SE-601 Series Electrocardiograph User Manual System Setup
- 97 -
Figure 9-23 SE-601A More Setup Window
After setup, press Enter to confirm. Then the System Setup screen appears.
For SE-601A, if you press Esc to exit the More Setup window after setup, a hint will pop up to
prompt you to save these modifications.
9.9.1 Choosing a Language
For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Language item in the More
Setup window, and then press F1 or F2 to select a language.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Language item in the
More Setup window, and then press F3 or F4 to select a language.
9.9.2 Setting Pacemaker Detection Sensitivity
For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Pacemaker Detection
Sensitivity item in the More Setup window, and then press F1 or F2 to select Low or High.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Pacemaker Detection
Sensitivity item in the More Setup window, and then press F3 or F4 to select Low or High.
When Pacemaker Detection Sensitivity is set to High, the pacemaker signals are easy to be
detected.
When Pacemaker Detection Sensitivity is set to low, the pacemaker signals are not easy to be
detected.
For more information about the detectable parameter ranges, please refer to Appendix 1,
“Technical Specifications”.
9.9.3 Setting Save Option
For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Save Option item in the
More Setup window, and then press F1 or F2 to select Off or On.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Save Option item in
the More Setup window, and then press F3 or F4 to select Off or On.
When Save Option is set to On, ECG data in the AUTO or RHYT mode will be saved into the
flash memory of the device automatically after an ECG report is printed out.
When Save Option is set to Off, ECG data will not be saved into the flash memory of the device.
NOTE: In the period sampling mode and in the auto style of 6×2 adjust, the ECG data
will not be saved.