- 190 -
A.13 Interfaces
A.13.1 Analog Output (Optional)
Bandwidth (-3 dB; reference
frequency: 10 Hz)
Monitor: 0.5 Hz to 40 Hz
Diagnosis/Diagnosis 1: 0.05 Hz to 40 Hz
Surgery: 1 Hz to 20 Hz
Enhanced: 2 Hz ~18 Hz
Customized: When Low-pass Filter < 40 Hz, Bandwidth
is High-pass Filter ~ Low-pass Filter; When Low-pass
Filter > 40 Hz, Bandwidth is High-pass ~40 Hz.
Maximum Transmission Delay
(Diagnosis Mode)
500 ms
Sensitivity
1 V/1 mV ±10%
PACE Rejection/ Enhancement
No PACE Rejection or Enhancement
Waveform Display
Consistent with the calculation leads.
Compliant with Standard and
Directive
Complies with the requirements in terms of short circuit
protection and leakage current in EN60601-1.
Output Impedance
< 500 Ω
Interface Type
PS2 connector
NOTE:
While using analog output, set the calculation lead as following:
1) In 3 Electrodes mode, set to Lead I, Lead II, or Lead III.
2) In 5 Electrodes mode, set to Lead I, Lead II, Lead III or Lead V.
3) In 6 Electrodes mode, set to I, II, III, and leads corresponding to Va, Vb.
4) In 10 Electrodes mode, set to Lead I, Lead II, Lead III or Lead V1~V6.
A.13.2 Defibrillator Synchronization (Optional)
Output Impedance
< 500 Ω
Maximum Time Delay
35 ms (R-wave peak to leading edge of pulse)
Waveform
Rectangular wave
Amplitude
High level: 3.5 V to 5.5 V, providing a maximum of 1 mA
output current;
Low level: < 0.5 V, receiving a maximum of 5 mA input
current