Product Manual
Patient Monitor User Manual Product Specification
- 184 -
Venipuncture pressure
Adult
Default: 60 mmHg
Options: 20 mmHg, 30 mmHg, 40 mmHg, 50 mmHg, 60
mmHg, 70 mmHg, 80 mmHg, 90 mmHg, 100 mmHg, 110
mmHg, 120 mmHg
Pediatric
Default: 40 mmHg
Options: 20 mmHg, 30 mmHg, 40 mmHg, 50 mmHg, 60
mmHg, 70 mmHg, 80 mmHg
Neonatal
Default: 30 mmHg
Options: 20 mmHg, 30 mmHg, 40 mmHg, 50 mmHg
A.7 SpO
2
Complies with ISO 80601-2-61: 2017.
Measuring Range
0% to 100%
Resolution
1%
☆Data Update Period
1 s
☆Accuracy
☆Adult /Pediatric
2% (70% to 100% SpO
2
)
Undefined (0% to 69% SpO
2
)
☆Neonate
3% (70% to 100% SpO
2
)
Undefined (0% to 69% SpO
2
)
Sensor
Red Light
(6603) nm
Infrared Light
(90510) nm
Emitted Light Energy
< 15 mW
PI
Measuring Range
0-10, invalid PI value is 0.
Resolution
1
NOTE:
The information about wavelength range can be especially useful to clinicians (for
instance, when photodynamic therapy is performed).