Product Manual

Patient Monitor User Manual Recording
- 143 -
procedure below:
Recording Type
Description/ Procedure
Continual real-time recording
Select at least one Rec waveform in Recorder Setup (A
maximum of three waveforms can be selected), select
Continual in R-T Rec Time. Press shortcut key on
the screen to start the recording. Press the shortcut again to
stop recording.
8-second real-time
recording/20-second real-time
recording
Select at least one Rec waveform in Recorder Setup (A
maximum of three waveforms can be selected), select 8 s or
20 s in R-T Rec Time, set Record Interval as needed,
press shortcut key on the screen to start the
recording. Press the shortcut again to stop recording or
when R-T Rec time ends, the monitor stops recording
automatically. The runtime for each wave is 8 seconds or 20
seconds. The record Interval can be set as: Off, 10 min, 20
min, 30 min, 40 min, 50 min, 1 h, 2 h, 3 h, 4 h. The
default recording time is 8s.
Trend graph recording
Select Menu > Review > Trend Graph, click Record to
start recording.
Trend table recording
Select Menu > Review > Trend Table, click Record to
start recording.
NIBP review recording
Select Menu > Review > NIBP Review, click Record to
start recording.
Arrhythmia review recording
Select Menu > Review > ARR Review, select one
arrhythmia alarm and click Record to start recording.
Alarm review recording
Select Menu > Review > Alarm Review, select one alarm
and click Record to start recording.
Drug calculation titration
recording
Select Menu > Common Function > Calculation > Drug
Dose > Titration, click Record to start recording.
Hemodynamic Calculation
result recording
Select Menu > Common Function > Calculation >
Hemodynamics, click Record to start recording.
Oxygenation Calculation result
recording
Select Menu > Common Function > Calculation >
Oxygenation, click Record to start recording.
Ventilation Calculation result
recording
Select Menu > Common Function Calculation >
Ventilation, click Record to start recording.