Patient Monitor User Manual Recording
- 144 -
Recording Type
Description/ Procedure
Renal Function Calculation
result recording
Select Menu > Common Function > Calculation > Renal
Function, click Record to start recording.
12-lead analysis recording
Select ECG Setup > 12-L Review, click Record to start
recording.
C.O. measurement recording
Select C.O. Option > C.O. Measure, click Record to start
recording.
Frozen waveform recording
In the Freeze window, click Record to start recording.
ST View recording
In the ST View window, click Record to start recording.
QT View recording
In the QT View window, click Record to start recording.
To manually stop recording, click Record again in the related windows.
The recorder will stop recording in the following situations:
The recording task is finished.
No paper in the recorder.
Malfunction stops the recorder from running properly.
NOTE:
1 You can also use the shortcut key on the screen to manually start or stop
recording.
2 For waveforms with sweep of 6.25 mm/s, after entering freeze status, the recording
speed will automatically be adjusted to 12.5 mm/s. Users can modify the recording
speed according to their actual needs. The options are 12.5 mm/s, 25 mm/s and 50
mm/s.
22.3 Recorder Operations and Status Messages
22.3.1 Record Paper Requirement
Only standard thermosensitive record paper can be used: otherwise the recorder may not function,
the recording quality may be poor, and the thermosensitive printhead may be damaged.
22.3.2 Proper Operation
When the recorder is working, the record paper goes out steadily. Do not pull the paper
outward with force: otherwise the recorder may be damaged.
Do not operate the recorder without record paper.