Product Manual

Patient Monitor User Manual Product Specification
- 186 -
IBP
Measure
Measuring
Range
Art
(0 to +300) mmHg
PA/PAWP
(-6 to +120) mmHg
CVP/RAP/LAP/ICP
(-10 to +40) mmHg
P1/P2
(-50 to +300) mmHg
Resolution
1 mmHg
Accuracy (not including sensor)
2 % or 1 mmHg, whichever is greater
ICP:
0 mmHg to 40 mmHg: 2 % or 1
mmHg, whichever is greater;
-10 mmHg to -1 mmHg: undefined
Pressure Unit
kPa, mmHg, cmH
2
O
Pressure sensor
Sensitivity
5 μV/V/mmHg
Impedance Range
300 Ω to 3000 Ω
Filter
DC~ 12.5 Hz; DC~ 40 Hz
Zero
Range: 200 mmHg
Pressure Calibration
Range
IBP (excluding ICP)
80 mmHg to 300 mmHg
ICP
10 mmHg to 40 mmHg
Volume Displacement
7.4 x 10
4
mm
3
/ 100 mmHg
A.11 CO
2
Complies with ISO 80601-2-55: 2011.
Intended
Patient
Adult, pediatric, neonatal
Measure
Parameters
EtCO
2
, FiCO
2
, AwRR
Unit
mmHg, %, kPa
Measuring
Range
EtCO
2
0 mmHg to 150 mmHg (0 % to 20%)
FiCO
2
0 mmHg to 50 mmHg
AwRR
2 rpm to 150 rpm
Resolution
EtCO
2
1 mmHg
FiCO
2
1 mmHg
AwRR
1 rpm