Patient Monitor User Manual Product Specification
- 185 -
A.8 TEMP
Complies with ISO 80601-2-56: 2018.
Technique
Thermal resistance
Position
Skin, oral cavity, rectum
Measure Parameter
T1, T2, TD (the absolute value of T2 minus T1)
Channel
X8: 1
X10/X12: 2
Sensor Type
YSI-10K and YSI-2.252K
Unit
C, F
Measuring Range
0 C to 50 C (32 ºF to 122 ºF)
Resolution
0.1 C (0.1 ºF)
☆Accuracy
1
0.3 C
Refresh Time
Every 1 s to 2 s
Temperature Calibration
At an interval of 5 to 10 minutes
Measuring Mode
Direct Mode
Transient Response Time
≤ 30 s
Note 1: The accuracy consists of two parts, as following:
Accuracy (not including sensor): ± 0.1 C
Sensor accuracy: ≤ ±0.2 C
A.9 PR
Measuring range
Accuracy
Resolution
PR (SpO
2
)
25 bpm to 300 bpm
2 bpm
1 bpm
PR
(NIBP)
40 bpm to 240 bpm
± 3 bpm or 3.5%, whichever is greater
1 bpm
PR (IBP)
20 bpm to 300 bpm
30 bpm to 300 bpm: 2 bpm or 2%,
whichever is greater;
20 bpm to 29 bpm: undefined
1 bpm
A.10 IBP
Complies with IEC 60601-2-34: 2011.
Technique
Direct invasive measurement
Channel
2 channels