series Electrocardiograph User Manual Managing Files Chapter 8 Managing Files If you want to save the ECG data in the electrocardiograph, you should set the Save Option item to On. The default value is On. Then the ECG data in the AUTO or RHYT mode will be saved on the File Manage screen automatically. To Set Save Option to On 1. When the main screen1 is displayed, press the function key F5 below to open the main screen2. Press the function key F1 below Setup to open the System Setup screen. 2.
SE-601 Series Electrocardiograph User Manual Managing Files Figure 8-1 SE-601B / SE-601C File Manage Screen1 Figure 8-2 SE-601B / SE-601C File Manage Screen2 Figure 8-3 SE-601B / SE-601C File Manage Screen3 Figure 8-4 SE-601A File Manage Screen1 - 61 -
SE-601 Series Electrocardiograph User Manual Managing Files Figure 8-5 SE-601A File Manage Screen2 Figure 8-6 SE-601A File Manage Screen3 On the File Manage screen, files can be printed, transmitted or deleted. If there is no file on the File Manage screen, the following dialog box will pop up when you press function keys. CAUTION 1. When files are being printed, transmitted, deleted or copied, you can not turn off the electrocardiograph. 2.
SE-601 Series Electrocardiograph User Manual Managing Files CAUTION It is forbidden to connect or disconnect a U disk or a USB printer during the transmission course. 8.2.1 Transmitting Files through the Serial Port 1. Connect the RS232 socket of the PC to the RS232 socket of the electrocardiograph with an RS232 cable. 2. Or if the PC has no RS232 socket, connect the USB socket of the PC to the RS232 socket of the electrocardiograph by using the RS232-USB assembly. Figure 8-7 RS232-USB Assembly 3.
SE-601 Series Electrocardiograph User Manual Managing Files 1) When the main screen1 is displayed, press the function key F5 below open the main screen2. to 2) Press the function key F1 below Setup to open the System Setup screen. 3) Press F1, F2, F3, F4, Tab or Shift + Tab to move the cursor. When the cursor is on Transmission, press Enter to open the Transmission Setup window. 4) For SE-601B/C, press F1 or F2 to select UART. For SE-601A, press F3 or F4 to select UART. 5) Press Enter to confirm.
SE-601 Series Electrocardiograph User Manual Managing Files NOTE: If you don’t set the transmission mode before transmission, the following hint will pop up. 1) When the main screen1 is displayed, press the function key F5 below open the main screen2. to 2) Press the function key F1 below Setup to open the System Setup screen. 3) Press F1, F2, F3, F4, Tab or Shift + Tab to move the cursor. When the cursor is on Transmission, press Enter to open the Transmission Setup window.
SE-601 Series Electrocardiograph User Manual Managing Files SE-601B/C Transmission Setup Window SE-601A Transmission Setup Window 8) Press Enter to confirm. 9) Press Esc or the function key F5 below to return to the main screen1. NOTE: 1. For details on configuring the wireless AP, please refer to the user manual delivered with the wireless AP. 2. There should be no shield in or around the room where the wireless AP is used, or else the wireless transmission may fail. 4.
SE-601 Series Electrocardiograph User Manual Managing Files 2) Press the function key F2 below File to open the File Manage screen1. Press the function key F5 below to display the File Manage screen2. 3) Press the function key F2 below AllToUSB to copy files from SE-601 to the ECGDATA folder of the U disk. Or press the function key F3 below USBToECG to copy files from the ECGDATA folder of the U disk to SE-601. 3.
SE-601 Series Electrocardiograph User Manual Managing Files NOTE: 1. If necessary, you can press Esc to cancel the operation during the copying course. 2. Please insert the U disk or the SD card reader that the manufacturer recommends. 3. Only the U disk with the capacity of 1GB or below is supported. Please set the format to FAT when formatting the U disk. 4. In order to protect the U disk, you should insert the U disk after power-on, and pull out the U disk before power-off. 8.
SE-601 Series Electrocardiograph User Manual Managing Files 4. Press Enter to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and then press Enter to confirm. 5. Press Esc to cancel the operation, or press Tab or Shift + Tab to move the cursor to the Cancel button, and then press Enter to cancel the operation. For details on inputting data, please refer to Section 3.2, “Entering Data”. 8.5 Deleting Files 1.
SE-601 Series Electrocardiograph User Manual Managing Files 8.6 Previewing a File (Only for SE-601B/C) When the main screen1 is displayed, press the function key F5 below to open the main screen2. Press the function key F2 below File to open the File Manage screen1. Press F1, F2, Shift + F1 or Shift + F2 to highlight a file, and then press the function key F4 below Select to select the file and open the File Manage screen3. Press the function key F5 below Preview to open the preview screen.
