Product Manual

Patient Monitor User Manual Product Specification
- 183 -
Mode
Manual, Auto, Continuous, Sequence
Measuring Interval in AUTO
Mode (unit: minutes)
1/2/3/4/5/10/15/30/60/90/120/180/240/360/480
and User Define
Continuous
5 min, interval is 5 s
Measuring Parameter
SYS, DIA, MAP, PR
Pressure Unit
kPa, mmHg, cmH
2
O
Measuring Range
Adult Mode
SYS: 25 mmHg to 290 mmHg
DIA: 10 mmHg to 250 mmHg
MAP: 15 mmHg to 260 mmHg
Pediatric Mode
SYS: 25 mmHg to 240 mmHg
DIA: 10 mmHg to 200 mmHg
MAP: 15 mmHg to 215 mmHg
Neonatal Mode
SYS: 25 mmHg to 140 mmHg
DIA: 10 mmHg to 115 mmHg
MAP: 15 mmHg to 125 mmHg
Alarm Type
SYS, DIA, MAP, PR (NIBP)
Cuff Pressure Measuring
Range
0 mmHg to 300 mmHg
Pressure Resolution
1 mmHg
Maximum Mean Error
±5 mmHg
Maximum Standard Deviation
8 mmHg
Maximum Measuring Period
Adult/Pediatric
120 s
Neonate
90 s
Typical Measuring Period
20 s to 35 s (depend on HR/motion disturbance)
Dual Independent Channel Overpressure Protection
Adult
(297±3) mmHg
Pediatric
(245±3 ) mmHg
Neonatal
(147±3) mmHg
Pre-inflation Pressure
Adult Mode
Range: 80/100/120/140/150/160/180/200/220/240 mmHg
Pediatric Mode
Range: 80/100/120/140/150/160/180/200 mmHg
Neonatal Mode
Range: 60/70/80/100/120 mmHg