Product Manual
Patient Monitor User Manual Product Specification
- 181 -
Maximum Start-up Alarm Time for Tachycardia
Ventricular Tachycardia
1 mV 206 bpm
Gain 0.5: 10 s
Gain 1.0: 10 s
Gain 2.0: 10 s
Ventricular Tachycardia
2 mV 195 bpm
Gain 0.5: 10 s
Gain 1.0: 10 s
Gain 2.0: 10 s
Response Time of Heart
Rate Meter to Change in
HR
HR range: 80 bpm to 120 bpm
Range : Within 11 s
HR range: 80 bpm to 40 bpm
Range : Within 11 s
☆Tall T-wave Rejection
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.17
minimum recommended 1.2 mV T-Wave amplitude
Accuracy of Heart Rate
Meter and Response to
Irregular Rhythm
Complied with IEC 60601-2-27: 2011, Sect. 201.7.9.2.9.101 b)
4), the HR value after 20 seconds of stabilization is displayed as
follows:
Ventricular bigeminy: 80 bpm±1 bpm
Slow alternating ventricular bigeminy: 60 bpm±1 bpm
Rapid alternating ventricular bigeminy: 120 bpm±1 bpm
Bidirectional systoles: 91 bpm±1 bpm
Time to Alarm for Heart
Rate alarm conditions
Asystole alarm: ≤ 10 s
HR low alarm: ≤ 10 s
HR high alarm: ≤ 10 s
Arrhythmia analyses
Asystole
V-Fib/V-Tach
Couplet
Vent Rhythm
PVC Bigeminy
PVC Trigeminy
Tachy
R on T
PVC
Irr Rhythm
Brady
Missed Beat
Pacer not Pacing
Vent Brady
Pacer not Capture
VEB
Run PVCs
Acc. Vent Rhythm
IPVC
Non-Sustain VT
Multiform PVCs
Pauses/min High
Pause
Afib
PAC Bigeminy
PVCs High
Low Voltage(Limb)
ExtremeBrady
PAC Trigeminy
Wide QRS Tachy
Sustain VT
ExtremeTachy
V-Tach
12-Lead ECG
Synchronization Analysis
Average parameters of heart beat
Heart rate (bpm)
Time limit of P wave (ms)