Product Manual

Patient Monitor User Manual Product Specification
- 179 -
Heart Rate
HR Calculation
Range
ADU: 15 bpm to 300 bpm
PED/NEO: 15 bpm to 350 bpm
Accuracy
±1% or 1 bpm, whichever is greater
Resolution
1 bpm
Sensitivity
300 μVPP
QRS Detection Range
The detection range has exceeded the requirement described in
the standard:
Width: 70 ms~120 ms for adult, 40 ms~120 ms for
Pediatric/neonate.
Amplitude: 0.5 mv~5 mv
In adult mode, these two signals are not responded:
1. when QRS amplitude of 0.15 mV or less is applied;
2. when QRS duration of 10 ms and QRS amplitude of 1 mV or
less is applied.
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.15.
PVC
Range
ADU: (0 to 300) PVCs/ min
PED/NEO: (0 to 350) PVCs/ min
Resolution
1 PVCs/min
Pauses/min
Range
ADU/PED/NEO: (0 to 30) pauses/min
Resolution
1 pause/min
ST value
Range
-2.0 mV to +2.0 mV
Accuracy
-0.8 mV to +0.8 mV: ±0.02 mV or 10%, whichever is greater.
Beyond this range: not specified.
Resolution
0.01 mV
QT measurement
Range
200 ms ~ 800 ms
Resolution
4 ms
Accuracy
± 30 ms