Product Manual
Patient Monitor User Manual Product Specification
- 182 -
PR interval (ms)
QRS interval (ms)
QT/QTC (ms)
P-QRS-T AXIS
A.5 RESP
Method
Impedance between RA-LL, RA-LA
Measurement lead
Options are lead I and II. The default is Lead II.
Calculation Type
Manual, Automatic
Baseline Impedance Range
200 Ω to 2500 Ω (with ECG cables of 1 KΩ resistance)
Measuring Sensitivity
Within the baseline impedance range: 0.3 Ω
Waveform Bandwidth
0.2 Hz to 2.5 Hz (-3 dB)
Respiration Excitation
Waveform
Sinusoid, 45.6 kHz (10%), < 350 μA
☆RR Measuring Range
☆Adult
0 rpm to120 rpm
☆Neo/Ped
0 rpm to150 rpm
Resolution
1 rpm
☆Accuracy
☆Adult
6 rpm to 120 rpm: 2 rpm
0 rpm to 5 rpm: not specified
☆Neo/Ped
6 rpm to 150 rpm: 2 rpm
0 rpm to 5 rpm: not specified
☆Gain Selection
0.25, 0.5, 1, 2, 3, 4, 5
☆Sweep
6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
☆Apnea Delay
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
A.6 NIBP
Complies with IEC 80601-2-30: 2009+A1: 2013
Technique
Oscillometry