SE-601 Series Electrocardiograph User Manual Managing Files If you want to view the measure information of the selected file, press the function key F3 below Measure to open the measure preview screen, as the following figure shows. Figure 8-10 Measure Preview Screen The measure preview screen shows patient name, patient ID, and measure information including heart rate, P duration, PR interval, QRS duration, QT/QTC interval, P/QRS/T axis, RV5/SV1 amplitude, RV5+SV1 amplitude, and RV6/SV2 amplitude.
SE-601 Series Electrocardiograph User Manual Managing Files 8.7 File Printing When the main screen1 is displayed, press the function key F5 below main screen 2. to open the Press the function key F2 below File to open the File Manage screen 1. Press F1, F2, Shift + F1 or Shift + F2 to highlight a file, and then press the function key F4 below Select to select the file and open the File Manage screen3. Press the START/STOP key to print the selected file.
SE-601 Series Electrocardiograph User Manual System Setup Chapter 9 System Setup After you turn on the device, the main screen1 pops up. Press the function key F5 below to open the main screen2. Then press the function key F1 below Setup to open the System Setup screen. Figure 9-1 SE-601B / SE-601C System Setup Screen Figure 9-2 SE-601A System Setup Screen After setup, press Esc or the function key F5 below to return to the main screen1. 9.
SE-601 Series Electrocardiograph User Manual System Setup Figure 9-3 SE-601B / SE-601C Work Mode Setup Window Figure 9-4 SE-601A Work Mode Setup Window After setup, press Enter to confirm. Then the System Setup screen appears. For SE-601A, if you press Esc to exit the Work Mode Setup window after setup, a hint will pop up to prompt you to save these modifications. 9.1.
SE-601 Series Electrocardiograph User Manual System Setup 9.1.2 Specifying Manual Style For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Manual Style item in the Work Mode Setup window, and then press F1 or F2 to select 3 channel or 6 channel. For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Manual Style item in the Work Mode Setup window, and then press F3 or F4 to select 3 channel or 6 channel.
SE-601 Series Electrocardiograph User Manual System Setup When Auto Style is set to 6×2+1r, ECG waves of 12 leads are printed in 2 groups of 6, and the ECG wave of the rhythm lead selected in the Rhythm Lead1 item is on the bottom of the ECG report. When Auto Style is set to 6×2 adjust, ECG waves of 12 leads are printed in 2 groups of 6. 6×2 adjust means that the time period can be adjusted according to Sample Time, which is different from other styles.
SE-601 Series Electrocardiograph User Manual System Setup 9.1.6 Specifying Recording Sequence For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Record Sequence item in the Work Mode Setup window, and then press F1 or F2 to select Sequential or Simultaneous. For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Record Sequence item in the Work Mode Setup window, and then press F3 or F4 to select Sequential or Simultaneous.
SE-601 Series Electrocardiograph User Manual System Setup 9.2 Filter Setup Press F1, F2, F3, F4, Tab or Shift + Tab to move the cursor to Filter on the System Setup screen, and then press Enter to open the Filter Setup window. In the Filter Setup window, you can set AC Filter, EMG Filter, DFT Filter, and Lowpass Filter. Figure 9-5 SE-601B / SE-601C Filter Setup Window Figure 9-6 SE-601A Filter Setup Window After setup, press Enter to confirm. Then the System Setup screen appears.
SE-601 Series Electrocardiograph User Manual System Setup EMG Filter suppresses disturbances caused by strong muscle tremor. The cutoff frequency can be set to 25Hz, 35Hz or 45Hz. Select Off to turn off the function. 9.2.3 Setting DFT Filter For SE-601B/C, press Tab or Shift + Tab to move the cursor to the DFT Filter item in the Filter Setup window. Press F1 or F2 to select an option. For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the DFT Filter item in the Filter Setup window.
SE-601 Series Electrocardiograph User Manual System Setup Figure 9-7 SE-601B / SE-601C Record Setup Window Figure 9-8 SE-601A Record Setup Window After setup, press Enter to confirm. Then the System Setup screen appears. For SE-601A, if you press Esc to exit the Record Setup window after setup, a hint will pop up to prompt you to save these modifications. NOTE: The Patient Info, Template, Measure, Minnesota Code, Analysis, and Position Marker items are valid only in the AUTO mode. 9.3.
