Product Manual
Patient Monitor User Manual Monitoring CO
2
- 118 -
2. Connect the sampling cannula or the sampling line to the water trap.
3. Set Work Mode to Measure.
4. For intubated patients, an airway adapter is required. For non-intubated patients, place the
nasal cannula or the sampling mask onto the patient.
Place the Nasal Cannula
CAUTION
1 The water trap collects water drops condensed in the sampling line and therefore
prevents them from entering the module. If the water trap is nearly filled, you should
replace it to avoid blocking the airway.
2 Based on a sample gas temperature of 37 C, a room temperature of 23 C and
sample relative humidity of 100%, the water trap will be filled after approximately 90
hours with the flowrate of 100 ml/min, approximately 130 hours with the flowrate of
70 ml/min, and approximately 180 hours with the flowrate of 50 ml/min. In clinical
practice, the water trap can be used for a longer time before it is filled. It is
recommended to replace the water trap once every month.
3 When replacing the water trap or suspecting the measurement value, please check if
the O-rings of the water trap holder are normal and well installed. If the O-rings get
damaged or loose, contact EDAN’s service staff.
4 To prevent the module from abnormal work, please ensure the water trap detection
button is not mistakenly touched.
5 Please replace and discard the water trap when blocking. Don’t reuse it, otherwise
the reading is not accurate and even the device may be damaged.