Product Manual

Patient Monitor User Manual Product Specification
- 178 -
Multichannel Crosstalk
5% of the input signal
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.5.
Frequency and Impulse
Response
Frequency response:
Input a 5 Hz, 1 mV sine wave signal, and the output signal
amplitude remains within the range of 71 % to 110 % at 0.67 Hz
and 40 Hz.
Input a 1 Hz, 1.5 mV 200 ms triangular wave input signal, and
the output shall be within 11.25 mm~15 mm.
Impulse response:
Displacement value: 0.1 mV
Slope: 0.3 mV/s following the end of the pulse.
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.8.
Sampling Frequency
1000 Hz
Sampling Channel Switch
Time
< 80 μS
A/D Precision
24 Bits (Minimum resolution: 0.077uV/LSB)
ESU Protection
Cut mode: 300 W
Coagulation mode: 100 W
Restore time: ≤ 10 s
Electrosurgical
Interference Suppression
Test according to ANSI/AAMI EC13:2002, Sect. 5.2.9.14.
Complied with ANSI/AAMI EC13:2002, Sect. 4.2.9.14.
Minimum Input Slew Rate
(Lead II)
> 2.5 V/s
Baseline Reset Time
< 3 s
Pace Pulse
Pulse Indicator
Pulse is marked if the requirements of IEC 60601-2-27: 2011,
Sect. 201.12.1.101.12 are met:
Amplitude: ±2 mV to ±700 mV
Width: 0.1 ms to2.0 ms
Ascending time: 10 μs to 100 μs
Pulse Rejection
Pulse is rejected if the requirements of IEC 60601-2-27: 2011,
Sect. 201.12.1.101.13 are met:
Amplitude: ±2 mV to ±700 mV
Width: 0.1 ms to 2.0 ms
Ascending time: 10 μs to 100 μs
Pace Pulse Detecting Lead: one among I, II, III, AVR, AVL, AVF, V1, V2, V3,V4, V5, V6