Product Manual
Patient Monitor User Manual Product Specification
- 188 -
Zero
Calibration
Support
Calibration
Support (It is recommend to be operated by trained personal.)
☆Alarm
EtCO
2
, FiCO
2
, AwRR
☆ Apnea
delay
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
Data Sample
Rate
100 Hz
EtCO
2
Change
1
AwRR ≤ 80 rpm, meet the accuracy
mentioned above;
AwRR > 80 rpm, EtCO
2
descends 8%;
AwRR > 120 rpm, EtCO
2
descends
10%;
with 2 m gas sampling tube, sample
gas flowrate: 100 ml/min)
AwRR ≤ 60 rpm, meet the accuracy
mentioned above;
AwRR > 60 rpm, EtCO
2
descends 8%;
AwRR > 90 rpm, EtCO
2
descends 10%;
AwRR > 120 rpm, EtCO
2
descends
15%;
with 2 m gas sampling tube, sample
gas flowrate: 70 ml/min)
AwRR ≤ 20 rpm, meet the accuracy
mentioned above;
AwRR > 20 rpm, EtCO
2
descends 8%;
AwRR > 40 rpm, not applicable;
with 2 m gas sampling tube, sample
gas flowrate: 50 ml/min)
Note 1: Use a test device equivalent to EN ISO 80601-2-55 fig 201.101 to measure at 1:2 I/E
ratio. Respiration rate accuracy is determined by frequency of device, and end-tidal reading
change refers to the nominal value.
Interfering Gas Effects:
Gas
Gas Level (%)
Quantitative Effect/Comments
Nitrous oxide
60%
None
Halothane
4%
None
Enflurane
5%
None
Isoflurane
5%
None
Sevoflurane
5%
None
Xenon
Not applicable
Not applicable
Hehelium
Not applicable
Not applicable
Metered dose inhaler propellants
Not applicable
Not applicable
Desflurane
15%
None