Product Manual
Patient Monitor User Manual Recording
- 145 -
22.3.3 Paper Out
When the Recorder Out OF Paper alarm is displayed, the recorder cannot start. Please insert
record paper properly.
22.3.4 Replacing Paper
1. Pull outwards the upper arc part of the recorder casing to release the casing, shown in the
following figure.
2. Insert a new roll of paper into the paper cassette, printing side facing upwards.
3. Ensure proper position and tidy margin.
4. Pull about 2 cm of the paper out, and close the recorder casing.
NOTE:
Be careful when inserting papers. Avoid damaging the thermo-sensitive print head.
Unless when inserting papers or shooting troubles, do not leave the recorder catch open.
22.3.5 Removing Paper Jam
When the recorder functions or sounds improperly, you should open the recorder casing to check
for a paper jam. Remove the paper jam in the following way:
■ Cut the record paper from the feeding edge.
■ Open the recorder casing.
■ Re-insert the paper.
NOTE:
1 If the monitor is not configured with the recorder function, it will indicate Recorder
Setup Needed after is pressed.
2 Do not touch the thermo-sensitive print head when performing continuous recording.