SE-601 Series Electrocardiograph User Manual System Setup NOTE: 1. During the USB printing course,pressing the START/STOP key again can not stop printing ECG reports. 2. USB printing only works in the auto real-time sampling mode, auto pre-sampling mode, and auto trigger sampling mode. It also supports copy printing, and file printing in the AUTO mode on the File Manage screen. 3. Make sure that paper is installed in the USB printer before printing.
SE-601 Series Electrocardiograph User Manual System Setup NOTE: The side with black markers should be on the bottom. When Paper Marker is set to Style2, only the paper with black markers in the bottom right corner of each page can be used, which is shown in the following figure. NOTE: The style of the loaded thermal paper should consist with the set paper style; otherwise the advance of the paper may be blocked.
SE-601 Series Electrocardiograph User Manual System Setup When Paper Marker is set to No, 1) The device does not roll back the recorder paper to the start point of each page at the beginning of the printing course, and it does not advance the paper to the next black marker at the end of the printing course. 2) Before printing, you can press the Space key to advance the paper for about 2cm. 9.3.
SE-601 Series Electrocardiograph User Manual System Setup 9.3.6 Selecting Template For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Template item in the Record Setup window, and then press F1 or F2 to select Off or On. For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Template item in the Record Setup window, and then press F3 or F4 to select Off or On. In the AUTO mode, when Template is set to On, average template will be printed in the ECG reports.
SE-601 Series Electrocardiograph User Manual System Setup In the AUTO mode, when Analysis is set to On, the auto diagnosis information will be printed in the ECG reports. In the AUTO mode, when Analysis is set to Off, the auto diagnosis information will not be printed in the ECG reports. 9.3.10 Selecting Position Marker For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Position Marker item in the Record Setup window, and then press F1 or F2 to select Off or On.
SE-601 Series Electrocardiograph User Manual System Setup Figure 9-10 SE-601A Lead Setup Window After setup, press Enter to confirm. Then the System Setup screen appears. For SE-601A, if you press Esc to exit the Lead Setup window after setup, a hint will pop up to prompt you to save these modifications. 9.4.1 Setting Rhythm Lead1/2/3 For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3 item in the Lead Setup window, and then press F1 or F2 to select an option.
SE-601 Series Electrocardiograph User Manual System Setup 9.5 Transmission Setup NOTE: To transmit ECG data to the PC, the Smart ECG Viewer software of the manufacturer must be installed in the PC. Press F1, F2, F3, F4, Tab or Shift + Tab to move the cursor to Transmission on the System Setup screen, and then press Enter to open the Transmission Setup window. In the Transmission Setup window, you can set Transmission Mode, Remote IP, Local IP, Gateway and Subnet Mask.
SE-601 Series Electrocardiograph User Manual System Setup To Set Remote IP Set the REMOTE IP item to the local IP of Smart ECG Viewer. To Set Local IP Set the first three sections of the Local IP item to the first three sections of the local IP of Smart ECG Viewer. The last section of the LOCAL IP item can be set at random, but it can’t be the same as the last section of the local IP of Smart ECG Viewer.
SE-601 Series Electrocardiograph User Manual System Setup Figure 9-14 SE-601C Display & Sound Setup Window Figure 9-15 SE-601A Sound Setup Window After setup, press Enter to confirm. Then the System Setup screen appears. For SE-601A, if you press Esc to exit the Display&Sound Setup window after setup, a hint will pop up to prompt you to save these modifications. 9.6.
SE-601 Series Electrocardiograph User Manual System Setup For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Key Volume item in the Sound Setup Window, and then press F3 or F4 to select On or Off. 9.6.5 Setting Hint Volume When the electrocardiograph displays a hint such as Lead Off, Overload, Battery Weak etc., there is a sound. When Hint Volume is set to Off, there is no hint sound.
SE-601 Series Electrocardiograph User Manual System Setup Figure 9-16 SE-601B / SE-601C Patient Question Setup Window Figure 9-17 SE-601A Patient Question Setup Window After setup, press Enter to confirm. Then the System Setup screen appears. For SE-601A, if you press Esc to exit the Patient Question window after setup, a hint will pop up to prompt you to save these modifications. 9.7.
SE-601 Series Electrocardiograph User Manual System Setup For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the ID Hint item in the Patient Question window, and then press F3 or F4 to select On or Off. In the AUTO or RHYT mode, when ID Mode is set to Manu and ID Hint is set to On, if you do not input the patient ID, the hint will pop up to remind you to input the patient ID when you press the START/STOP key. 9.7.
