Product Manual
Patient Monitor User Manual Monitoring CO
2
- 120 -
WARNING
Safety and effectiveness of the respiration measurement method in the detection of apnea,
particularly the apnea of prematurity and apnea of infancy, has not been established.
16.6 Setting CO
2
Waveform
Open the menu CO
2
Waveform Setup by clicking on the CO
2
waveform area:
Choose Mode and set it to Curve or Filled from the pop-up list;
Choose Sweep and select a suitable setting from the pop-up list. The bigger the value is, the
wider the waveform will be.
16.7 Intubation Mode
Intubation mode is suitable for CO
2
monitoring. During general anesthesia, the monitor can be set
to intubation mode to eliminate unnecessary alarms. In intubation mode, CO
2
-related
physiological alarm including CO
2
APNEA will be turned off.
To enter intubation mode, follow these steps:
1. Click Intubation Mode in CO
2
Setup;
2. Select Duration in Intubation Mode, there are two selections: 3 min and 5 min. The default
setting is 3 min.
3. Click Start, the monitor will start the intubation mode. During the intubation mode, the
monitor will display the intubation mode and remaining time in the form of text.
When countdown finishes or clicking End in Intubation Mode menu, the monitor will exit the
intubation mode; After exiting intubation mode, the monitor will respond the physiological alarm
related to CO
2
.
NOTE:
The intubation mode is not available when CO
2
module is in Standby mode.