Patient Monitor User Manual Care and Cleaning
- 156 -
26.2.1 Cleaning the Monitor
WARNING
Before cleaning the monitor, make sure that the monitor is switched off and disconnected
from the power line.
To surface-clean the monitor, follow these steps:
1. Switch off the monitor and disconnect it from the power line.
2. Wipe the entire exterior surface, including the screen, of the equipment using a soft cloth
dampened with the cleaning solution thoroughly until no visible contaminants remain.
3. After cleaning, wipe off the cleaning solution with a fresh cloth or towel dampened with tap
water until no visible cleaning agent remains.
4. Dry the monitor in a ventilated and cool place.
26.2.2 Cleaning the Reusable Accessories
26.2.2.1 Cleaning the ECG Cable Assembly
1. Wipe the cable assembly with a soft cloth dampened with the cleaning solution until no
visible contaminants remain.
2. After cleaning, wipe off the cleaning solution with a fresh cloth or towel dampened with tap
water until no visible cleaning agent remains.
3. Wipe off residual moisture with a dry cloth.
4. Leave the cable assembly to air dry.
26.2.2.2 Cleaning the Blood Pressure Cuff
Cleaning the Cuff:
1. Take out the air bladder before cleaning.
2. Hand wash the cuff with the cleaning solution; clean the air bladder with a soft cloth
dampened with the cleaning solution until no visible contaminants remain.
3. Rinse the cuff and after cleaning, wipe off the cleaning solution with a fresh cloth or towel
dampened with tap water until no visible cleaning agent remains.
4. Wipe off residual moisture with a dry cloth.
5. Air dry the cuff thoroughly after cleaning.
Replacing the Air Bladder:
After cleaning, replace the air bladder into the cuff following the steps below:
1. Roll the bladder lengthwise and insert it into the cuff from the large opening at one end of
the cuff.