SE-601 Series Electrocardiograph User Manual System Setup 9.7.6 Selecting Next Patient For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Next Patient item in the Patient Question window, and then press F1 or F2 to select Off or On. For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Next Patient item in the Patient Question window, and then press F3 or F4 to select Off or On.
SE-601 Series Electrocardiograph User Manual System Setup When Prompt is set to Reviewed by, if Doctor in the Patient Question window is set to On, and the doctor name is input in the Patient Information window, Report Reviewed by and the doctor’s name is printed on the bottom of ECG reports. When Prompt is set to Unconfirmed Report, Unconfirmed Report is printed on the bottom of ECG reports, and doctor name will not be printed on the bottom of ECG reports. 9.7.
SE-601 Series Electrocardiograph User Manual System Setup For SE-601A, if you press Esc to exit the Date&Time Setup window after setup, a hint will pop up to prompt you to save these modifications. 9.8.1 Setting Current Date/Current Time Press Tab or Shift + Tab to move the cursor to the Current Date or Current Time textbox in the Date&Time Setup window, and then input the date or time manually.
SE-601 Series Electrocardiograph User Manual System Setup 9.8.5 Setting Power-Off Time Press Tab or Shift + Tab to move the cursor to the Power Off textbox in the Date&Time Setup window, and then input the power-off time manually. When Power Off is set to 000 Minutes, this function will not be effective. NOTE: 1. Power-off time is counted from the time when you last press the keys on the keyboard. 2. Only when the device is powered by the battery, can the set automatic power-off time be effective. 9.8.
SE-601 Series Electrocardiograph User Manual System Setup Figure 9-23 SE-601A More Setup Window After setup, press Enter to confirm. Then the System Setup screen appears. For SE-601A, if you press Esc to exit the More Setup window after setup, a hint will pop up to prompt you to save these modifications. 9.9.1 Choosing a Language For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Language item in the More Setup window, and then press F1 or F2 to select a language.
SE-601 Series Electrocardiograph User Manual System Setup 9.9.4 Entering Institution Press Tab or Shift + Tab to move the cursor to the Institution textbox in the More Setup window, and then manually input the institution name. For SE-601B/C, the range is within 20 characters. For SE-601A, the range is within 10 characters. 9.9.5 Restoring Default Settings For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Default item in the More Setup window, and then press F1 or F2 to select No or Yes.
SE-601 Series Electrocardiograph User Manual 23 24 25 26 27 28 29 30 31 System Setup ID Mode ID hint/Gender/Height/Weight/BP Race/Medication/Ward NO/Doctor/Technician/Next Patient Prompt Period Interval Period Duration Power Off LCD Off Save Option Auto On Off Reviewed By 001 060 010 010 On 9.9.6 Setting Extern Input/Extern Output For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Extern Input or Extern Output item in the More Setup window, and then press F1 or F2 to select Off or On.
SE-601 Series Electrocardiograph User Manual Switching Off the Electrocardiograph Chapter 10 Switching Electrocardiograph Off the When the battery is used, press the key for 1 second or more to display the hint System is shutting down… on the screen. Then the device will be off a few seconds later. When the mains supply is used, press the key for 1second or more to display the hint System is shutting down… on the screen. Then the device will be off a few seconds later.
SE-601 Series Electrocardiograph User Manual Hint Information Chapter 11 Hint Information Hint information and the corresponding causes provided by the electrocardiograph are listed in Table 11-1. Table 11-1 Hint Information and Causes Hint Information Causes Lead off Electrodes fall off the patient or the patient cable falls off the unit, or a high polarization voltage occurs. Battery Weak The battery is weak. No Paper Recorder paper runs out or is not loaded or the casing is not firmly closed.
SE-601 Series Electrocardiograph User Manual Troubleshooting Chapter 12 Troubleshooting 1. Operating Problems Q1: I was trying to select a file from the file list on the File Manage screen, but the file was in the middle of the long list. Is there any way to make the selection faster? A1: Actually, the system provides a method for fast moving: pressing Shift + F1 or Shift + F2 can move the cursor up or down in the file list very fast.
SE-601 Series Electrocardiograph User Manual Troubleshooting 2. Printing Problems Q1: I was encountered with paper-jam, what was I supposed to do? A1: If it happened for the first time, it might be the result of an inappropriate placement of the paper. In this case, please open the recorder casing, pull the paper out of the paper tray, tear the pages with rumples, and then put the paper in the paper tray again, adjust the position of the paper carefully and close the casing.
SE-601 Series Electrocardiograph User Manual Troubleshooting Q4: I want to print the hospital name in the report, but I can’t find the place to enter it, where is it? A4: Please open the More Setup window, and move the cursor to the Institution textbox, and then input the hospital name. The content you input in this textbox will be printed in the report. For details, please refer to Section 9.9.4, “Entering Institution”.
SE-601 Series Electrocardiograph User Manual Troubleshooting 4. Main Unit Problems Q1: After power-on, the ECG stays on the logo screen and doesn’t open the main screen. I have restarted the machine several times, but there is no better change. A1: The reason for this problem might be: there is a key pressed down, without springing up. Find that key, and make it spring up, the problem should be solved.
SE-601 Series Electrocardiograph User Manual Cleaning, Care and Maintenance Chapter 13 Cleaning, Care and Maintenance 13.1 Cleaning CAUTION Turn off the power before cleaning and disinfection. The mains supply must be switched off if it is in use. 13.1.1 Cleaning the Main Unit and the Patient Cable To clean the electrocardiograph: 1. Unplug the AC power cord. 2. Wipe the external surfaces of the electrocardiograph with a soft cloth dampened in any of the approved cleaning solutions listed below.
SE-601 Series Electrocardiograph User Manual Cleaning, Care and Maintenance Dirty and soiled thermal print head will deteriorate the printing definition. So it should be cleaned at least once a month regularly. CAUTION Prevent the detergent from seeping into the main unit while cleaning. Do not immerse the unit or the patient cable into liquid under any circumstances. 13.2 Disinfection Disinfection of the main unit is not necessary need in daily maintenance, it is only necessary in operating room.
SE-601 Series Electrocardiograph User Manual Cleaning, Care and Maintenance 2) Recharge The 6-channel electrocardiograph is equipped with the recharge control circuit together with the battery. When the unit is connected to the mains supply, the battery will be recharged automatically. Then the battery recharging indicator ( ) and the mains supply indicator ( ) flashes in the top will be lit at the same time. During the recharging course, the symbol right corner of the LCD screen.
SE-601 Series Electrocardiograph User Manual Cleaning, Care and Maintenance 13.3.2 Recorder Paper NOTE: Recorder paper provided by the manufacturer should be used. Other paper may shorten the life of the thermal print head. The deteriorated print head may lead to illegible ECG reports and block the advance of the paper. Storage Requirements: ♦ Recorder paper should be stored in a dry, dark and cool area, avoiding excessive temperature, humidity and sunshine.
SE-601 Series Electrocardiograph User Manual Cleaning, Care and Maintenance g) Test the enclosure leakage current according to IEC/EN/UL 60601-1: Limit: NC 100μA, SFC 300μA. h) Test the patient leakage current according to IEC/EN 60601-1: Limit: NC a.c. 10μA, d.c. 10μA; SFC a.c. 50μA, d.c. 50μA. i) Test the patient auxiliary current according to IEC/EN 60601-1: Limit: NC a.c. 10μA, d.c. 10μA; SFC a.c. 50μA, d.c. 50μA.
SE-601 Series Electrocardiograph User Manual ♦ Cleaning, Care and Maintenance After long-term use, the surfaces of electrodes will be oxidized because of erosion and other causes. By this time, electrodes should be replaced to achieve high-quality ECG records. CAUTION The device and accessories are to be disposed of according to local regulations after their useful lives. Alternatively, they can be returned to the dealer or the manufacturer for recycling or proper disposal.
SE-601 Series Electrocardiograph User Manual Accessories Chapter 14 Accessories WARNING Only the patient cable and other accessories supplied by the manufacturer can be used. Or else, the performance and electric shock protection can not be guaranteed. Accessory ECG Cable (European) ECG Cable (American) Part Number 01.57.107581 (Snap Style) 01.57.107583 (Grabber Style) 01.57.107582 (Snap Style) 01.57.107584 (Grabber Style) ECG Cable (European) 01.57.471016 ECG Cable (American) 01.57.
SE-601 Series Electrocardiograph User Manual Smart ECG Viewer Smart ECG Viewer Software Key Accessories 03.24.38952 03.24.38953 11.18.47116 (TINY-SPRO) 12.01.47194 (USB) RS232 download Cable 11.13.20117 Ethernet download cable 11.13.20096 Wireless AP 11.17.047338 External Ink-jet Printer (HP Deskjet D2668) 11.18.052215 U disk 11.18.078204 SD Card (1G) 11.18.052199 SD Card Reader 11.17.047324 ECG bag 01.56.110351 MT-202 Trolley 03.28.111848 12V Vehicle-carried Inverter 11.21.
SE-601 Series Electrocardiograph User Manual Warranty & Service Chapter 15 Warranty & Service 15.1 Warranty EDAN warrants that EDAN’s products meet the labeled specifications of the products and will be free from defects in materials and workmanship that occur within warranty period. The warranty is void in cases of: a) Damage caused by mishandling during shipping. b) Subsequent damage caused by improper use or maintenance. c) Damage caused by alteration or repair by anyone not authorized by EDAN.
SE-601 Series Electrocardiograph User Manual Technical Specifications Appendix 1 Technical Specifications A1.1 Safety Specifications Anti-electric-shock type: IEC/EN 60601-1+A1+A2, IEC/EN 60601-1-2+A1, IEC/EN60601-2-25, ANSI/AAMI EC11, IEC/EN 60601-2-51, UL 60601-1, CAN/CSA C22.2 No.601.
SE-601 Series Electrocardiograph User Manual Technical Specifications A1.2 Environment Specifications Temperature: Transport & Storage Working -20℃ (-4ºF) ~ +55℃ (+131ºF) +5ºC (+41ºF) ~ +40ºC (+104ºF) 25%~93% 25%~80% Non-Condensing Non-Condensing 700hPa ~1060hPa 860hPa ~1060hPa Relative Humidity: Atmospheric Pressure: A1.3 Physical Specifications Dimensions 310mm×322mm×101mm (12.2in×12.7in×3.98in) Weight Approx. 2.6kg (5.7lbs) (excluding recorder paper and battery) Display SE-601A: 3.
SE-601 Series Electrocardiograph User Manual Technical Specifications A1.5 Performance Specifications Recording Recorder: Thermal dot-matrix recorder Printing Density 8 dots per mm / 200 dots per inch (amplitude axes) 40 dots per mm / 1000 dots per inch (time axes, @ 25 mm/s) Recorder Paper: Folded thermal paper: 110mm×140mm×144pages Effective Width: 104mm Paper Speed: 5mm/s, 6.25mm/s, 10mm/s, 12.
SE-601 Series Electrocardiograph User Manual Multichannel crosstalk Technical Specifications ≤0.5mm AC Filter: On/Off Filter DFT Filter: 0.05Hz / 0.15Hz / 0.25Hz / 0.32Hz / 0.5Hz / 0.67Hz EMG Filter: 25Hz/35Hz/45Hz/OFF LOWPASS Filter:150Hz/100Hz/75Hz CMRR ≥115dB Sampling Frequency 1000 Hz Pacemaker Detection Amplitude ±2 to ±700 mV Width 0.1 to 2.
SE-601 Series Electrocardiograph User Manual EMC Information Appendix 2 EMC Information Guidance and manufacture’s declaration - electromagnetic emissionsfor all EQUIPMENT and SYSTEMS Guidance and manufacture’s declaration - electromagnetic emission The SE-601 Electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or the user of the SE-601 Electrocardiograph should assure that it is used in such an environment.
SE-601 Series Electrocardiograph User Manual EMC Information Guidance and manufacture’s declaration - electromagnetic immunity for all EQUIPMENT and SYSTEMS Guidance and manufacture’s declaration - electromagnetic immunity The SE-601 Electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or the user of SE-601 Electrocardiograph should assure that it is used in such an environment.
SE-601 Series Electrocardiograph User Manual EMC Information Guidance and manufacture’s declaration - electromagnetic immunity for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING Guidance and manufacture’s declaration - electromagnetic immunity The SE-601 Electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or the user of SE-601 Electrocardiograph should assure that it is used in such an environment.
SE-601 Series Electrocardiograph User Manual EMC Information NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
SE-601 Series Electrocardiograph User Manual EMC Information Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM - for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and the SE-601 Electrocardiograph The SE-601 Electrocardiograph is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
SE-601 Series Electrocardiograph User Manual Abbreviation Appendix 3 Abbreviation Abbr English LCD Liquid Crystal Display BP Blood Pressure ECG Electrocardiogram/Electrocardiograph HR Heart Rate aVF Left Foot Augmented Lead aVL Left Arm Augmented Lead aVR Right Arm Augmented Lead LA Left Arm LL Left Leg RA Right Arm RL Right Leg ID Identification AC Alternating Current USB Universal Serial Bus AGC Auto Gain Control SFC Single Fault Condition NC Normal Condition - 124 -